Temporal relationship between haemodynamic changes and activation of closed-loop stimulation during a tilt-induced vasovagal syncope.
Close-loop stimulation
Pacemaker therapy
Tilt table test
Vasovagal syncope
Journal
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
ISSN: 1532-2092
Titre abrégé: Europace
Pays: England
ID NLM: 100883649
Informations de publication
Date de publication:
01 Feb 2024
01 Feb 2024
Historique:
received:
08
12
2023
accepted:
06
02
2024
medline:
26
2
2024
pubmed:
10
2
2024
entrez:
10
2
2024
Statut:
ppublish
Résumé
A dual-chamber pacemaker with closed-loop stimulation (CLS) mode is effective in reducing syncopal recurrences in patients with asystolic vasovagal syncope (VVS). In this study, we explored the haemodynamic and temporal relationship of CLS during a tilt-induced vasovagal reflex. Twenty patients underwent a tilt test under video recording 3.9 years after CLS pacemaker implantation. Three patients were excluded from the analysis because of no VVS induced by the tilt test (n = 1) and protocol violation (n = 2). In 14 of the remaining 17 patients, CLS pacing emerged during the pre-syncopal phase of circulatory instability when the mean intrinsic heart rate (HR) was 88 ± 12 b.p.m. and systolic blood pressure (SBP) was 108 ± 19 mmHg. The CLS pacing rate thereafter rapidly increased to 105 ± 14 b.p.m. within a median of 0.1 min [inter-quartile range (IQR), 0.1-0.7 min] when the SBP was 99 ± 21 mmHg. At the time of maximum vasovagal effect (syncope or pre-syncope), SBP was 63 ± 17 mmHg and the CLS rate was 95 ± 13 b.p.m. The onset of CLS pacing was 1.7 min (IQR, 1.5-3.4) before syncope or lowest SBP. The total duration of CLS pacing was 5.0 min (IQR, 3.3-8.3). Closed-loop stimulation pacing was not observed in three patients who had a similar SBP decrease from 142 ± 22 mmHg at baseline to 69 ± 4 mmHg at the time of maximum vasovagal effect, but there was no significant increase in HR (59 ± 1 b.p.m.). The reproducibility of a vasovagal reflex was high. High-rate CLS pacing was observed early during the pre-syncopal phase in most patients and persisted, although attenuated, at the time of maximum vasovagal effect. ClinicalTrials.gov identifier: NCT06038708.
Identifiants
pubmed: 38340330
pii: 7606226
doi: 10.1093/europace/euae045
pmc: PMC10886438
pii:
doi:
Banques de données
ClinicalTrials.gov
['NCT06038708']
Types de publication
Clinical Study
Journal Article
Video-Audio Media
Langues
eng
Sous-ensembles de citation
IM
Subventions
Organisme : BIOTRONIK SE & Co . KG, Center for Clinical Research
Commentaires et corrections
Type : CommentIn
Informations de copyright
© The Author(s) 2024. Published by Oxford University Press on behalf of the European Society of Cardiology.
Déclaration de conflit d'intérêts
Conflict of interest: A.G., D.G., and P.N. are employees of BIOTRONIK Italia, an affiliate of Biotronik SE & Co. KG, the sponsor of the study. The other authors have no conflict of interest to declare.
Références
Int J Cardiol. 2018 Nov 1;270:149-153
pubmed: 29980368
Heart Rhythm. 2018 Jun;15(6):921-929
pubmed: 29246828
Pacing Clin Electrophysiol. 1994 May;17(5 Pt 1):882-91
pubmed: 7517523
Pacing Clin Electrophysiol. 2000 Oct;23(10 Pt 1):1502-8
pubmed: 11060870
Cardiol Clin. 2013 Feb;31(1):67-74
pubmed: 23217688
Europace. 2018 May 1;20(5):859-866
pubmed: 28407148
Circ Res. 2020 Aug 14;127(5):e126-e138
pubmed: 32460687
Eur Heart J. 2021 Feb 1;42(5):508-516
pubmed: 33279955
Pacing Clin Electrophysiol. 1992 Nov;15(11 Pt 2):1778-86
pubmed: 1279547
Arrhythm Electrophysiol Rev. 2021 Dec;10(4):244-249
pubmed: 35106176
Europace. 2010 Jul;12(7):968-71
pubmed: 20382968
Europace. 2020 Feb 1;22(2):314-319
pubmed: 31713631
J Am Coll Cardiol. 2017 Oct 3;70(14):1720-1728
pubmed: 28958328
Pacing Clin Electrophysiol. 2000 Nov;23(11 Pt 1):1632-7
pubmed: 11138300
Europace. 2000 Oct;2(4):339-42
pubmed: 11194602
Pacing Clin Electrophysiol. 2004 Jan;27(1):24-32
pubmed: 14720151
Europace. 2023 Feb 16;25(2):263-269
pubmed: 36796797
Europace. 2023 Dec 6;25(12):
pubmed: 38042981
Int J Cardiol. 2021 Jul 15;335:52-54
pubmed: 33887343
Pacing Clin Electrophysiol. 2012 Aug;35(8):990-8
pubmed: 22680238
Eur Heart J. 1992 Nov;13 Suppl E:104-12
pubmed: 1478202
Europace. 2023 Feb 16;25(2):762-766
pubmed: 36351661
N Engl J Med. 1993 Apr 15;328(15):1085-90
pubmed: 8455666