Word Recognition with a Cochlear Implant in Relation to Prediction and Electrode Position.
WRS prediction
cochlear implant
digital volume tomography
insertion depth
lateral wall
word recognition score
Journal
Journal of clinical medicine
ISSN: 2077-0383
Titre abrégé: J Clin Med
Pays: Switzerland
ID NLM: 101606588
Informations de publication
Date de publication:
28 Dec 2023
28 Dec 2023
Historique:
received:
20
11
2023
revised:
11
12
2023
accepted:
21
12
2023
medline:
11
1
2024
pubmed:
11
1
2024
entrez:
11
1
2024
Statut:
epublish
Résumé
the word recognition score (WRS) achieved with cochlear implants (CIs) varies widely. To account for this, a predictive model was developed based on patients' age and their pre-operative WRS. This retrospective study aimed to find out whether the insertion depth of the nucleus lateral-wall electrode arrays contributes to the deviation of the CI-achieved WRS from the predicted WRS. patients with a pre-operative maximum WRS > 0 or a pure-tone audiogram ≥80 dB were included. The insertion depth was determined via digital volume tomography. fifty-three patients met the inclusion criteria. The median WRS achieved with the CI was 70%. The comparison of pre- and post-operative scores achieved with a hearing aid and a CI respectively in the aided condition showed a median improvement of 65 percentage points (pp). A total of 90% of the patients improved by at least 20 pp. The majority of patients reached or exceeded the prediction, with a median absolute error of 11 pp. No significant correlation was found between the deviation from the predicted WRS and the insertion depth. our data support a previously published model for the prediction of the WRS after cochlear implantation. For the lateral-wall electrode arrays evaluated, the insertion depth did not influence the WRS with a CI.
Sections du résumé
BACKGROUND
BACKGROUND
the word recognition score (WRS) achieved with cochlear implants (CIs) varies widely. To account for this, a predictive model was developed based on patients' age and their pre-operative WRS. This retrospective study aimed to find out whether the insertion depth of the nucleus lateral-wall electrode arrays contributes to the deviation of the CI-achieved WRS from the predicted WRS.
MATERIALS AND METHODS
METHODS
patients with a pre-operative maximum WRS > 0 or a pure-tone audiogram ≥80 dB were included. The insertion depth was determined via digital volume tomography.
RESULTS
RESULTS
fifty-three patients met the inclusion criteria. The median WRS achieved with the CI was 70%. The comparison of pre- and post-operative scores achieved with a hearing aid and a CI respectively in the aided condition showed a median improvement of 65 percentage points (pp). A total of 90% of the patients improved by at least 20 pp. The majority of patients reached or exceeded the prediction, with a median absolute error of 11 pp. No significant correlation was found between the deviation from the predicted WRS and the insertion depth.
CONCLUSIONS
CONCLUSIONS
our data support a previously published model for the prediction of the WRS after cochlear implantation. For the lateral-wall electrode arrays evaluated, the insertion depth did not influence the WRS with a CI.
Identifiants
pubmed: 38202190
pii: jcm13010183
doi: 10.3390/jcm13010183
pii:
doi:
Types de publication
Journal Article
Langues
eng
Subventions
Organisme : Cochlear Research and Development Limited
ID : IIR 2341