Verification of a Benchtop Hematology Analyzer with a 5-Part Differential Count: There is Nothing Wrong with Being Small.
Journal
Clinical laboratory
ISSN: 1433-6510
Titre abrégé: Clin Lab
Pays: Germany
ID NLM: 9705611
Informations de publication
Date de publication:
01 Feb 2024
01 Feb 2024
Historique:
medline:
12
2
2024
pubmed:
12
2
2024
entrez:
12
2
2024
Statut:
ppublish
Résumé
The benchtop ADVIA 560 AL hematology analyzer (Siemens Healthineers Tarrytown, NY, USA) offers a small footprint and ease of operation making it suitable for satellite laboratories and intensive care units. A verification study of this analyzer was performed. Between- and intra-run precision, carry-over, linearity, and throughput were evaluated on the ADVIA 560 AL. Accuracy was assessed on 94 patient samples by comparing the results obtained on the ADVIA 560 AL to the results on the reference Sysmex XN1000 analyzer (Sysmex Corporation, Kobe, Japan). The ADVIA 560 AL showed acceptable imprecision on control material and minimal bias in comparison to the XN 1000 on patient samples with a throughput of 60 samples per hour. The percentage carryover was not significant and the linearity was within acceptable limits. The ADVIA 560 AL bench-top analyzer is suitable for acute care centers and satellite laboratories owing to its small footprint, ease of use, and reproducible and accurate results.
Sections du résumé
BACKGROUND
BACKGROUND
The benchtop ADVIA 560 AL hematology analyzer (Siemens Healthineers Tarrytown, NY, USA) offers a small footprint and ease of operation making it suitable for satellite laboratories and intensive care units. A verification study of this analyzer was performed.
METHODS
METHODS
Between- and intra-run precision, carry-over, linearity, and throughput were evaluated on the ADVIA 560 AL. Accuracy was assessed on 94 patient samples by comparing the results obtained on the ADVIA 560 AL to the results on the reference Sysmex XN1000 analyzer (Sysmex Corporation, Kobe, Japan).
RESULTS
RESULTS
The ADVIA 560 AL showed acceptable imprecision on control material and minimal bias in comparison to the XN 1000 on patient samples with a throughput of 60 samples per hour. The percentage carryover was not significant and the linearity was within acceptable limits.
CONCLUSIONS
CONCLUSIONS
The ADVIA 560 AL bench-top analyzer is suitable for acute care centers and satellite laboratories owing to its small footprint, ease of use, and reproducible and accurate results.