Evaluation of 2D ion chamber arrays for patient specific quality assurance using a static phantom at a 0.35 T MR-Linac.

MR-Linac ion chamber array patient specific quality assurance

Journal

Zeitschrift fur medizinische Physik
ISSN: 1876-4436
Titre abrégé: Z Med Phys
Pays: Germany
ID NLM: 100886455

Informations de publication

Date de publication:
05 Jan 2024
Historique:
received: 27 02 2023
revised: 31 10 2023
accepted: 18 12 2023
medline: 7 1 2024
pubmed: 7 1 2024
entrez: 6 1 2024
Statut: aheadofprint

Résumé

Patient specific quality assurance (QA) in MR-Linacs can be performed with MR-compatible ion chamber arrays. However, the presence of a static magnetic field can alter the angular response of such arrays substantially. This works investigates the suitability of two ion chamber arrays, an air-filled and a liquid-filled array, for patient specific QA at a 0.35 T MR-Linac using a static phantom. In order to study the angular response, the two arrays were placed in a static, solid phantom and irradiated with 9.96 × 9.96 cm The air-filled array showed asymmetric angular response changes of central chamber dose of up to 18% and down to local 3 mm / 3% gamma rates of 20%, while only minor differences within 3% (excluding parallel irradiation and beams through the couch edges) were found for the liquid-filled ion chamber array without rotating the phantom. Patient plan QA using the liquid-filled array yielded a median local 3 mm / 3% 3D gamma passing rate of 99.8% (range 96.9%-100%). A liquid-filled ionization chamber array in combination with a static phantom can be used for efficient patient specific plan QA in a single measurement set-up in a 0.35 T MR-Linac, while the air-filled ion chamber array phantom shows large angular response changes and has its limitations regarding patient specific QA measurements.

Identifiants

pubmed: 38184375
pii: S0939-3889(23)00150-2
doi: 10.1016/j.zemedi.2023.12.003
pii:
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Informations de copyright

Copyright © 2023 The Author(s). Published by Elsevier GmbH.. All rights reserved.

Auteurs

C K Renkamp (CK)

Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany; Heidelberg Institute of Radiation Oncology (HIRO), National Center for Radiation Oncology (NCRO), Heidelberg, Germany. Electronic address: katharina.renkamp@med.uni-heidelberg.de.

D Eulenstein (D)

PTW Freiburg GmbH, Freiburg, Germany.

M Sebald (M)

Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany; Heidelberg Institute of Radiation Oncology (HIRO), National Center for Radiation Oncology (NCRO), Heidelberg, Germany.

F Schlüter (F)

Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany; Heidelberg Institute of Radiation Oncology (HIRO), National Center for Radiation Oncology (NCRO), Heidelberg, Germany.

C Buchele (C)

Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany; Heidelberg Institute of Radiation Oncology (HIRO), National Center for Radiation Oncology (NCRO), Heidelberg, Germany; Medical Faculty, University of Heidelberg, Heidelberg, Germany.

C Rippke (C)

Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany; Heidelberg Institute of Radiation Oncology (HIRO), National Center for Radiation Oncology (NCRO), Heidelberg, Germany; Medical Faculty, University of Heidelberg, Heidelberg, Germany.

J Debus (J)

Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany; Heidelberg Institute of Radiation Oncology (HIRO), National Center for Radiation Oncology (NCRO), Heidelberg, Germany; Medical Faculty, University of Heidelberg, Heidelberg, Germany; Clinical Cooperation Unit Radiation Oncology, German Cancer Research Center (DKFZ), Heidelberg, Germany; National Center for Tumor diseases (NCT), Heidelberg, Germany; Heidelberg Ion-Beam Therapy Center (HIT), Heidelberg, Germany; German Cancer Consortium (DKTK), Core-Center Heidelberg, Heidelberg, Germany.

S Klüter (S)

Department of Radiation Oncology, Heidelberg University Hospital, Heidelberg, Germany; Heidelberg Institute of Radiation Oncology (HIRO), National Center for Radiation Oncology (NCRO), Heidelberg, Germany.

Classifications MeSH