Regulation (EU) 2017/746 (IVDR): practical implementation of annex I in pathology.

Die Verordnung (EU) 2017/746 (IVDR) in der Praxis: Umsetzung von Anhang I in der Pathologie. Englische Version.
In-house manufacturing Laboratory-developed tests Quality Assurance in health care Quality Management Regulatory requirements

Journal

Pathologie (Heidelberg, Germany)
ISSN: 2731-7196
Titre abrégé: Pathologie (Heidelb)
Pays: Germany
ID NLM: 9918384887506676

Informations de publication

Date de publication:
Nov 2023
Historique:
accepted: 17 10 2023
pubmed: 4 12 2023
medline: 4 12 2023
entrez: 4 12 2023
Statut: ppublish

Résumé

Regulation (EU) 2017/746 on in vitro diagnostic medical devices (IVDR) imposes several conditions on pathology departments that develop and use in-house in vitro diagnostic medical devices (IH-IVDs). However, not all of these conditions need to be implemented immediately after the IVDR entered into force on 26 May 2022. Based on an amending regulation of the European Parliament and the Council of the European Union, the requirements for IH-IVDs will be phased in. Conformity with the essential safety and performance requirements of annex I must be ensured from May 2022. With this article, we would like to present the practical implementation of the currently valid conditions for IH-IVDs at the Institute of Pathology at the University Hospital of Heidelberg, in order to provide possible assistance to other institutions. In addition to the intensive work on the requirements for IH-IVDs, several guidance documents and handouts provide orientation for the implementation and harmonisation of the requirements for healthcare institutions mentioned in Article 5 (5). Exchange in academic network structures is also of great importance for the interpretation and practical implementation of the IVDR. For university and nonuniversity institutions, ensuring conformity with the IVDR represents a further challenge in terms of personnel and time, in addition to the essential tasks of patient care, teaching and research and the further development of methods for optimal and targeted diagnostics, as well as the maintenance of the constantly evolving quality management system. HINTERGRUND: Die Verordnung (EU) 2017/746 über In-vitro-Diagnostika (IVDR) stellt mehrere Bedingungen an Pathologische Institute, die hausinterne In-vitro-Diagnostika (IH-IVD) entwickeln und anwenden. Diese Bedingungen müssen jedoch nicht alle unmittelbar mit dem Geltungsbeginn der IVDR zum 26.05.2022 umgesetzt worden sein. Auf der Grundlage einer Änderungsverordnung des Europäischen Parlaments und des Rates der Europäischen Union werden die Anforderungen an IH-IVD stufenweise eingeführt. Die Konformität mit den grundlegenden Sicherheits- und Leistungsanforderungen gemäß Anhang I muss seit Mai 2022 gewährleistet sein. Mit diesem Artikel möchten wir die praktische Umsetzung der aktuell gültigen Bedingungen für IH-IVD im Pathologischen Institut des Universitätsklinikums Heidelberg vorstellen und damit mögliche Hilfestellung für andere Einrichtungen geben. Neben der intensiven Auseinandersetzung mit den Anforderungen an IH-IVD geben mehrere Handreichungen und Hilfestellungen eine Orientierungshilfe zur Umsetzung und Harmonisierung der in Artikel 5 (5) genannten Anforderungen an Gesundheitseinrichtungen. Auch der Austausch in akademischen Netzwerkstrukturen ist für die Interpretation und die praktische Umsetzung der IVDR von großer Bedeutung. Für universitäre und nicht-universitäre Einrichtungen stellt die Sicherstellung der IVDR-Konformität – neben den wesentlichen Aufgaben in der Krankenversorgung, in der Lehre und der Forschung und Weiterentwicklung von Methoden zur optimalen und zielgerichteten Diagnostik sowie der Aufrechterhaltung des sich stetig weiterentwickelnden Qualitätsmanagementsystems  – eine weitere personelle und zeitliche Herausforderung dar.

Sections du résumé

BACKGROUND BACKGROUND
Regulation (EU) 2017/746 on in vitro diagnostic medical devices (IVDR) imposes several conditions on pathology departments that develop and use in-house in vitro diagnostic medical devices (IH-IVDs). However, not all of these conditions need to be implemented immediately after the IVDR entered into force on 26 May 2022. Based on an amending regulation of the European Parliament and the Council of the European Union, the requirements for IH-IVDs will be phased in. Conformity with the essential safety and performance requirements of annex I must be ensured from May 2022.
OBJECTIVES OBJECTIVE
With this article, we would like to present the practical implementation of the currently valid conditions for IH-IVDs at the Institute of Pathology at the University Hospital of Heidelberg, in order to provide possible assistance to other institutions.
CONCLUSIONS CONCLUSIONS
In addition to the intensive work on the requirements for IH-IVDs, several guidance documents and handouts provide orientation for the implementation and harmonisation of the requirements for healthcare institutions mentioned in Article 5 (5). Exchange in academic network structures is also of great importance for the interpretation and practical implementation of the IVDR. For university and nonuniversity institutions, ensuring conformity with the IVDR represents a further challenge in terms of personnel and time, in addition to the essential tasks of patient care, teaching and research and the further development of methods for optimal and targeted diagnostics, as well as the maintenance of the constantly evolving quality management system.
ZUSAMMENFASSUNG UNASSIGNED
HINTERGRUND: Die Verordnung (EU) 2017/746 über In-vitro-Diagnostika (IVDR) stellt mehrere Bedingungen an Pathologische Institute, die hausinterne In-vitro-Diagnostika (IH-IVD) entwickeln und anwenden. Diese Bedingungen müssen jedoch nicht alle unmittelbar mit dem Geltungsbeginn der IVDR zum 26.05.2022 umgesetzt worden sein. Auf der Grundlage einer Änderungsverordnung des Europäischen Parlaments und des Rates der Europäischen Union werden die Anforderungen an IH-IVD stufenweise eingeführt. Die Konformität mit den grundlegenden Sicherheits- und Leistungsanforderungen gemäß Anhang I muss seit Mai 2022 gewährleistet sein.
ZIEL DER ARBEIT UNASSIGNED
Mit diesem Artikel möchten wir die praktische Umsetzung der aktuell gültigen Bedingungen für IH-IVD im Pathologischen Institut des Universitätsklinikums Heidelberg vorstellen und damit mögliche Hilfestellung für andere Einrichtungen geben.
SCHLUSSFOLGERUNGEN UNASSIGNED
Neben der intensiven Auseinandersetzung mit den Anforderungen an IH-IVD geben mehrere Handreichungen und Hilfestellungen eine Orientierungshilfe zur Umsetzung und Harmonisierung der in Artikel 5 (5) genannten Anforderungen an Gesundheitseinrichtungen. Auch der Austausch in akademischen Netzwerkstrukturen ist für die Interpretation und die praktische Umsetzung der IVDR von großer Bedeutung. Für universitäre und nicht-universitäre Einrichtungen stellt die Sicherstellung der IVDR-Konformität – neben den wesentlichen Aufgaben in der Krankenversorgung, in der Lehre und der Forschung und Weiterentwicklung von Methoden zur optimalen und zielgerichteten Diagnostik sowie der Aufrechterhaltung des sich stetig weiterentwickelnden Qualitätsmanagementsystems  – eine weitere personelle und zeitliche Herausforderung dar.

Autres résumés

Type: Publisher (ger)
HINTERGRUND: Die Verordnung (EU) 2017/746 über In-vitro-Diagnostika (IVDR) stellt mehrere Bedingungen an Pathologische Institute, die hausinterne In-vitro-Diagnostika (IH-IVD) entwickeln und anwenden. Diese Bedingungen müssen jedoch nicht alle unmittelbar mit dem Geltungsbeginn der IVDR zum 26.05.2022 umgesetzt worden sein. Auf der Grundlage einer Änderungsverordnung des Europäischen Parlaments und des Rates der Europäischen Union werden die Anforderungen an IH-IVD stufenweise eingeführt. Die Konformität mit den grundlegenden Sicherheits- und Leistungsanforderungen gemäß Anhang I muss seit Mai 2022 gewährleistet sein.

Identifiants

pubmed: 38047950
doi: 10.1007/s00292-023-01274-6
pii: 10.1007/s00292-023-01274-6
pmc: PMC10713655
doi:

Types de publication

Journal Article Review

Langues

eng

Sous-ensembles de citation

IM

Pagination

86-95

Informations de copyright

© 2023. The Author(s).

Références

Pathologe. 2020 Dec;41(Suppl 2):129-133
pubmed: 33263810
Pathologie (Heidelb). 2022 Sep;43(5):351-364
pubmed: 35925306
Pathologie (Heidelb). 2022 Sep;43(5):338-345
pubmed: 35925317

Auteurs

Andy Kahles (A)

Institute of Pathology, University Hospital Heidelberg, Im Neuenheimer Feld 224, 69120, Heidelberg, Germany. andy.kahles@med.uni-heidelberg.de.

Hannah Goldschmid (H)

Institute of Pathology, University Hospital Heidelberg, Im Neuenheimer Feld 224, 69120, Heidelberg, Germany.

Anna-Lena Volckmar (AL)

Institute of Pathology, University Hospital Heidelberg, Im Neuenheimer Feld 224, 69120, Heidelberg, Germany.

Carolin Ploeger (C)

Institute of Pathology, University Hospital Heidelberg, Im Neuenheimer Feld 224, 69120, Heidelberg, Germany.

Daniel Kazdal (D)

Institute of Pathology, University Hospital Heidelberg, Im Neuenheimer Feld 224, 69120, Heidelberg, Germany.

Roland Penzel (R)

Institute of Pathology, University Hospital Heidelberg, Im Neuenheimer Feld 224, 69120, Heidelberg, Germany.

Jan Budczies (J)

Institute of Pathology, University Hospital Heidelberg, Im Neuenheimer Feld 224, 69120, Heidelberg, Germany.

Christa Flechtenmacher (C)

Institute of Pathology, University Hospital Heidelberg, Im Neuenheimer Feld 224, 69120, Heidelberg, Germany.

Ulrich M Gassner (UM)

Faculty of Law, University of Augsburg, Augsburg, Germany.

Monika Brüggemann (M)

2nd Internal Medicine Department, Hematology Lab Kiel, University Hospital Schleswig-Holstein (UKSH), Kiel, Germany.

Michael Vogeser (M)

Institute of Laboratory Medicine, University Hospital, LMU Munich, Munich, Germany.

Peter Schirmacher (P)

Institute of Pathology, University Hospital Heidelberg, Im Neuenheimer Feld 224, 69120, Heidelberg, Germany.

Albrecht Stenzinger (A)

Institute of Pathology, University Hospital Heidelberg, Im Neuenheimer Feld 224, 69120, Heidelberg, Germany. albrecht.stenzinger@med.uni-heidelberg.de.

Classifications MeSH