Evaluation of Busulfan as a Third-Party Immunoassay on a Clinical Chemistry Analyzer.


Journal

The journal of applied laboratory medicine
ISSN: 2576-9456
Titre abrégé: J Appl Lab Med
Pays: England
ID NLM: 101693884

Informations de publication

Date de publication:
09 Feb 2024
Historique:
received: 20 07 2023
accepted: 28 11 2023
medline: 9 2 2024
pubmed: 9 2 2024
entrez: 9 2 2024
Statut: aheadofprint

Résumé

Busulfan is widely used in conditioning regimens to prepare patients for hematopoietic stem cell transplantation. Therapeutic drug monitoring (TDM) is critical due to large inter- and intra-individual variability in busulfan pharmacokinetics, and the risk of adverse consequences of toxicity including hepatic veno-occlusive disease. Busulfan is most commonly measured by liquid chromatography-mass spectrometry (LC-MS/MS), which is not as widely available in clinical laboratories as automated routine clinical chemistry analyzers. The objective was to perform analytical verification of a busulfan immunoassay on the Abbott Alinity c platform. The MyCare Oncology busulfan immunoassay was configured as a third-party reagent on the Abbott Alinity c. Imprecision, linearity, sample carryover, and onboard stability of reagent studies were evaluated. The performance of the busulfan immunoassay using the Abbott Alinity c was compared to the Beckman Coulter AU480 using sodium heparinized plasma, as well as to LC-MS/MS using lithium heparinized plasma. The imprecision goal of 8% was met, and linearity within the analytical measurement range of 240 to 1700 ng/mL was verified. Sample carryover was negligible, and the reagents were stable onboard for at least 84 days. The busulfan immunoassay correlated well with LC-MS/MS (slope = 0.949, y-intercept = -7.8 ng/mL, r2 = 0.9935) and the Beckman Coulter AU480 (slope = 1.090, y-intercept = -34.5 ng/mL, r2 = 0.9988). This study demonstrated successful analytical verification of a busulfan third-party immunoassay on the Abbott Alinity c platform. The ability to perform TDM of busulfan on a routine clinical chemistry analyzer will positively impact turnaround times to improve patient outcomes.

Sections du résumé

BACKGROUND BACKGROUND
Busulfan is widely used in conditioning regimens to prepare patients for hematopoietic stem cell transplantation. Therapeutic drug monitoring (TDM) is critical due to large inter- and intra-individual variability in busulfan pharmacokinetics, and the risk of adverse consequences of toxicity including hepatic veno-occlusive disease. Busulfan is most commonly measured by liquid chromatography-mass spectrometry (LC-MS/MS), which is not as widely available in clinical laboratories as automated routine clinical chemistry analyzers. The objective was to perform analytical verification of a busulfan immunoassay on the Abbott Alinity c platform.
METHODS METHODS
The MyCare Oncology busulfan immunoassay was configured as a third-party reagent on the Abbott Alinity c. Imprecision, linearity, sample carryover, and onboard stability of reagent studies were evaluated. The performance of the busulfan immunoassay using the Abbott Alinity c was compared to the Beckman Coulter AU480 using sodium heparinized plasma, as well as to LC-MS/MS using lithium heparinized plasma.
RESULTS RESULTS
The imprecision goal of 8% was met, and linearity within the analytical measurement range of 240 to 1700 ng/mL was verified. Sample carryover was negligible, and the reagents were stable onboard for at least 84 days. The busulfan immunoassay correlated well with LC-MS/MS (slope = 0.949, y-intercept = -7.8 ng/mL, r2 = 0.9935) and the Beckman Coulter AU480 (slope = 1.090, y-intercept = -34.5 ng/mL, r2 = 0.9988).
CONCLUSIONS CONCLUSIONS
This study demonstrated successful analytical verification of a busulfan third-party immunoassay on the Abbott Alinity c platform. The ability to perform TDM of busulfan on a routine clinical chemistry analyzer will positively impact turnaround times to improve patient outcomes.

Identifiants

pubmed: 38332638
pii: 7604025
doi: 10.1093/jalm/jfad134
pii:
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Subventions

Organisme : Busulfan MyCare Oncology

Informations de copyright

© Association for Diagnostics & Laboratory Medicine 2024. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.

Auteurs

Janet R Zhou (JR)

Department of Laboratory Medicine and Pathobiology, University of Toronto, Toronto, ON, Canada.

Andrea To (A)

Division of Clinical Biochemistry, University Health Network, Toronto, ON, Canada.

James Fritz (J)

Saladax Biomedical, Inc., Bethlehem, PA, United States.

Benjamin Jung (B)

Department of Laboratory Medicine and Pathobiology, University of Toronto, Toronto, ON, Canada.
Division of Clinical Biochemistry, Department of Paediatric Laboratory Medicine, Hospital for Sick Children, Toronto, ON, Canada.

Mehrdad Yazdanpanah (M)

Division of Clinical Biochemistry, Department of Paediatric Laboratory Medicine, Hospital for Sick Children, Toronto, ON, Canada.

Vathany Kulasingam (V)

Department of Laboratory Medicine and Pathobiology, University of Toronto, Toronto, ON, Canada.
Division of Clinical Biochemistry, University Health Network, Toronto, ON, Canada.

Classifications MeSH