Résumé

To carry out the preclinical and histological evaluation of a novel nanotechnology-based microshunt for drainage glaucoma surgery. Twelve New Zealand White rabbits were implanted with a novel microshunt and followed up for 6 weeks. The new material composite consists of the silicone polydimethylsiloxane (PDMS) and tetrapodal Zinc Oxide (ZnO-T) nano-/microparticles. The microshunts were inserted ab externo to connect the subconjunctival space with the anterior chamber. Animals were euthanized after 2 and 6 weeks for histological evaluation. Ocular health and implant position were assessed at postoperative days 1, 3, 7 and twice a week thereafter by slit lamp biomicroscopy. Intraocular pressure (IOP) was measured using rebound tonometry. A good tolerability was observed in both short- and medium-term follow-up. Intraocular pressure was reduced following surgery but increased to preoperative levels after 2 weeks. No clinical or histological signs of inflammatory or toxic reactions were seen; the fibrotic encapsulation was barely noticeable after two weeks and very mild after six weeks. The new material composite PDMS/ZnO-T is well tolerated and the associated foreign body fibrotic reaction quite mild. The new microshunt reduces the IOP for 2 weeks. Further research will elucidate a tube-like shape to improve and prolong outflow performance and longer follow-up to exclude medium-term adverse effects.

Identifiants

pubmed: 38396005
doi: 10.1038/s41598-024-54960-7
pii: 10.1038/s41598-024-54960-7
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

4452

Subventions

Organisme : federal Ministry of Education and Research in Germany (BMBF)
ID : 16KN062022
Organisme : federal Ministry of Education and Research in Germany (BMBF)
ID : 16KN062022
Organisme : federal Ministry of Education and Research in Germany (BMBF)
ID : 16KN062022
Organisme : federal Ministry of Education and Research in Germany (BMBF)
ID : 16KN062022
Organisme : federal Ministry of Education and Research in Germany (BMBF)
ID : 16KN062022
Organisme : federal Ministry of Education and Research in Germany (BMBF)
ID : 16KN062022
Organisme : federal Ministry of Education and Research in Germany (BMBF)
ID : 16KN062022
Organisme : federal Ministry of Education and Research in Germany (BMBF)
ID : 16KN062022
Organisme : federal Ministry of Education and Research in Germany (BMBF)
ID : 16KN062022
Organisme : federal Ministry of Education and Research in Germany (BMBF)
ID : 16KN062022

Informations de copyright

© 2024. The Author(s).

Références

Sihota, R., Angmo, D., Ramaswamy, D. & Dada, T. Simplifying, “target” intraocular pressure for different stages of primary open-angle glaucoma and primary angle-closure glaucoma. Indian J. Ophthalmol. 66, 495–505 (2018).
doi: 10.4103/ijo.IJO_1130_17 pubmed: 29582808 pmcid: 5892050
Lamoureux, E. L. et al. Comparing the effectiveness of selective laser trabeculoplasty with topical medication as initial treatment (the Glaucoma Initial Treatment Study): Study protocol for a randomised controlled trial. Trials 16, 406 (2015).
doi: 10.1186/s13063-015-0924-6 pubmed: 26362541 pmcid: 4567808
European Glaucoma Society Terminology and Guidelines for Glaucoma, 5th Edition. Br. J. Ophthalmol. 105, 1–169 (2021).
Chong, R. S., Crowston, J. G. & Wong, T. T. Experimental models of glaucoma filtration surgery. Acta Ophthalmol. 99, 9–15 (2021).
doi: 10.1111/aos.14485 pubmed: 32715621
Aref, A. A., Gedde, S. J. & Budenz, D. L. Glaucoma drainage implant surgery. Dev. Ophthalmol. 59, 43–52 (2017).
doi: 10.1159/000458485 pubmed: 28442686
Yu, D.-Y. et al. The critical role of the conjunctiva in glaucoma filtration surgery. Prog. Retinal Eye Res. 28, 303–328 (2009).
doi: 10.1016/j.preteyeres.2009.06.004
Zhao, X., Liu, S., Han, Y., Wang, Y. & Lin, Q. Preparation of 5-fluorouracil loaded chitosan microtube via in situ precipitation for glaucoma drainage device application: In vitro and in vivo investigation. J. Biomater. Sci. Polym. Ed. 32, 1849–1864 (2021).
doi: 10.1080/09205063.2021.1946460 pubmed: 34156326
Koz, O. G. et al. The effect of paclitaxel on conjunctival wound healing: a pilot study. J. Glaucoma 16, 610–615 (2007).
doi: 10.1097/IJG.0b013e31815c89b9 pubmed: 18091179
Takeuchi, K. et al. Solid hyaluronic acid film and the prevention of postoperative fibrous scar formation in experimental animal eyes. Arch Ophthalmol. (Chicago, Ill. : 1960) 127, 460–464 (2009).
Ishida, K. et al. Evaluation of bleb characteristics after trabeculectomy and glaucoma implant surgery in the rabbit. Ophthal. Res. 64, 68–76 (2021).
doi: 10.1159/000509135
van Mechelen, R. J. S. et al. Wound healing response after bleb-forming glaucoma surgery with a SIBS microshunt in rabbits. Translat. Vis. Sci. Technol. 11, 29 (2022).
doi: 10.1167/tvst.11.8.29
Chatzara, A., Chronopoulou, I., Theodossiadis, G., Theodossiadis, P. & Chatziralli, I. XEN implant for glaucoma treatment: A review of the literature. Semin. Ophthalmol. 34, 93–97 (2019).
doi: 10.1080/08820538.2019.1581820 pubmed: 30789072
Sonntag, S. R. et al. Zinc Oxide tetrapods modulate wound healing and cytokine release in vitro-a new antiproliferative substance in glaucoma filtering surgery. Life (Basel, Switzerland) 12 (2022).
Gapeeva, A. et al. Tetrapodal ZnO-based composite stents for minimally invasive glaucoma surgery. ACS Biomater. Sci. Eng. 9, 1352–1361 (2023).
doi: 10.1021/acsbiomaterials.2c01203 pubmed: 36776118
Ceresnakova, M., Dully, M., Murray, D., Soulimane, T. & Hudson, S. P. Stent conditioned media for in vitro evaluation of hydrophobic stent coatings. Toxicol. Vitro: Int. J. Publ. Assoc. BIBRA 76, 105212 (2021).
doi: 10.1016/j.tiv.2021.105212
Fujimoto, T. et al. Intraocular pressure-lowering effects of trabeculectomy versus microshunt insertion in rabbit eyes. Translat. Vis. Sci. Technol. 10, 9 (2021).
doi: 10.1167/tvst.10.9.9
Sonntag, S. R. et al. In vitro evaluation of zinc oxide tetrapods as a new material component for glaucoma implants. Life (Basel, Switzerland) 12 (2022).
Nakamura, K. et al. Tissue reactivity to, and stability of, glaucoma drainage device materials placed under rabbit conjunctiva. Translat. Vis. Sci. Technol. 11, 9 (2022).
doi: 10.1167/tvst.11.4.9
Acosta, A. C. et al. A newly designed glaucoma drainage implant made of poly(styrene-b-isobutylene-b-styrene): Biocompatibility and function in normal rabbit eyes. Archives of ophthalmology (Chicago, Ill. : 1960) 124, 1742–1749 (2006).
Lenzhofer, M., Hohensinn, M., Strohmaier, C. & Reitsamer, H. A. Subkonjunktivale minimalinvasive Glaukomchirurgie: Verfahren und klinische Ergebnisse. Der Ophthalmologe : Zeitschrift der Deutschen Ophthalmologischen Gesellschaft 115, 381–387 (2018).
doi: 10.1007/s00347-018-0669-1 pubmed: 29500714
Fea, A. M. et al. Early experience with the New XEN63 implant in primary open-angle glaucoma patients: Clinical outcomes. J. Clin. Med. 10 (2021).
Shute, T. S. et al. Biocompatibility of a novel microfistula implant in nonprimate mammals for the surgical treatment of glaucoma. Invest. Ophthalmol. Vis. Sci. 57, 3594–3600 (2016).
doi: 10.1167/iovs.16-19453 pubmed: 27391549
Scheres, L. M. J. et al. XEN® Gel Stent compared to PRESERFLO™ MicroShunt implantation for primary open-angle glaucoma: two-year results. Acta Ophthalmol. 99, e433–e440 (2021).
doi: 10.1111/aos.14602 pubmed: 32909682
Lee, R. M. H., Bouremel, Y., Eames, I., Brocchini, S. & Khaw, P. T. The implications of an ab interno versus ab externo surgical approach on outflow resistance of a subconjunctival drainage device for intraocular pressure control. Translat. Vis. Sci. Technol. 8, 58 (2019).
doi: 10.1167/tvst.8.3.58
García Feijoó, J. et al. A European study of the performance and safety of MINIject in patients with medically uncontrolled open-angle glaucoma (STAR-II). J. Glaucoma 29, 864–871 (2020).
Grierson, I. et al. A novel suprachoroidal microinvasive glaucoma implant: in vivo biocompatibility and biointegration. BMC Biomed. Eng. 2, 10 (2020).
doi: 10.1186/s42490-020-00045-1 pubmed: 33073174 pmcid: 7556975

Auteurs

Stefanie Gniesmer (S)

Department of Ophthalmology, University of Luebeck, Luebeck, Germany. stefanie.gniesmer@uksh.de.

Svenja Rebecca Sonntag (SR)

Department of Ophthalmology, University of Luebeck, Luebeck, Germany.

Anna Gapeeva (A)

Institute for Materials Science, Christian-Albrechts-University of Kiel, Kiel, Germany.

Ala Cojocaru (A)

Institute for Materials Science, Christian-Albrechts-University of Kiel, Kiel, Germany.
Phi-Stone AG, Kiel, Germany.

Sören Kaps (S)

Institute for Materials Science, Christian-Albrechts-University of Kiel, Kiel, Germany.

Rainer Adelung (R)

Institute for Materials Science, Christian-Albrechts-University of Kiel, Kiel, Germany.

Judith Sewing (J)

Department of Ophthalmology, University of Luebeck, Luebeck, Germany.

Aysegül Tura (A)

Department of Ophthalmology, University of Luebeck, Luebeck, Germany.

Salvatore Grisanti (S)

Department of Ophthalmology, University of Luebeck, Luebeck, Germany.

Swaantje Grisanti (S)

Department of Ophthalmology, University of Luebeck, Luebeck, Germany.

Classifications MeSH