Effect of intraocular lens material and haptic design on anterior capsule contraction after cataract surgery: A systematic review and meta-analysis.

Anterior capsule contraction Cataract Intraocular lens Meta-analysis

Journal

Graefe's archive for clinical and experimental ophthalmology = Albrecht von Graefes Archiv fur klinische und experimentelle Ophthalmologie
ISSN: 1435-702X
Titre abrégé: Graefes Arch Clin Exp Ophthalmol
Pays: Germany
ID NLM: 8205248

Informations de publication

Date de publication:
13 Oct 2023
Historique:
received: 20 02 2023
accepted: 01 09 2023
revised: 16 05 2023
medline: 13 10 2023
pubmed: 13 10 2023
entrez: 13 10 2023
Statut: aheadofprint

Résumé

To compare anterior capsule contraction (ACC) after cataract surgery with implantation of intraocular lens (IOLs) of different materials and designs. We searched three electronic databases for relevant studies published up to January 1, 2023. Five randomized controlled trails (RCTs) and three cohort studies involving 1,221 eyes were included in quantitative synthesis. We extracted data, assessed their quality independently, and calculated standard mean difference (SMD) using a random-effects model. Six RCTs and one retrospective cohort were included in information summary. The contraction of the anterior capsule opening area in the hydrophilic group was larger than that of the hydrophobic group from one month to one year postoperatively (P < 0.001 and P < 0.001, respectively). Specifically, the hydrophilic group showed greater contraction of the anterior capsule opening area at one month postoperatively (Standardized mean difference [SMD] = -0.73, 95% confidence interval [CI] = -0.93 to -0.52), three months (SMD = -1.04, 95% CI = -1.32 to -0.75), six months (SMD = -0.99, 95% CI = -1.24 to -0.74) and one year (SMD = -1.33, 95% CI = -2.50 to -0.16). As of one year postoperatively, the anterior capsular opening area showed a trend of decreasing over time in both groups (P = 0.046 and P = 0.050, respectively). In information summary, three studies indicated no relationship between haptic design and ACC, while the other four studies reported that the number and shape of haptic would affect ACC. This meta-analysis suggested that the postoperative ACC after the implantation of hydrophobic IOLs was less than that induced by hydrophilic IOLs. Haptic design may also affect the degree of ACC.

Identifiants

pubmed: 37831171
doi: 10.1007/s00417-023-06230-x
pii: 10.1007/s00417-023-06230-x
doi:

Types de publication

Journal Article Review

Langues

eng

Sous-ensembles de citation

IM

Subventions

Organisme : National Natural Science Foundation of China
ID : 82271063
Organisme : National Natural Science Foundation of China
ID : 81670833
Organisme : National Natural Science Foundation of China
ID : 81870641
Organisme : National Natural Science Foundation of China
ID : 8207939
Organisme : National Natural Science Foundation of China
ID : 81300641
Organisme : Fundamental Research Funds of the Central Universities
ID : 2019QNA7026
Organisme : Zhejiang Province Key Research and Development Program
ID : 2019C03091
Organisme : Zhejiang Province Key Research and Development Program
ID : 2020C03035

Informations de copyright

© 2023. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.

Références

World Health Organization (2022) Blindness and vision impairment. World Health Organization.  https://www.who.int/news-room/fact-sheets/detail/blindness-and-visual-impairment . Accessed 23 Apr, 2023
Center for Disease Control and Prevention (2022) Common Eye Disorders and Diseases.  https://www.cdc.gov/visionhealth/basics/ced/index.html . Accessed 23 Apr 2023
NVISION (2022) The Statistics on Eye Disease in America. NVISION Eye Centers.  https://www.nvisioncenters.com/education/eye-disease-statistics/ . Accessed 23 Apr, 2023
Chen X, Xu J, Chen X, Yao K (2021) Cataract: Advances in surgery and whether surgery remains the only treatment in future. Adv Ophthalmol Pract Res 1:2667–3762
Davison JA (1993) Capsule contraction syndrome. J Cataract Refract Surg 19:582–589
pubmed: 8229711 doi: 10.1016/S0886-3350(13)80004-1
Hansen SO, Crandall AS, Olson RJ (1993) Progressive constriction of the anterior capsular opening following intact capsulorhexis. J Cataract Refract Surg 19:77–82
pubmed: 8426328 doi: 10.1016/S0886-3350(13)80287-8
Deokule SP, Mukherjee SS, Chew CK (2006) Neodymium:YAG laser anterior capsulotomy for capsular contraction syndrome. Ophthalmic Surg Lasers Imaging 37:99–105
pubmed: 16583630 doi: 10.3928/1542-8877-20060301-04
Hayashi H, Hayashi K, Nakao F, Hayashi F (1998) Anterior capsule contraction and intraocular lens dislocation in eyes with pseudoexfoliation syndrome. Br J Ophthalmol 82:1429–1432
pubmed: 9930277 pmcid: 1722461 doi: 10.1136/bjo.82.12.1429
Hayashi Y, Kato S, Fukushima H, Numaga J, Kaiya T, Tamaki Y, Oshika T (2004) Relationship between anterior capsule contraction and posterior capsule opacification after cataract surgery in patients with diabetes mellitus. J Cataract Refract Surg 30:1517–1520
pubmed: 15210231 doi: 10.1016/j.jcrs.2003.11.045
Matsuda H, Kato S, Hayashi Y, Amano S, Hattori T, Yuguchi T, Kaiya T, Oshika T (2001) Anterior capsular contraction after cataract surgery in vitrectomized eyes. Am J Ophthalmol 132:108–109
pubmed: 11438064 doi: 10.1016/S0002-9394(00)00937-5
Michael K, O’Colmain U, Vallance JH, Cormack TG (2010) Capsule contraction syndrome with haptic deformation and flexion. J Cataract Refract Surg 36:686–689
pubmed: 20362866 doi: 10.1016/j.jcrs.2009.09.042
Kato S, Suzuki T, Hayashi Y, Numaga J, Hattori T, Yuguchi T, Kaiya T, Oshika T (2002) Risk factors for contraction of the anterior capsule opening after cataract surgery. J Cataract Refract Surg 28:109–112
pubmed: 11777718 doi: 10.1016/S0886-3350(01)00901-4
Hayashi K, Hayashi H (2005) Intraocular lens factors that may affect anterior capsule contraction. Ophthalmology 112:286–292
pubmed: 15691565 doi: 10.1016/j.ophtha.2004.11.013
Miyake K, Ota I, Miyake S, Maekubo K (1996) Correlation between intraocular lens hydrophilicity and anterior capsule opacification and aqueous flare. J Cataract Refract Surg 22(Suppl 1):764–769
pubmed: 9279669 doi: 10.1016/S0886-3350(96)80159-3
Hayashi K, Hayashi H, Nakao F, Hayashi F (2001) Anterior capsule contraction and intraocular lens decentration and tilt after hydrogel lens implantation. Br J Ophthalmol 85:1294–1297
pubmed: 11673291 pmcid: 1723752 doi: 10.1136/bjo.85.11.1294
Cochener B, Jacq PL, Colin J (1999) Capsule contraction after continuous curvilinear capsulorhexis: poly(methyl methacrylate) versus silicone intraocular lenses. J Cataract Refract Surg 25:1362–1369
pubmed: 10511936 doi: 10.1016/S0886-3350(99)00227-8
Dahlhauser KF, Wroblewski KJ, Mader TH (1998) Anterior capsule contraction with foldable silicone intraocular lenses. J Cataract Refract Surg 24:1216–1219
pubmed: 9768395 doi: 10.1016/S0886-3350(98)80014-X
Wang Y, Wang W, Zhu Y, Xu J, Luo C, Yao K (2022) Comparison Study of Anterior Capsule Contraction of Hydrophilic and Hydrophobic Intraocular Lenses Under the Same Size Capsulotomy. Transl Vis Sci Technol 11:24
pubmed: 35040916 pmcid: 8764210 doi: 10.1167/tvst.11.1.24
Sacu S, Menapace R, Findl O (2006) Effect of optic material and haptic design on anterior capsule opacification and capsulorrhexis contraction. Am J Ophthalmol 141:488–493
pubmed: 16490494 doi: 10.1016/j.ajo.2005.10.041
Tsinopoulos IT, Tsaousis KT, Kymionis GD, Symeonidis C, Grentzelos MA, Diakonis VF, Adaloglou M, Dimitrakos SA (2010) Comparison of anterior capsule contraction between hydrophobic and hydrophilic intraocular lens models. Graefes Arch Clin Exp Ophthalmol 248:1155–1158
pubmed: 20414785 doi: 10.1007/s00417-010-1373-2
Kim SY, Yang JW, Lee YC, Kim SY (2013) Effect of haptic material and number of intraocular lens on anterior capsule contraction after cataract surgery. Korean J Ophthalmol 27:7–11
pubmed: 23372373 pmcid: 3550319 doi: 10.3341/kjo.2013.27.1.7
Choi M, Lazo MZ, Kang M, Lee J, Joo CK (2018) Effect of number and position of intraocular lens haptics on anterior capsule contraction: a randomized, prospective trial. BMC Ophthalmol 18:78
pubmed: 29558909 pmcid: 5859398 doi: 10.1186/s12886-018-0742-1
Moher D, Liberati A, Tetzlaff J, Altman DG, Group P (2009) Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. PLoS Med 6:e1000097
Jadad AR, Moore RA, Carroll D, Jenkinson C, Reynolds DJ, Gavaghan DJ, McQuay HJ (1996) Assessing the quality of reports of randomized clinical trials: is blinding necessary? Control Clin Trials 17:1–12
pubmed: 8721797 doi: 10.1016/0197-2456(95)00134-4
The Newcastle-Ottawa Scale (NOS) for assessing the quality if nonrandomized studies in meta-analyses. [online] Available at: http://www.ohri.ca/programs/clinical_epidemiology/oxford.htm . Accessed 23 Apr 2023
DerSimonian R, Kacker R (2007) Random-effects model for meta-analysis of clinical trials: an update. Contemp Clin Trials 28:105–114
pubmed: 16807131 doi: 10.1016/j.cct.2006.04.004
Huedo-Medina TB, Sanchez-Meca J, Marin-Martinez F, Botella J (2006) Assessing heterogeneity in meta-analysis: Q statistic or I2 index? Psychol Methods 11:193–206
pubmed: 16784338 doi: 10.1037/1082-989X.11.2.193
Begg CB, Mazumdar M (1994) Operating characteristics of a rank correlation test for publication bias. Biometrics 50:1088–1101
pubmed: 7786990 doi: 10.2307/2533446
Egger M, Davey Smith G, Schneider M, Minder C (1997) Bias in meta-analysis detected by a simple, graphical test. BMJ 315:629–634
pubmed: 9310563 pmcid: 2127453 doi: 10.1136/bmj.315.7109.629
Sutton AJ, Duval SJ, Tweedie RL, Abrams KR, Jones DR (2000) Empirical assessment of effect of publication bias on meta-analyses. BMJ 320:1574–1577
pubmed: 10845965 pmcid: 27401 doi: 10.1136/bmj.320.7249.1574
Cui XCM, YH, Fu XX (2009) Relationship of optic materials with anterior and posterior capsule opacification as well as capsulorhexis contraction of lenses. J Clin Rehabil Tissue Eng Res 13:4875-4878
Georgopoulos M, Menapace R, Findl O, Rainer G, Petternel V, Kiss B (2001) Posterior continuous curvilinear capsulorhexis with hydrogel and silicone intraocular lens implantation: development of capsulorhexis size and capsule opacification. J Cataract Refract Surg 27:825–832
pubmed: 11408126 doi: 10.1016/S0886-3350(00)00793-8
Hayashi K, Hayashi H, Nakao F, Hayashi F (1997) Reduction in the area of the anterior capsule opening after polymethylmethacrylate, silicone, and soft acrylic intraocular lens implantation. Am J Ophthalmol 123:441–447
pubmed: 9124240 doi: 10.1016/S0002-9394(14)70169-2
Yang S, Lim SA, Na KS, Joo CK (2017) Comparison of Anterior Capsule Stability Following Implantation of Three Single Piece Acrylic Intraocular Lenses with Different Haptic Design. Korean J Ophthalmol 31:32–38
pubmed: 28243021 pmcid: 5327172 doi: 10.3341/kjo.2017.31.1.32
Mingels A, Koch J, Lommatzsch A, Pauleikhoff D, Heiligenhaus A (2007) Comparison of two acrylic intraocular lenses with different haptic designs in patients with combined phacoemulsification and pars plana vitrectomy. Eye (Lond) 21:1379–1383
pubmed: 16763658 doi: 10.1038/sj.eye.6702446
Patel CK, Ormonde S, Rosen P, Bron AJ (2000) Post-operative changes in the capsulorhexis aperture: a prospective, randomised comparison between loop and plate haptic silicone intraocular lenses. Eye (Lond) 14(Pt 2):185–189
pubmed: 10845014 doi: 10.1038/eye.2000.51
Nguyen CL, Francis IC (2017) Mechanical anterior lens capsule polishing under viscoelastic during phacoemulsification cataract surgery. Clin Exp Ophthalmol 45:654–656
pubmed: 28165186 doi: 10.1111/ceo.12924
Shah SK, Praveen MR, Kaul A, Vasavada AR, Shah GD, Nihalani BR (2009) Impact of anterior capsule polishing on anterior capsule opacification after cataract surgery: a randomized clinical trial. Eye (Lond) 23:1702–1706
pubmed: 19079142 doi: 10.1038/eye.2008.355
Corydon C, Lindholt M, Knudsen EB, Graakjaer J, Corydon TJ, Dam-Johansen M (2007) Capsulorhexis contraction after cataract surgery: comparison of sharp anterior edge and modified anterior edge acrylic intraocular lenses. J Cataract Refract Surg 33:796–799
pubmed: 17466850 doi: 10.1016/j.jcrs.2007.01.020
Nagata T, Minakata A, Watanabe I (1998) Adhesiveness of AcrySof to a collagen film. J Cataract Refract Surg 24:367–370
pubmed: 9559473 doi: 10.1016/S0886-3350(98)80325-8
Abela-Formanek C, Amon M, Kahraman G, Schauersberger J, Dunavoelgyi R (2011) Biocompatibility of hydrophilic acrylic, hydrophobic acrylic, and silicone intraocular lenses in eyes with uveitis having cataract surgery: Long-term follow-up. J Cataract Refract Surg 37:104–112
pubmed: 21183105 doi: 10.1016/j.jcrs.2010.07.038
Richter-Mueksch S, Kahraman G, Amon M, Schild-Burggasser G, Schauersberger J, Abela-Formanek C (2007) Uveal and capsular biocompatibility after implantation of sharp-edged hydrophilic acrylic, hydrophobic acrylic, and silicone intraocular lenses in eyes with pseudoexfoliation syndrome. J Cataract Refract Surg 33:1414–1418
pubmed: 17662434 doi: 10.1016/j.jcrs.2007.05.009
Saika S, Miyamoto T, Ohnishi Y (2003) Histology of anterior capsule opacification with a polyHEMA/HOHEXMA hydrophilic hydrogel intraocular lens compared to poly(methyl methacrylate), silicone, and acrylic lenses. J Cataract Refract Surg 29:1198–1203
pubmed: 12842690 doi: 10.1016/S0886-3350(02)01988-0
Kranitz K, Takacs A, Mihaltz K, Kovacs I, Knorz MC, Nagy ZZ (2011) Femtosecond laser capsulotomy and manual continuous curvilinear capsulorrhexis parameters and their effects on intraocular lens centration. J Refract Surg 27:558–563
pubmed: 21710951 doi: 10.3928/1081597X-20110623-03
Grewal DS, Schultz T, Basti S, Dick HB (2016) Femtosecond laser-assisted cataract surgery–current status and future directions. Surv Ophthalmol 61:103–131
pubmed: 26409902 doi: 10.1016/j.survophthal.2015.09.002
Nagy ZZ, Kranitz K, Takacs AI, Mihaltz K, Kovacs I, Knorz MC (2011) Comparison of intraocular lens decentration parameters after femtosecond and manual capsulotomies. J Refract Surg 27:564–569
pubmed: 21688765 doi: 10.3928/1081597X-20110607-01
Weiblinger RP (1986) Review of the clinical literature on the use of the Nd:YAG laser for posterior capsulotomy. J Cataract Refract Surg 12:162–170
pubmed: 3517297 doi: 10.1016/S0886-3350(86)80034-7
Spalton DJ (1999) Posterior capsular opacification after cataract surgery. Eye (Lond) 13(Pt 3b):489–492
pubmed: 10627830 doi: 10.1038/eye.1999.127
Gonvers M, Sickenberg M, van Melle G (1997) Change in capsulorhexis size after implantation of three types of intraocular lenses. J Cataract Refract Surg 23:231–238
pubmed: 9113574 doi: 10.1016/S0886-3350(97)80346-X
Moreno-Montanes J, Sanchez-Tocino H, Rodriguez-Conde R (2002) Complete anterior capsule contraction after phacoemulsification with acrylic intraocular lens and endocapsular ring implantation. J Cataract Refract Surg 28:717–719
pubmed: 11955919 doi: 10.1016/S0886-3350(01)01231-7
Efremov YM, Bakhchieva NA, Shavkuta BS, Frolova AA, Kotova SL, Novikov IA, Akovantseva AA, Avetisov KS, Avetisov SE, Timashev PS (2020) Mechanical properties of anterior lens capsule assessed with AFM and nanoindenter in relation to human aging, pseudoexfoliation syndrome, and trypan blue staining. J Mech Behav Biomed Mater 112:104081
pubmed: 32961392 doi: 10.1016/j.jmbbm.2020.104081
Pedrigi RM, Dziezyc J, Humphrey JD (2009) Altered mechanical behavior and properties of the human anterior lens capsule after cataract surgery. Exp Eye Res 89:575–580
pubmed: 19523944 doi: 10.1016/j.exer.2009.06.001

Auteurs

Lu Chen (L)

Eye Center, The Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, 310009, Zhejiang Province, China.
Zhejiang Provincial Key Laboratory of Ophthalmology, Zhejiang Provincial Clinical Research Center for Eye Diseases, Zhejiang Provincial Engineering Institute On Eye Diseases, Hangzhou, 310009, Zhejiang Province, China.

Yueyang Zhong (Y)

Eye Center, The Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, 310009, Zhejiang Province, China.
Zhejiang Provincial Key Laboratory of Ophthalmology, Zhejiang Provincial Clinical Research Center for Eye Diseases, Zhejiang Provincial Engineering Institute On Eye Diseases, Hangzhou, 310009, Zhejiang Province, China.

Ke Yao (K)

Eye Center, The Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, 310009, Zhejiang Province, China. xlren@zju.edu.cn.
Zhejiang Provincial Key Laboratory of Ophthalmology, Zhejiang Provincial Clinical Research Center for Eye Diseases, Zhejiang Provincial Engineering Institute On Eye Diseases, Hangzhou, 310009, Zhejiang Province, China. xlren@zju.edu.cn.

Qiuli Fu (Q)

Eye Center, The Second Affiliated Hospital, School of Medicine, Zhejiang University, Hangzhou, 310009, Zhejiang Province, China. 2313009@zju.edu.cn.
Zhejiang Provincial Key Laboratory of Ophthalmology, Zhejiang Provincial Clinical Research Center for Eye Diseases, Zhejiang Provincial Engineering Institute On Eye Diseases, Hangzhou, 310009, Zhejiang Province, China. 2313009@zju.edu.cn.

Classifications MeSH