Characterization of the on-board megavoltage imager in a magnetic resonance-guided radiotherapy machine for beam output checks.

Dosimetry MRI-Linac MV imager Quality assurance Radiotherapy Unity

Journal

Physics and imaging in radiation oncology
ISSN: 2405-6316
Titre abrégé: Phys Imaging Radiat Oncol
Pays: Netherlands
ID NLM: 101704276

Informations de publication

Date de publication:
Jan 2024
Historique:
received: 17 11 2023
revised: 12 02 2024
accepted: 12 02 2024
medline: 26 2 2024
pubmed: 26 2 2024
entrez: 26 2 2024
Statut: epublish

Résumé

We characterized the on-board megavoltage imager (MVI) of a magnetic resonance-guided radiotherapy machine for beam output checks. Linearity and repeatability of its dose response were investigated. Alignment relative to the beam under clinical circumstances was evaluated for a year using daily measurements. Linearity and short-term repeatability were excellent. Long-term repeatability drifted 0.8 % per year, which can be overcome by monthly cross calibrations. Long-term alignment was stable. Thus, the MVI has suitable characteristics for beam output checks.

Identifiants

pubmed: 38405429
doi: 10.1016/j.phro.2024.100558
pii: S2405-6316(24)00028-9
pmc: PMC10891347
doi:

Types de publication

Journal Article

Langues

eng

Pagination

100558

Informations de copyright

© 2024 The Author(s).

Déclaration de conflit d'intérêts

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Auteurs

Guido C Hilgers (GC)

Radiotherapiegroep, Arnhem/Deventer, The Netherlands.

Marijke Ikink (M)

Radiotherapiegroep, Arnhem/Deventer, The Netherlands.

Ilona Potters (I)

Radiotherapiegroep, Arnhem/Deventer, The Netherlands.

Danny Schuring (D)

Radiotherapiegroep, Arnhem/Deventer, The Netherlands.

André W Minken (AW)

Radiotherapiegroep, Arnhem/Deventer, The Netherlands.

Classifications MeSH