Torque forces of expandable titanium vertebral body replacement cages during expansion and subsidence in the osteoporotic lumbar spine.

Osteoporosis Strain gauges Subsidence Titanium expandable cages Torque forces Vertebral body replacement

Journal

Clinical biomechanics (Bristol, Avon)
ISSN: 1879-1271
Titre abrégé: Clin Biomech (Bristol, Avon)
Pays: England
ID NLM: 8611877

Informations de publication

Date de publication:
09 Apr 2024
Historique:
received: 15 07 2023
revised: 02 04 2024
accepted: 05 04 2024
medline: 11 4 2024
pubmed: 11 4 2024
entrez: 10 4 2024
Statut: aheadofprint

Résumé

The application of expandable titanium-cages has gained widespread use in vertebral body replacement for indications such as burst fractures, tumors and infectious destruction. However, torque forces necessary for a satisfactory expansion of these implants and for subsidence of them into the adjacent vertebrae are unknown within the osteoporotic spine. Six fresh-frozen human, osteoporotic, lumbar spines were dorsally instrumented with titanium implants (L2-L4) and a partial corpectomy of L3 was performed. An expandable titanium-cage was inserted ventrally and expanded by both residents and senior surgeons until fixation was deemed sufficient, based on haptic feedback. Torque forces for expansion were measured in Nm. Expansion was then continued until cage subsidence occurred. Torque forces necessary for subsidence were recorded. Strain of the dorsal rods during expansion was measured with strain gauges. The mean torque force for fixation of cages was 1.17 Nm (0.9 Nm for residents, 1.4 Nm for senior surgeons, p = .06). The mean torque force for subsidence of cages was 3.1 Nm (p = .005). Mean peak strain of the dorsal rods was 970 μm/m during expansion and 1792 μm/m at subsidence of cages (p = .004). The use of expandable titanium-cages for vertebral body replacement seems to be a primarily safe procedure even within the osteoporotic spine as torque forces required for subsidence of cages are nearly three times higher than those needed for fixation. Most of the expansion load is absorbed by straining of the dorsal instrumentation. Rod materials other than titanium may alter the torque forces found in this study.

Sections du résumé

BACKGROUND BACKGROUND
The application of expandable titanium-cages has gained widespread use in vertebral body replacement for indications such as burst fractures, tumors and infectious destruction. However, torque forces necessary for a satisfactory expansion of these implants and for subsidence of them into the adjacent vertebrae are unknown within the osteoporotic spine.
METHODS METHODS
Six fresh-frozen human, osteoporotic, lumbar spines were dorsally instrumented with titanium implants (L2-L4) and a partial corpectomy of L3 was performed. An expandable titanium-cage was inserted ventrally and expanded by both residents and senior surgeons until fixation was deemed sufficient, based on haptic feedback. Torque forces for expansion were measured in Nm. Expansion was then continued until cage subsidence occurred. Torque forces necessary for subsidence were recorded. Strain of the dorsal rods during expansion was measured with strain gauges.
FINDINGS RESULTS
The mean torque force for fixation of cages was 1.17 Nm (0.9 Nm for residents, 1.4 Nm for senior surgeons, p = .06). The mean torque force for subsidence of cages was 3.1 Nm (p = .005). Mean peak strain of the dorsal rods was 970 μm/m during expansion and 1792 μm/m at subsidence of cages (p = .004).
INTERPRETATION CONCLUSIONS
The use of expandable titanium-cages for vertebral body replacement seems to be a primarily safe procedure even within the osteoporotic spine as torque forces required for subsidence of cages are nearly three times higher than those needed for fixation. Most of the expansion load is absorbed by straining of the dorsal instrumentation. Rod materials other than titanium may alter the torque forces found in this study.

Identifiants

pubmed: 38599132
pii: S0268-0033(24)00071-8
doi: 10.1016/j.clinbiomech.2024.106239
pii:
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

106239

Informations de copyright

Copyright © 2024 The Author(s). Published by Elsevier Ltd.. All rights reserved.

Déclaration de conflit d'intérêts

Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Auteurs

Krishnan Sircar (K)

Department of Orthopedic and Trauma Surgery, University of Cologne, Faculty of Medicine and University Hospital Cologne, Kerpener-Strasse 62, 50937 Cologne, Germany. Electronic address: krishnan.sircar@uk-koeln.de.

Maximilian Weber (M)

Department of Orthopedic and Trauma Surgery, University of Cologne, Faculty of Medicine and University Hospital Cologne, Kerpener-Strasse 62, 50937 Cologne, Germany.

Sebastian G Walter (SG)

Department of Orthopedic and Trauma Surgery, University of Cologne, Faculty of Medicine and University Hospital Cologne, Kerpener-Strasse 62, 50937 Cologne, Germany.

Nadine Ott (N)

Department of Orthopedic and Trauma Surgery, University of Cologne, Faculty of Medicine and University Hospital Cologne, Kerpener-Strasse 62, 50937 Cologne, Germany.

Andreas Prescher (A)

Institute of Molecular and Cellular Anatomy, RWTH Aachen University Hospital, Wendlingweg 2, 52070 Aachen, Germany.

Peer Eysel (P)

Department of Orthopedic and Trauma Surgery, University of Cologne, Faculty of Medicine and University Hospital Cologne, Kerpener-Strasse 62, 50937 Cologne, Germany.

Nikolaus Kernich (N)

Department of Orthopedic and Trauma Surgery, University of Cologne, Faculty of Medicine and University Hospital Cologne, Kerpener-Strasse 62, 50937 Cologne, Germany.

Classifications MeSH