Scintillation and Cherenkov photon counting detectors with analog silicon photomultipliers for TOF-PET.
Analog SiPMs
Photon Counting Detector
TOF-PET
Journal
Physics in medicine and biology
ISSN: 1361-6560
Titre abrégé: Phys Med Biol
Pays: England
ID NLM: 0401220
Informations de publication
Date de publication:
22 Jan 2024
22 Jan 2024
Historique:
medline:
22
1
2024
pubmed:
22
1
2024
entrez:
22
1
2024
Statut:
aheadofprint
Résumé
Standard signal processing approaches for scintillation detectors in positron emission tomography (PET) derive accurate estimates for 511 keV photon time of interaction and energy imparted to the detection media from aggregate characteristics of electronic pulse shapes. The ultimate realization of a scintillation detector for PET is one that provides a unique timestamp and position for each detected scintillation photon. Detectors with these capabilities enable advanced concepts for three-dimensional (3D) position and time of interaction estimation with methods that exploit the spatiotemporal arrival time kinetics of individual scintillation photons. 

In this work, we show that taking into consideration the temporal photon emission density of a scintillator, the channel density of an analog silicon photomultiplier (SiPM) array, and employing fast electronic readout with digital signal processing, a detector that counts and timestamps scintillation photons can be realized. To demonstrate this approach, a prototype detector was constructed, comprising multichannel electronic readout for a bismuth germanate (BGO) scintillator coupled to an SiPM array.

In proof-of-concept measurements, we were able to count and provide unique timestamps for 66% of all optical photons, where the remaining 34% (two-or-more-photon pulses) are also independently counted, but each photon bunch shares a common timestamp. We show this detector concept can implement 3D positioning of 511 keV photon interactions and thereby enable corrections for time of interaction estimators. The detector achieved 17.6% energy resolution at 511 keV and 237±10 ps full-width-at-half-maximum (FWHM) coincidence time resolution (CTR) (fast spectral component) versus a reference detector. We outline the methodology, readout, and approach for achieving this detector capability in first-ever, proof-of-concept measurements for scintillation photon counting detector with analog silicon photomultipliers.

The presented detector concept is a promising design for large area, high sensitivity TOF-PET detector modules that can implement advanced event positioning and time of interaction estimators, which could push state-of-the-art performance.