Evaluation of Busulfan as a Third-Party Immunoassay on a Clinical Chemistry Analyzer.
Journal
The journal of applied laboratory medicine
ISSN: 2576-9456
Titre abrégé: J Appl Lab Med
Pays: England
ID NLM: 101693884
Informations de publication
Date de publication:
09 Feb 2024
09 Feb 2024
Historique:
received:
20
07
2023
accepted:
28
11
2023
medline:
9
2
2024
pubmed:
9
2
2024
entrez:
9
2
2024
Statut:
aheadofprint
Résumé
Busulfan is widely used in conditioning regimens to prepare patients for hematopoietic stem cell transplantation. Therapeutic drug monitoring (TDM) is critical due to large inter- and intra-individual variability in busulfan pharmacokinetics, and the risk of adverse consequences of toxicity including hepatic veno-occlusive disease. Busulfan is most commonly measured by liquid chromatography-mass spectrometry (LC-MS/MS), which is not as widely available in clinical laboratories as automated routine clinical chemistry analyzers. The objective was to perform analytical verification of a busulfan immunoassay on the Abbott Alinity c platform. The MyCare Oncology busulfan immunoassay was configured as a third-party reagent on the Abbott Alinity c. Imprecision, linearity, sample carryover, and onboard stability of reagent studies were evaluated. The performance of the busulfan immunoassay using the Abbott Alinity c was compared to the Beckman Coulter AU480 using sodium heparinized plasma, as well as to LC-MS/MS using lithium heparinized plasma. The imprecision goal of 8% was met, and linearity within the analytical measurement range of 240 to 1700 ng/mL was verified. Sample carryover was negligible, and the reagents were stable onboard for at least 84 days. The busulfan immunoassay correlated well with LC-MS/MS (slope = 0.949, y-intercept = -7.8 ng/mL, r2 = 0.9935) and the Beckman Coulter AU480 (slope = 1.090, y-intercept = -34.5 ng/mL, r2 = 0.9988). This study demonstrated successful analytical verification of a busulfan third-party immunoassay on the Abbott Alinity c platform. The ability to perform TDM of busulfan on a routine clinical chemistry analyzer will positively impact turnaround times to improve patient outcomes.
Sections du résumé
BACKGROUND
BACKGROUND
Busulfan is widely used in conditioning regimens to prepare patients for hematopoietic stem cell transplantation. Therapeutic drug monitoring (TDM) is critical due to large inter- and intra-individual variability in busulfan pharmacokinetics, and the risk of adverse consequences of toxicity including hepatic veno-occlusive disease. Busulfan is most commonly measured by liquid chromatography-mass spectrometry (LC-MS/MS), which is not as widely available in clinical laboratories as automated routine clinical chemistry analyzers. The objective was to perform analytical verification of a busulfan immunoassay on the Abbott Alinity c platform.
METHODS
METHODS
The MyCare Oncology busulfan immunoassay was configured as a third-party reagent on the Abbott Alinity c. Imprecision, linearity, sample carryover, and onboard stability of reagent studies were evaluated. The performance of the busulfan immunoassay using the Abbott Alinity c was compared to the Beckman Coulter AU480 using sodium heparinized plasma, as well as to LC-MS/MS using lithium heparinized plasma.
RESULTS
RESULTS
The imprecision goal of 8% was met, and linearity within the analytical measurement range of 240 to 1700 ng/mL was verified. Sample carryover was negligible, and the reagents were stable onboard for at least 84 days. The busulfan immunoassay correlated well with LC-MS/MS (slope = 0.949, y-intercept = -7.8 ng/mL, r2 = 0.9935) and the Beckman Coulter AU480 (slope = 1.090, y-intercept = -34.5 ng/mL, r2 = 0.9988).
CONCLUSIONS
CONCLUSIONS
This study demonstrated successful analytical verification of a busulfan third-party immunoassay on the Abbott Alinity c platform. The ability to perform TDM of busulfan on a routine clinical chemistry analyzer will positively impact turnaround times to improve patient outcomes.
Identifiants
pubmed: 38332638
pii: 7604025
doi: 10.1093/jalm/jfad134
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Subventions
Organisme : Busulfan MyCare Oncology
Informations de copyright
© Association for Diagnostics & Laboratory Medicine 2024. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.