Multileaf collimator characterization and modeling for a 1.5 T MR-linac using static synchronous and asynchronous sweeping gaps.

MLC Characterization MR-linac sweeping gap tests

Journal

Physics in medicine and biology
ISSN: 1361-6560
Titre abrégé: Phys Med Biol
Pays: England
ID NLM: 0401220

Informations de publication

Date de publication:
27 Feb 2024
Historique:
medline: 27 2 2024
pubmed: 27 2 2024
entrez: 27 2 2024
Statut: aheadofprint

Résumé

The Elekta Unity MR-linac delivers step-and-shoot intensity modulated radiotherapy plans using a multileaf collimator (MLC) based on the Agility MLC used on conventional Elekta linacs. Currently, details of the physical Unity MLC and the computational model within its treatment planning system (TPS) Monaco are lacking in published literature. Recently, a novel approach to characterize the physical properties of MLCs was introduced using dynamic synchronous and asynchronous sweeping gap (aSG) tests. Our objective was to develop a step-and-shoot version of the dynamic aSG test to characterize the Unity MLC and the computational MLC models in the Monaco and RayStation TPSs.
Approach: Dynamic aSG were discretized into a step-and-shoot aSG by investigating the number of segments/sweep and the minimal number of monitor units (MU) per segment. The step-and-shoot aSG tests were compared to the dynamic aSG tests on a conventional linac at a source-to-detector distance of 143.5 cm, mimicking the Unity configuration. the step-and-shoot aSG tests were used to characterize the Unity MLC through measurements and dose calculations in both TPSs.
Main results: The step-and-shoot aSGs tests with 100 segments and 5 MU/segment gave results very similar to the dynamic aSG experiments. The effective tongue-and-groove width of the Unity gradually increased up to 1.4 cm from the leaf tip end. The MLC models in RayStation and Monaco agreed with experimental data within 2.0% and 10%, respectively. The largest discrepancies in Monaco were found for aSG tests with > 10 mm leaf interdigitation, which are non-typical for clinical plans. 
Significance: The step-and-shoot aSG tests accurately characterize the MLC in step-and-shoot delivery mode. The MLC model in RayStation 2023B accurately describes the tongue-and-groove and leaf tip effects whereas Monaco overestimates the tongue-and-groove shadowing further away from the leaf tip end.

Identifiants

pubmed: 38412538
doi: 10.1088/1361-6560/ad2d7d
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Informations de copyright

© 2024 Institute of Physics and Engineering in Medicine.

Auteurs

Roel Kierkels (R)

Department of Radiation Oncology, Radiotherapiegroep, P.O. Box 60160, Arnhem, 6800 JD, NETHERLANDS.

Victor Hernandez (V)

Department of Medical Physics, Hospital Universitari Sant Joan de Reus, IISPV, 43204 Tarragona, Reus, Tarragona, 43204, SPAIN.

Jordi Saez (J)

Radiation Oncology, Hospital Clinic de Barcelona, Villarroel 170, Barcelona, 08036, SPAIN.

Agnes Angerud (A)

RaySearch Laboratories AB, Eugeniavägen 18, Stockholm, Stockholm, 11368, SWEDEN.

Guido Hilgers (G)

Radiation Oncology Department, Radiotherapiegroep, P.O. Box 60160, Arnhem, 6800 JD, NETHERLANDS.

Kathrin Surmann (K)

Department of Radiation Oncology, Radiotherapiegroep, P.O. Box 60160, Arnhem, 6800 JD, NETHERLANDS.

Danny Schuring (D)

Radiation Oncology Department, Radiotherapiegroep, P.O. Box 60160, Arnhem, 6800 JD, NETHERLANDS.

André Minken (A)

Radiation Oncology Department, Radiotherapiegroep, P.O. Box 60160, Arnhem, 6800 JD, NETHERLANDS.

Classifications MeSH