The impact of audiovisual breathing guidance on respiratory-triggered cardiac magnetic resonance cine imaging.
Cardiovascular magnetic resonance imaging
Cine imaging
Motion compensation
Respiratory motion
Journal
Magnetic resonance imaging
ISSN: 1873-5894
Titre abrégé: Magn Reson Imaging
Pays: Netherlands
ID NLM: 8214883
Informations de publication
Date de publication:
03 Apr 2024
03 Apr 2024
Historique:
received:
05
09
2023
revised:
15
03
2024
accepted:
02
04
2024
medline:
6
4
2024
pubmed:
6
4
2024
entrez:
5
4
2024
Statut:
aheadofprint
Résumé
In patients who have difficulty holding their breath, a free breathing (FB) respiratory-triggered (RT) bSSFP cine technique may be used. However, this technique may have inferior image quality and a longer scan time than breath-hold (BH) bSSFP cine acquisitions. This study examined the effect of an audiovisual breathing guidance (BG) system on RT bSSFP cine image quality, scan time, and ventricular measurements. This study evaluated a BG system that provides audiovisual instructions and feedback on the timing of inspiration and expiration to the patient during image acquisition using input from the respiratory bellows to guide them toward a regular breathing pattern with extended end-expiration. In this single-center prospective study in patients undergoing a clinical cardiac magnetic resonance examination, a ventricular short-axis stack of bSSFP cine images was acquired using 3 techniques in each patient: 1) FB and RT (FBRT), 2) BG system and RT (BGRT), and 3) BH. The 3 acquisitions were compared for image quality metrics (endocardial edge definition, motion artifact, and blood-to-myocardial contrast) scored on a Likert scale, scan time, and ventricular volumes and mass. Thirty-two patients (19 females; median age 21 years, IQR 18-32) completed the study protocol. For scan time, BGRT was faster than FBRT (163 s vs. 345 s, p < 0.001). Endocardial edge definition, motion artifact, and blood-to-myocardial contrast were all better for BGRT than FBRT (p < 0.001). Left ventricular (LV) end-systolic volume (ESV) was smaller (3%, p = 0.02) and LV ejection fraction (EF) was larger (0.5%, p = 0.003) with BGRT than with FBRT. There was no significant difference in LV end-diastolic volume (EDV), LV mass, right ventricular (RV) EDV, RV ESV, and RV EF. Scan times were shorter for BGRT compared to BH. Endocardial edge definition and blood-to-myocardial contrast were better for BH than BGRT. Compared to BH, the LV EDV, LV ESV, RV EDV, and RV ESV were mildly smaller (all differences <7%) for BGRT. The addition of a BG system to RT bSSFP cine acquisitions decreased the scan time and improved image quality. Further exploration of this BG approach is warranted in more diverse populations and with other free breathing sequences.
Sections du résumé
BACKGROUND
BACKGROUND
In patients who have difficulty holding their breath, a free breathing (FB) respiratory-triggered (RT) bSSFP cine technique may be used. However, this technique may have inferior image quality and a longer scan time than breath-hold (BH) bSSFP cine acquisitions. This study examined the effect of an audiovisual breathing guidance (BG) system on RT bSSFP cine image quality, scan time, and ventricular measurements.
METHODS
METHODS
This study evaluated a BG system that provides audiovisual instructions and feedback on the timing of inspiration and expiration to the patient during image acquisition using input from the respiratory bellows to guide them toward a regular breathing pattern with extended end-expiration. In this single-center prospective study in patients undergoing a clinical cardiac magnetic resonance examination, a ventricular short-axis stack of bSSFP cine images was acquired using 3 techniques in each patient: 1) FB and RT (FBRT), 2) BG system and RT (BGRT), and 3) BH. The 3 acquisitions were compared for image quality metrics (endocardial edge definition, motion artifact, and blood-to-myocardial contrast) scored on a Likert scale, scan time, and ventricular volumes and mass.
RESULTS
RESULTS
Thirty-two patients (19 females; median age 21 years, IQR 18-32) completed the study protocol. For scan time, BGRT was faster than FBRT (163 s vs. 345 s, p < 0.001). Endocardial edge definition, motion artifact, and blood-to-myocardial contrast were all better for BGRT than FBRT (p < 0.001). Left ventricular (LV) end-systolic volume (ESV) was smaller (3%, p = 0.02) and LV ejection fraction (EF) was larger (0.5%, p = 0.003) with BGRT than with FBRT. There was no significant difference in LV end-diastolic volume (EDV), LV mass, right ventricular (RV) EDV, RV ESV, and RV EF. Scan times were shorter for BGRT compared to BH. Endocardial edge definition and blood-to-myocardial contrast were better for BH than BGRT. Compared to BH, the LV EDV, LV ESV, RV EDV, and RV ESV were mildly smaller (all differences <7%) for BGRT.
CONCLUSIONS
CONCLUSIONS
The addition of a BG system to RT bSSFP cine acquisitions decreased the scan time and improved image quality. Further exploration of this BG approach is warranted in more diverse populations and with other free breathing sequences.
Identifiants
pubmed: 38579974
pii: S0730-725X(24)00107-3
doi: 10.1016/j.mri.2024.04.001
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Informations de copyright
Copyright © 2024. Published by Elsevier Inc.
Déclaration de conflit d'intérêts
Declaration of competing interest Jihye Jang and Annerieke Heuvelink-Marck are employees of Philips. Other authors have no competing interests relevant to this article to disclose.