Late cardiac tamponade after a helix-based active fixation leadless pacemaker implantation.

cardiac tamponade helix‐based active fixation leadless pacemaker leadless pacemakerAveir‐VR

Journal

Journal of arrhythmia
ISSN: 1880-4276
Titre abrégé: J Arrhythm
Pays: Japan
ID NLM: 101263026

Informations de publication

Date de publication:
Feb 2024
Historique:
received: 06 10 2023
revised: 17 12 2023
accepted: 21 12 2023
medline: 9 2 2024
pubmed: 9 2 2024
entrez: 9 2 2024
Statut: epublish

Résumé

Although the late cardiac tamponade in leadless pacemaker implantation (LPI) is rare, we encountered such an incident in patient with AVEIR-VR™ system on hemodialysis and warfarinization. When LPI with active fixation system, we should aim for successful single-attempt deployment using electrical premapping to prevent cardiac tamponade including the late phase.

Identifiants

pubmed: 38333402
doi: 10.1002/joa3.12987
pii: JOA312987
pmc: PMC10848613
doi:

Types de publication

Journal Article

Langues

eng

Pagination

177-179

Informations de copyright

© 2024 The Authors. Journal of Arrhythmia published by John Wiley & Sons Australia, Ltd on behalf of Japanese Heart Rhythm Society.

Déclaration de conflit d'intérêts

Authors declare no conflict of interests for this article.

Auteurs

Ryuki Chatani (R)

Department of Cardiovascular Medicine Kurashiki Central Hospital Kurashiki Japan.

Hiroshi Tasaka (H)

Department of Cardiovascular Medicine Kurashiki Central Hospital Kurashiki Japan.

Kenta Yoshida (K)

Department of Cardiovascular Medicine Kurashiki Central Hospital Kurashiki Japan.

Mitsuru Yoshino (M)

Department of Cardiovascular Medicine Kurashiki Central Hospital Kurashiki Japan.

Kazushige Kadota (K)

Department of Cardiovascular Medicine Kurashiki Central Hospital Kurashiki Japan.

Classifications MeSH