Electrocardiogram Cream Reduces Skin-Electrode Impedance Upon Neuromuscular Monitoring Using TOF-Cuff®.

electrocardiogram cream impedance muscle relaxant neuromuscular block monitoring neuromuscular monitoring

Journal

Cureus
ISSN: 2168-8184
Titre abrégé: Cureus
Pays: United States
ID NLM: 101596737

Informations de publication

Date de publication:
Sep 2023
Historique:
accepted: 04 09 2023
medline: 6 10 2023
pubmed: 6 10 2023
entrez: 6 10 2023
Statut: epublish

Résumé

Background Mechanistic insight into the high failure rate of TOF-Cuff® (RGB Medical Devices, Madrid, Spain) measurements on the lower leg is unclear. Aims We aimed to determine whether materials applied to pseudo-skin can reduce the impedance between a model arm and TOF-Cuff® electrodes and whether a material between TOF-Cuff® electrodes and the patient's skin surface decreases the skin-TOF-Cuff® electrode impedance within the appropriate range. Methods This was a combination of an in vitro study using non-living materials and a prospective observational clinical study. Eight patients aged > 70 years who had undergone elective surgery were eligible. One of the primary outcomes was whether water, electrocardiogram (ECG) cream, or ECG gel applied on the pseudo-skin could reduce the impedance between the model arm and the TOF-Cuff® electrodes in the in vitro study. Another was whether a material between the TOF-Cuff® electrodes and the patient's skin surface decreased the skin-TOF-Cuff® electrode impedance to an appropriate level of less than 5,000 Ω in the clinical study. Results The application of water, ECG cream, and ECG gel similarly reduced the impedance values within the electrical circuit in the in vitro study. ECG cream application between the patient's skin surface and the TOF-Cuff® electrodes decreased the skin-TOF-Cuff® electrode impedance (median (interquartile range (IQR)) Ω) from 8,600 (6,450 to 9,775) to 2,000 (1,600 to 2,600) (P = 0.012) in surgical patients. Conclusion ECG cream application between the patient's skin surface and the TOF-Cuff® electrodes decreased the skin-TOF-Cuff® electrode impedance appropriately, and thus, the application can facilitate precise TOF-Cuff® measurements in patients.

Identifiants

pubmed: 37799239
doi: 10.7759/cureus.44670
pmc: PMC10550304
doi:

Types de publication

Journal Article

Langues

eng

Pagination

e44670

Informations de copyright

Copyright © 2023, Sugimura et al.

Déclaration de conflit d'intérêts

The authors have declared that no competing interests exist.

Références

Anesthesiology. 2023 Jan 1;138(1):13-41
pubmed: 36520073
Eur J Anaesthesiol. 2020 Jan;37(1):38-43
pubmed: 31592900
Br J Anaesth. 2019 Feb;122(2):e22-e24
pubmed: 30686321
Anesthesiology. 1988 Jul;69(1):129-31
pubmed: 3389551
Clin Neurophysiol. 2020 Jan;131(1):243-258
pubmed: 31761717
J Invest Dermatol. 1984 Jan;82(1):97-100
pubmed: 6690633
Acta Anaesthesiol Scand. 2005 Nov;49(10):1552-8
pubmed: 16223405
J Anesth. 2020 Feb;34(1):79-85
pubmed: 31760477
Acta Anaesthesiol Scand. 2020 Feb;64(2):173-179
pubmed: 31593293
Minerva Anestesiol. 2020 Jul;86(7):704-711
pubmed: 32068983

Auteurs

Sho Sugimura (S)

Departments of Anesthesiology and Intensive Care, Hamamatsu University School of Medicine, Hamamatsu, JPN.

Huynh V Khanh (HV)

Departments of Anesthesiology and Intensive Care, Hamamatsu University School of Medicine, Hamamatsu, JPN.

Shingo Kawashima (S)

Departments of Anesthesiology and Intensive Care, Hamamatsu University School of Medicine, Hamamatsu, JPN.

Yoshiki Nakajima (Y)

Departments of Anesthesiology and Intensive Care Medicine, Hamamatsu University School of Medicine, Hamamatsu, JPN.

Hiroyuki Kinoshita (H)

Departments of Anesthesiology and Intensive Care, Hamamatsu University School of Medicine, Hamamatsu, JPN.
Department of Dental Anesthesiology, Tokushima University Graduate School of Biomedical Sciences, Tokushima, JPN.

Classifications MeSH