Efficiency of CONTRAfluran™ in reducing sevoflurane pollution from maintenance anaesthesia in minimal flow end-tidal control mode for laparoscopic surgery.

CONTRAfluran™ environment greenhouse gas sevoflurane volatile capture technology

Journal

Anaesthesia
ISSN: 1365-2044
Titre abrégé: Anaesthesia
Pays: England
ID NLM: 0370524

Informations de publication

Date de publication:
12 Apr 2024
Historique:
accepted: 08 03 2024
medline: 12 4 2024
pubmed: 12 4 2024
entrez: 12 4 2024
Statut: aheadofprint

Résumé

Recommendations exist that aim to mitigate the substantial ecological impact of anaesthesia. One option is to use anaesthetic gas capturing technology at anaesthesia workstation exhausts to harvest and recycle volatile agents. However, the efficiency of such technology is mainly unverified in vivo. The efficiency of CONTRAfluran™ in capturing sevoflurane from an anaesthesia workstation exhaust (when set to minimal flow and end-tidal control mode) was evaluated in 70 adult patients scheduled for general or bariatric laparoscopic surgery. The weight of the sevoflurane vaporiser and CONTRAfluran canister was measured before and after each case, to calculate total sevoflurane consumption and retention. Retention was measured after the minimal flow maintenance phase and after the high flow washout phase. The total retention efficiency was the fraction of all consumed sevoflurane captured by the CONTRAfluran canister. The primary objective was to examine the retention efficiency of CONTRAfluran in a clinical surgical setting, where all feasible strategies to minimise sevoflurane consumption and optimise the efficacy of CONTRAfluran were utilised. The secondary objective was to analyse the correlation between mass transfer and the duration of the case. Mean (SD) volume of sevoflurane captured using CONTRAfluran was 4.82 (1.41) ml, representing 45% (95%CI 42-48%) of all sevoflurane administered. The highest amount of retention was found during the washout phase. Retention efficiency did not correlate with the duration of the case. Over half of the sevoflurane administered was not captured by the CONTRAfluran canister when minimal flow techniques were used, likely due to residual accumulation of sevoflurane in the patient after tracheal extubation or, to a lesser extent, due to ventilation system leakage. However, as every prevented emission is commendable, CONTRAfluran may be a potentially valuable tool for reducing the environmental footprint of sevoflurane-based anaesthesia.

Sections du résumé

BACKGROUND BACKGROUND
Recommendations exist that aim to mitigate the substantial ecological impact of anaesthesia. One option is to use anaesthetic gas capturing technology at anaesthesia workstation exhausts to harvest and recycle volatile agents. However, the efficiency of such technology is mainly unverified in vivo.
METHODS METHODS
The efficiency of CONTRAfluran™ in capturing sevoflurane from an anaesthesia workstation exhaust (when set to minimal flow and end-tidal control mode) was evaluated in 70 adult patients scheduled for general or bariatric laparoscopic surgery. The weight of the sevoflurane vaporiser and CONTRAfluran canister was measured before and after each case, to calculate total sevoflurane consumption and retention. Retention was measured after the minimal flow maintenance phase and after the high flow washout phase. The total retention efficiency was the fraction of all consumed sevoflurane captured by the CONTRAfluran canister. The primary objective was to examine the retention efficiency of CONTRAfluran in a clinical surgical setting, where all feasible strategies to minimise sevoflurane consumption and optimise the efficacy of CONTRAfluran were utilised. The secondary objective was to analyse the correlation between mass transfer and the duration of the case.
RESULTS RESULTS
Mean (SD) volume of sevoflurane captured using CONTRAfluran was 4.82 (1.41) ml, representing 45% (95%CI 42-48%) of all sevoflurane administered. The highest amount of retention was found during the washout phase. Retention efficiency did not correlate with the duration of the case.
CONCLUSIONS CONCLUSIONS
Over half of the sevoflurane administered was not captured by the CONTRAfluran canister when minimal flow techniques were used, likely due to residual accumulation of sevoflurane in the patient after tracheal extubation or, to a lesser extent, due to ventilation system leakage. However, as every prevented emission is commendable, CONTRAfluran may be a potentially valuable tool for reducing the environmental footprint of sevoflurane-based anaesthesia.

Identifiants

pubmed: 38606765
doi: 10.1111/anae.16289
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Informations de copyright

© 2024 Association of Anaesthetists.

Références

Kampman JM, Sperna Weiland NH. Anaesthesia and environment: impact of a green anaesthesia on economics. Curr Opin Anaesthesiol 2023; 36: 188–195. https://doi.org/10.1097/aco.0000000000001243.
Devlin‐Hegedus JA, McGain F, Harris RD, Sherman JD. Action guidance for addressing pollution from inhalational anaesthetics. Anaesthesia 2022; 77: 1023–1029. https://doi.org/10.1111/anae.15785.
Buhre W, De Robertis E, Gonzalez‐Pizarro P. The Glasgow declaration on sustainability in Anaesthesiology and Intensive Care. Eur J Anaesthesiol 2023; 40: 461–464. https://doi.org/10.1097/eja.0000000000001862.
Kalmar AF, Verdonck P, Saxena S, Mulier J. Proper use of CONTRAfluran™ for optimal desorption and reuse of volatile anaesthetics. Br J Anaesth 2023; 131: e71–e72. https://doi.org/10.1016/j.bja.2023.06.046.
Cavalloa G, Cicerib E, Metrangoloa P, Piacevolic Q, Resnatia G, Terraneoa G. Sevoflurane: impurities and stability testing. J Fluor Chem 2019; 226: 109363. https://doi.org/10.1016/j.jfluchem.2019.109363.
Hu EP, Yap A, Davies JF, Goyagi T, McGain F. Global practices in desflurane use. Br J Anaesth 2023. Epub 13 September. https://doi.org/10.1016/j.bja.2023.08.018.
Gandhi J, Barker K, Cross S, Goddard A, Vaghela M, Cooper A. Volatile capture technology in sustainable anaesthetic practice: a narrative review. Anaesthesia 2024; 79: 261–269. https://doi.org/10.1111/anae.16207.
Laster MJ, Fang Z, Eger EI 2nd. Specific gravities of desflurane, enflurane, halothane, isoflurane, and sevoflurane. Anesth Analg 1994; 78: 1152–1153. https://doi.org/10.1213/00000539‐199406000‐00022.
Tokumine J, Nitta K, Higa T, Gushiken K, Okayama H, Sugahara K. Leakage in anesthesia circuits ‐ a comparison between 1998 and 2005 investigations. Masui 2007; 56: 453–458.
Hinterberg J, Beffart T, Gabriel A, Holzschneider M, Tartler TM, Schaefer MS, Kienbaum P. Efficiency of inhaled anaesthetic recapture in clinical practice. Br J Anaesth 2022; 129: e79–e81. https://doi.org/10.1016/j.bja.2022.04.009.
Intergovernmental Panel on Climate Change. Climate Change 2021: The Physical Science Basis; Chapter 7: The Earth's energy budget, climate feedbacks, and climate sensitivity ‐ Supplementary Material. 2021. https://www.ipcc.ch/report/ar6/wg1/downloads/report/IPCC_AR6_WGI_FGD_Chapter07_SM.pdf (accessed 19/12/2023).
Shelton C, Barker K, Winter BJ. Efficiency of inhaled anaesthetic recapture in clinical practice. Br J Anaesth 2022; 129: e114–e116. https://doi.org/10.1016/j.bja.2022.07.035.
Hu X, Pierce JMT, Taylor T, Morrissey K. The carbon footprint of general anaesthetics: a case study in the UK. Resour Conserv Recycl 2021; 167: 105411. https://doi.org/10.1016/j.resconrec.2021.105411.
Shiralkar S, Field E, Murphy E, Shelton C. The role of volatile capture technology in desflurane disposal from decommissioned vaporisers. Anaesthesia 2023; 78: 1298–1300. https://doi.org/10.1111/anae.16044.
Gandhi J, Baxter I. Efficiency of inhaled anaesthetic recapture in clinical practice. Br J Anaesth 2023; 130: e464–e465. https://doi.org/10.1016/j.bja.2023.02.013.
Vaghela M, Kay RH, Jones L, Greig P. Inhalational anaesthetics: an assessment of agent delivery and capture. Anaesthesia 2023; 78: 784–785. https://doi.org/10.1111/anae.15981.
Bernat M, Boyer A, Roche M, et al. Reducing the carbon footprint of general anaesthesia: a comparison of total intravenous anaesthesia vs. a mixed anaesthetic strategy in 47,157 adult patients. Anaesthesia 2024; 79: 309–317. https://doi.org/10.1111/anae.16221.
Sherman JD, Barrick B. Total intravenous anesthetic versus inhaled anesthetic: pick your poison. Anesth Analg 2019; 128: 13–15. https://doi.org/10.1213/ane.0000000000003898.

Auteurs

Harold Mulier (H)

Department of Anaesthesiology, University Hospitals Leuven, Leuven, Belgium.
Department of Anaesthesia and Critical Care, AZ Sint-Jan Brugge, Bruges, Belgium.

Michel M R F Struys (MMRF)

Department of Anaesthesiology, University Medical Center Groningen and University of Groningen, Groningen, The Netherlands.
Department of Basic and Applied Medical Sciences, Ghent University, Gent, Belgium.

Hugo Vereecke (H)

Department of Anaesthesia and Critical Care, AZ Sint-Jan Brugge, Bruges, Belgium.
Department of Anaesthesiology, University Medical Center Groningen and University of Groningen, Groningen, The Netherlands.

Steffen Rex (S)

Department of Anaesthesiology, University Hospitals Leuven, Leuven, Belgium.
Department of Cardiovascular Sciences, KU Leuven - University of Leuven, Leuven, Belgium.

An Teunkens (A)

Department of Anaesthesiology, University Hospitals Leuven, Leuven, Belgium.
Department of Cardiovascular Sciences, KU Leuven - University of Leuven, Leuven, Belgium.

Alain F Kalmar (AF)

Department of Anaesthesia and Critical Care, AZ Sint-Jan Brugge, Bruges, Belgium.
Department of Electronics and Information Systems, IBiTech, Ghent University, Gent, Belgium.

Classifications MeSH