Improved Signal Processing for Mass Shifting Ions in Charge Detection Mass Spectrometry.
Journal
Journal of the American Society for Mass Spectrometry
ISSN: 1879-1123
Titre abrégé: J Am Soc Mass Spectrom
Pays: United States
ID NLM: 9010412
Informations de publication
Date de publication:
29 Jan 2024
29 Jan 2024
Historique:
medline:
30
1
2024
pubmed:
30
1
2024
entrez:
29
1
2024
Statut:
aheadofprint
Résumé
The quality of data in charge detection mass spectrometry depends on accurate determination of ion charge. While the method of selective temporal overview of resonant ions (STORI) has proven to be highly enabling for determining the charge of ions that survive for variable amounts of time, it assumes that the ion frequency exactly matches the frequency being used in the calculation. Any mismatches result in low charge estimates. To address this, the misSTORI method was developed to correct these discrepancies. This can significantly reduce the charge measurement errors for samples with unstable masses. As an example, the misSTORI approach can eliminate a 5.7% charge determination error for a VP3-only AAV capsid that shifts 25 ppm in mass.
Identifiants
pubmed: 38286823
doi: 10.1021/jasms.3c00435
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM