Validation of a deep-learning segmentation model for adult and pediatric head and neck radiotherapy in different patient positions.
Artificial Intelligence
Auto-segmentation
Autocontouring
Head and neck cancer
Organs-at-risk
Radiation therapy
Journal
Physics and imaging in radiation oncology
ISSN: 2405-6316
Titre abrégé: Phys Imaging Radiat Oncol
Pays: Netherlands
ID NLM: 101704276
Informations de publication
Date de publication:
Jan 2024
Jan 2024
Historique:
received:
10
05
2023
revised:
15
12
2023
accepted:
18
12
2023
medline:
15
1
2024
pubmed:
15
1
2024
entrez:
15
1
2024
Statut:
epublish
Résumé
Autocontouring for radiotherapy has the potential to significantly save time and reduce interobserver variability. We aimed to assess the performance of a commercial autocontouring model for head and neck (H&N) patients in eight orientations relevant to particle therapy with fixed beam lines, focusing on validation and implementation for routine clinical use. Autocontouring was performed on sixteen organs at risk (OARs) for 98 adult and pediatric patients with 137 H&N CT scans in eight orientations. A geometric comparison of the autocontours and manual segmentations was performed using the Hausdorff Distance 95th percentile, Dice Similarity Coefficient (DSC) and surface DSC and compared to interobserver variability where available. Additional qualitative scoring and dose-volume-histogram (DVH) parameters analyses were performed for twenty patients in two positions, consisting of scoring on a 0-3 scale based on clinical usability and comparing the mean (D For the geometric analysis, the model performance in head-first-supine straight and hyperextended orientations was in the same range as the interobserver variability. HD95, DSC and surface DSC was heterogeneous in other orientations. No significant geometric differences were found between pediatric and adult autocontours. The qualitative scoring yielded a median score of ≥ 2 for 13/16 OARs while 7/32 DVH parameters were significantly different. For head-first-supine straight and hyperextended scans, we found that 13/16 OAR autocontours were suited for use in daily clinical practice and subsequently implemented. Further development is needed for other patient orientations before implementation.
Sections du résumé
Background and purpose
UNASSIGNED
Autocontouring for radiotherapy has the potential to significantly save time and reduce interobserver variability. We aimed to assess the performance of a commercial autocontouring model for head and neck (H&N) patients in eight orientations relevant to particle therapy with fixed beam lines, focusing on validation and implementation for routine clinical use.
Materials and methods
UNASSIGNED
Autocontouring was performed on sixteen organs at risk (OARs) for 98 adult and pediatric patients with 137 H&N CT scans in eight orientations. A geometric comparison of the autocontours and manual segmentations was performed using the Hausdorff Distance 95th percentile, Dice Similarity Coefficient (DSC) and surface DSC and compared to interobserver variability where available. Additional qualitative scoring and dose-volume-histogram (DVH) parameters analyses were performed for twenty patients in two positions, consisting of scoring on a 0-3 scale based on clinical usability and comparing the mean (D
Results
UNASSIGNED
For the geometric analysis, the model performance in head-first-supine straight and hyperextended orientations was in the same range as the interobserver variability. HD95, DSC and surface DSC was heterogeneous in other orientations. No significant geometric differences were found between pediatric and adult autocontours. The qualitative scoring yielded a median score of ≥ 2 for 13/16 OARs while 7/32 DVH parameters were significantly different.
Conclusions
UNASSIGNED
For head-first-supine straight and hyperextended scans, we found that 13/16 OAR autocontours were suited for use in daily clinical practice and subsequently implemented. Further development is needed for other patient orientations before implementation.
Identifiants
pubmed: 38222671
doi: 10.1016/j.phro.2023.100527
pii: S2405-6316(23)00118-5
pmc: PMC10787237
doi:
Types de publication
Journal Article
Langues
eng
Pagination
100527Informations de copyright
© 2023 The Author(s).
Déclaration de conflit d'intérêts
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.