Sensing of concealed ventricular extrasystoles by pacemakers. Fifty years later.
concealed ventricular extrasystole
defective pacemaker lead
pacemaker
pacemaker oversensing
ventricular electrogram
Journal
Journal of cardiovascular electrophysiology
ISSN: 1540-8167
Titre abrégé: J Cardiovasc Electrophysiol
Pays: United States
ID NLM: 9010756
Informations de publication
Date de publication:
Dec 2023
Dec 2023
Historique:
revised:
17
10
2023
received:
30
06
2023
accepted:
24
10
2023
medline:
11
12
2023
pubmed:
15
11
2023
entrez:
15
11
2023
Statut:
ppublish
Résumé
Whether a pacemaker can sense concealed ventricular extrasystoles still remains debatable since its occurrence was first proposed in 1972. It must remain a diagnosis of exclusion if it really exists. Isoelectric complexes and all the causes of oversensing especially discrete false signals generated by a defective pacemaker lead must be excluded before concealed ventricular extrasystoles can be postulated.
Types de publication
Journal Article
Review
Langues
eng
Sous-ensembles de citation
IM
Pagination
2607-2612Informations de copyright
© 2023 Wiley Periodicals LLC.
Références
Platt SB, Fisher JD, Zilo P, Kim SG, Ferrigk KJ, Roth JA. Mutually independent atrial, junctional, and ventricular rhythms following radiofrequency ablation in a patient with postoperative junctional ectopic tachycardia. Pacing Clin Electrophysiol. 1994;17:1306-1310.
Khan A, Ameen A, Maraj I, Turitto G, El-Sherif NE. A His bundle extrasystole can both induce and reverse 2:1 atrioventricular block. J Interv Card Electrophysiol. 2011;32:121-123.
Mossami RA, Mason DT, Amsterdam EA, Salel AF. Apparent malfunction of demand pacemaker caused by nonpropagated (concealed) ventricular extrasystoles. Chest. 1972;61:426-431.
Levine PA, Pirzada FA. Pacemaker oversensing: a possible example of concealed ventricular extrasystoles. Pacing Clin Electrophysiol. 1981;4:199-203.
Barold SS. Inapparent signals to demand pacemakers. Chest. 1973;63:467-468.
Wiehmeyer J. Korrektur einer T-wellen induzierten Schrittmacherbradykardie durch Chinidin. DMW - Deutsche Medizinische Wochenschrift. 1975;100:1172-1176.
Barold SS, Falkoff MD, Ong LS, Heinle RA. Oversensing by single-chamber pacemakers: mechanisms, diagnosis, and treatment. Cardiol Clin. 1985;3:565-585.
Scher DL, Man DC, Edery T, Lu S. Isoelectric PVCs diagnosed with a monitor with simultaneous ECG, and ICD intracardiac electrogram and marker channels. Pacing Clin Electrophysiol. 2009;32:137-139.
Swerdlow CD, Ploux S, Poole JE, Nair SG, Himes A, Ellenbogen KA. Interpreting device diagnostics for lead failure. Heart Rhythm. 2022;19:154-164.
Swerdlow CD, Koneru JN, Gunderson B, Kroll MW, Ploux S, Ellenbogen KA. Impedance in the diagnosis of lead malfunction. Circ Arrhythm Electrophysiol. 2020;13(2):e008092.
Furman S, De Caprio V. Electrode causation of pacemaker inhibition. Chest. 1977;72:117-119.
Mugica J, Dubos M, Ducongé B, Lazarus B, Hertzog P. Arrhythmia in patients with demand pacemakers caused by faulty conduction material in the electrode. 36 cases-value of Holter monitoring. Arch Mal Coeur Vaiss. 1980;73:72-78 in French.
Bowers M, Berger M. Direct recording of concealed ventricular extrasystoles producing pacer inhibition. J Cardiovasc Electrophysiol. 2013;24:1179-1182.