A readout system for highly sensitive diamond detectors for FLASH dosimetry.
Beam monitoring
Diamond detectors
Electron FLASH
Front-end/readout electronics
Synchronous detection
Ultra-high dose-per-pulse
Journal
Physics and imaging in radiation oncology
ISSN: 2405-6316
Titre abrégé: Phys Imaging Radiat Oncol
Pays: Netherlands
ID NLM: 101704276
Informations de publication
Date de publication:
Jan 2024
Jan 2024
Historique:
received:
13
09
2023
revised:
13
01
2024
accepted:
15
01
2024
medline:
6
2
2024
pubmed:
6
2
2024
entrez:
6
2
2024
Statut:
epublish
Résumé
Accurate dosimetry of ultra-high dose-rate beams using diamond detectors remains challenging, primarily due to the elevated photocurrent peaks exceeding the input dynamics of precision electrometers. This work aimed at demonstrating the effectiveness of compact gated-integration electronics in conditioning the current peaks (>20 mA) generated by a highly sensitive (
Identifiants
pubmed: 38317851
doi: 10.1016/j.phro.2024.100538
pii: S2405-6316(24)00008-3
pmc: PMC10839766
doi:
Types de publication
Journal Article
Langues
eng
Pagination
100538Informations de copyright
© 2024 The Author(s).
Déclaration de conflit d'intérêts
The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: Giuseppe Felici is a SIT S.p.A. shareholder. Lorenzo Galluzzo is an employee at SIT S.p.A.