Load distribution on intervertebral cages with and without posterior instrumentation.
Fusion
Intervertebral cage
PLIF
TLIF
Journal
The spine journal : official journal of the North American Spine Society
ISSN: 1878-1632
Titre abrégé: Spine J
Pays: United States
ID NLM: 101130732
Informations de publication
Date de publication:
02 Nov 2023
02 Nov 2023
Historique:
received:
10
05
2023
revised:
18
09
2023
accepted:
28
10
2023
pubmed:
5
11
2023
medline:
5
11
2023
entrez:
4
11
2023
Statut:
aheadofprint
Résumé
Posterior and transforaminal lumbar interbody fusion (PLIF, TLIF) are well-established procedures for spinal fusion. However, little is known about load sharing between cage, dorsal construct, and biological tissue within the instrumented lumbar spine. The aim of this study was to quantify the forces acting on cages under axial compression force with and without posterior instrumentation. Biomechanical cadaveric study. Ten lumbar spinal segments were tested under uniaxial compression using load cell instrumented intervertebral cages. The force was increased in 100N increments to 1000N or a force greater than 500N on one load cell. Each specimen was tested after unilateral PLIF (uPLIF), bilateral PLIF (bPLIF) and TLIF each with/without posterior instrumentation. Dorsal instrumentation was performed with 55N of compression per side. Cage insertion resulted in median cage preloads of 16N, 29N and 35N for uPLIF, bPLIF, and TLIF. The addition of compressed dorsal instrumentation increased the median preload to 224N, 328N, and 317N, respectively. With posterior instrumentation, the percentage of the external load acting on the intervertebral cage was less than 25% at 100N (uPLIF: 14.2%; bPLIF: 16%; TLIF: 11%), less than 45% at 500N (uPLIF: 31.8%; bPLIF: 41.1%; TLIF: 37.9%) and less than 50% at 1000N (uPLIF: 40.3%; bPLIF: 49.7%; TLIF: 43.4%). Without posterior instrumentation, the percentage of external load on the cages was significantly higher with values above 50% at 100N (uPLIF: 55.6%; bPLIF: 75.5%; TLIF: 66.8%), 500N (uPLIF: 71.7%; bPLIF: 79.2%; TLIF: 65.4%), and 1000N external load (uPLIF: 73%; bPLIF: 80.5%; TLIF: 66.1%). For absolute loads, preloads and external loads must be added together. Without posterior instrumentation, the intervertebral cages absorb more than 50% of the axial load and the load distribution is largely independent of the loading amplitude. With posterior instrumentation, the external load acting on the cages is significantly lower and the load distribution becomes load amplitude dependent, with a higher proportion of the load transferred by the cages at high loads. The bPLIF cages tend to absorb more force than the other two cage configurations. Cage instrumentation allows some of the compression force to be transmitted through the cage to the screws below, better distributing and reducing the overall force on the pedicle screws at the end of the construct and on the rods.
Sections du résumé
BACKGROUND CONTEXT
BACKGROUND
Posterior and transforaminal lumbar interbody fusion (PLIF, TLIF) are well-established procedures for spinal fusion. However, little is known about load sharing between cage, dorsal construct, and biological tissue within the instrumented lumbar spine.
PURPOSE
OBJECTIVE
The aim of this study was to quantify the forces acting on cages under axial compression force with and without posterior instrumentation.
STUDY DESIGN
METHODS
Biomechanical cadaveric study.
METHODS
METHODS
Ten lumbar spinal segments were tested under uniaxial compression using load cell instrumented intervertebral cages. The force was increased in 100N increments to 1000N or a force greater than 500N on one load cell. Each specimen was tested after unilateral PLIF (uPLIF), bilateral PLIF (bPLIF) and TLIF each with/without posterior instrumentation. Dorsal instrumentation was performed with 55N of compression per side.
RESULTS
RESULTS
Cage insertion resulted in median cage preloads of 16N, 29N and 35N for uPLIF, bPLIF, and TLIF. The addition of compressed dorsal instrumentation increased the median preload to 224N, 328N, and 317N, respectively. With posterior instrumentation, the percentage of the external load acting on the intervertebral cage was less than 25% at 100N (uPLIF: 14.2%; bPLIF: 16%; TLIF: 11%), less than 45% at 500N (uPLIF: 31.8%; bPLIF: 41.1%; TLIF: 37.9%) and less than 50% at 1000N (uPLIF: 40.3%; bPLIF: 49.7%; TLIF: 43.4%). Without posterior instrumentation, the percentage of external load on the cages was significantly higher with values above 50% at 100N (uPLIF: 55.6%; bPLIF: 75.5%; TLIF: 66.8%), 500N (uPLIF: 71.7%; bPLIF: 79.2%; TLIF: 65.4%), and 1000N external load (uPLIF: 73%; bPLIF: 80.5%; TLIF: 66.1%). For absolute loads, preloads and external loads must be added together.
CONCLUSIONS
CONCLUSIONS
Without posterior instrumentation, the intervertebral cages absorb more than 50% of the axial load and the load distribution is largely independent of the loading amplitude. With posterior instrumentation, the external load acting on the cages is significantly lower and the load distribution becomes load amplitude dependent, with a higher proportion of the load transferred by the cages at high loads. The bPLIF cages tend to absorb more force than the other two cage configurations.
CLINICAL SIGNIFICANCE
CONCLUSIONS
Cage instrumentation allows some of the compression force to be transmitted through the cage to the screws below, better distributing and reducing the overall force on the pedicle screws at the end of the construct and on the rods.
Identifiants
pubmed: 37924848
pii: S1529-9430(23)03485-X
doi: 10.1016/j.spinee.2023.10.017
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Informations de copyright
Copyright © 2023 The Author(s). Published by Elsevier Inc. All rights reserved.
Déclaration de conflit d'intérêts
Declaration of Competing Interest MF reports being a Consultant for Incremed (Balgrist University Startup), Zimmer Biomet, Medacta, and 25 Segments (Balgrist Startup). All the other authors report no conflicts of interest.