Occlusion Pressures of Tactical Pneumatic Tourniquet 2".

emergency treatment first aid hemorrhage tourniquet

Journal

Journal of special operations medicine : a peer reviewed journal for SOF medical professionals
ISSN: 1553-9768
Titre abrégé: J Spec Oper Med
Pays: United States
ID NLM: 101158402

Informations de publication

Date de publication:
01 Feb 2024
Historique:
medline: 1 2 2024
pubmed: 1 2 2024
entrez: 1 2 2024
Statut: aheadofprint

Résumé

The Tactical Pneumatic Tournqiuet 2" (TPT2, 5.1cm-wide deflated) allows total average applied pressure measurement, which should be useful toward development of emergency-use limb tourniquet certification devices. The TPT2 hand bulb was replaced with stopcocks and syringes, allowing filling with continuous pressure measurement. Forearm and mid-thigh applications involved two sets of five Doppler-based pulse gone/return pairs. Second set pulse gones were chosen a priori for occlusion pressures (preliminary work indicated greater consistency in second sets). All 68 forearms occluded (30 female, 38 male, median circumference 17.8cm, range 14.6-23.5cm; median second set of pulse gone tourniquet pressures 176mmHg, range 128-282mmHg). Fifty-five thighs occluded (median circumference 54.3cm, range 41.6-62.4cm; median systolic pressure 126mmHg, range 102-142mmHg; median second set of pulse gone pressures 574mmHg, range 274-1158mmHg). Thirteen thigh applications were stopped without occlusion because of concerning pressures combined with no indication of imminent occlusion and difficulties forcing more air into the TPT2 (3 female, 10 male, peak pressures from 958-1377mmHg, median 1220mmHg, p<.0001 versus occluded thighs; median circumference 63.3cm, range 55.0-72.9cm, p<.0001 versus occluded thighs; median systolic pressure 126mmHg, range 120-173mmHg, p<.019 versus occluded thighs). Thigh TPT2 impression widths on five subjects after occlusion were as follows: 3.5cm, occlusion 274mmHg; 2.8cm, occlusion 348mmHg; 2.9cm, occlusion 500mmHg; 2.8cm, occlusion 782mmHg; 2.7cm, occlusion 1114mmHg. Though probably useful to tourniquet certification, the required pressures for thigh occlusion make the TPT2 undesirable for any clinical use, emergency or otherwise.

Sections du résumé

BACKGROUND BACKGROUND
The Tactical Pneumatic Tournqiuet 2" (TPT2, 5.1cm-wide deflated) allows total average applied pressure measurement, which should be useful toward development of emergency-use limb tourniquet certification devices.
METHODS METHODS
The TPT2 hand bulb was replaced with stopcocks and syringes, allowing filling with continuous pressure measurement. Forearm and mid-thigh applications involved two sets of five Doppler-based pulse gone/return pairs. Second set pulse gones were chosen a priori for occlusion pressures (preliminary work indicated greater consistency in second sets).
RESULTS RESULTS
All 68 forearms occluded (30 female, 38 male, median circumference 17.8cm, range 14.6-23.5cm; median second set of pulse gone tourniquet pressures 176mmHg, range 128-282mmHg). Fifty-five thighs occluded (median circumference 54.3cm, range 41.6-62.4cm; median systolic pressure 126mmHg, range 102-142mmHg; median second set of pulse gone pressures 574mmHg, range 274-1158mmHg). Thirteen thigh applications were stopped without occlusion because of concerning pressures combined with no indication of imminent occlusion and difficulties forcing more air into the TPT2 (3 female, 10 male, peak pressures from 958-1377mmHg, median 1220mmHg, p<.0001 versus occluded thighs; median circumference 63.3cm, range 55.0-72.9cm, p<.0001 versus occluded thighs; median systolic pressure 126mmHg, range 120-173mmHg, p<.019 versus occluded thighs). Thigh TPT2 impression widths on five subjects after occlusion were as follows: 3.5cm, occlusion 274mmHg; 2.8cm, occlusion 348mmHg; 2.9cm, occlusion 500mmHg; 2.8cm, occlusion 782mmHg; 2.7cm, occlusion 1114mmHg.
CONCLUSIONS CONCLUSIONS
Though probably useful to tourniquet certification, the required pressures for thigh occlusion make the TPT2 undesirable for any clinical use, emergency or otherwise.