Feasibility of proton dosimetry overriding planning CT with daily CBCT elaborated through generative artificial intelligence tools.

CBCT CT Deep learning artifact correction cycleGAN dosimetry hounsfield unit recovery image-to-image translation limited FOV

Journal

Computer assisted surgery (Abingdon, England)
ISSN: 2469-9322
Titre abrégé: Comput Assist Surg (Abingdon)
Pays: England
ID NLM: 101681550

Informations de publication

Date de publication:
Dec 2024
Historique:
medline: 12 3 2024
pubmed: 12 3 2024
entrez: 12 3 2024
Statut: ppublish

Résumé

Radiotherapy commonly utilizes cone beam computed tomography (CBCT) for patient positioning and treatment monitoring. CBCT is deemed to be secure for patients, making it suitable for the delivery of fractional doses. However, limitations such as a narrow field of view, beam hardening, scattered radiation artifacts, and variability in pixel intensity hinder the direct use of raw CBCT for dose recalculation during treatment. To address this issue, reliable correction techniques are necessary to remove artifacts and remap pixel intensity into Hounsfield Units (HU) values. This study proposes a deep-learning framework for calibrating CBCT images acquired with narrow field of view (FOV) systems and demonstrates its potential use in proton treatment planning updates. Cycle-consistent generative adversarial networks (cGAN) processes raw CBCT to reduce scatter and remap HU. Monte Carlo simulation is used to generate CBCT scans, enabling the possibility to focus solely on the algorithm's ability to reduce artifacts and cupping effects without considering intra-patient longitudinal variability and producing a fair comparison between planning CT (pCT) and calibrated CBCT dosimetry. To showcase the viability of the approach using real-world data, experiments were also conducted using real CBCT. Tests were performed on a publicly available dataset of 40 patients who received ablative radiation therapy for pancreatic cancer. The simulated CBCT calibration led to a difference in proton dosimetry of less than 2%, compared to the planning CT. The potential toxicity effect on the organs at risk decreased from about 50% (uncalibrated) up the 2% (calibrated). The gamma pass rate at 3%/2 mm produced an improvement of about 37% in replicating the prescribed dose before and after calibration (53.78% vs 90.26%). Real data also confirmed this with slightly inferior performances for the same criteria (65.36% vs 87.20%). These results may confirm that generative artificial intelligence brings the use of narrow FOV CBCT scans incrementally closer to clinical translation in proton therapy planning updates.

Identifiants

pubmed: 38468391
doi: 10.1080/24699322.2024.2327981
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

2327981

Auteurs

Matteo Rossi (M)

Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milan, Italy.
Laboratory of Innovation in Sleep Medicine, Istituto Auxologico Italiano, Milan, Italy.

Gabriele Belotti (G)

Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milan, Italy.

Luca Mainardi (L)

Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milan, Italy.

Guido Baroni (G)

Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milan, Italy.
Bioengineering Unit, Clinical Department, National Center for Oncological Hadrontherapy (CNAO), Pavia, Italy.

Pietro Cerveri (P)

Department of Electronics, Information and Bioengineering, Politecnico di Milano, Milan, Italy.
Laboratory of Innovation in Sleep Medicine, Istituto Auxologico Italiano, Milan, Italy.

Classifications MeSH