Reduced Continuity Index with Proactive Esophageal Cooling Compared to Luminal Temperature Monitoring During Radiofrequency Ablation.


Journal

medRxiv : the preprint server for health sciences
Titre abrégé: medRxiv
Pays: United States
ID NLM: 101767986

Informations de publication

Date de publication:
12 Apr 2024
Historique:
medline: 22 4 2024
pubmed: 22 4 2024
entrez: 22 4 2024
Statut: epublish

Résumé

Proactive esophageal cooling is FDA cleared to reduce the likelihood of esophageal injury during radiofrequency ablation for treatment of atrial fibrillation (AF). Long-term follow-up data have also shown improved freedom from arrhythmia with proactive esophageal cooling compared to luminal esophageal temperature (LET) monitoring during pulmonary vein isolation (PVI). One hypothesized mechanism is improved lesion contiguity (as measured by the Continuity Index) with the use of cooling. We aimed to compare the Continuity Index of PVI cases using proactive esophageal cooling to those using LET monitoring. Continuity Index was calculated for PVI cases at two different hospitals within the same health system using a slightly modified Continuity Index to facilitate both real-time calculation during observation of PVI cases and retrospective determination from recorded cases. The results were then compared between proactively cooled cases and those using LET monitoring. Continuity Indices for a total of 101 cases were obtained; 77 cases using proactive esophageal cooling and 24 cases using traditional LET monitoring. With proactive esophageal cooling, the average Continuity Index was 2.7 (1.3 on the left pulmonary vein, and 1.5 on the right pulmonary vein). With LET monitoring, the average Continuity Index was 27.3 (14.3 on the left, and 12.9 on the right), for a difference of 24.6 (p < 0.001). Proactive esophageal cooling during PVI is associated with significantly improved lesion contiguity when compared to LET monitoring. This finding may offer a mechanism for the greater freedom from arrhythmia seen with proactive cooling in long-term follow-up.

Sections du résumé

Background UNASSIGNED
Proactive esophageal cooling is FDA cleared to reduce the likelihood of esophageal injury during radiofrequency ablation for treatment of atrial fibrillation (AF). Long-term follow-up data have also shown improved freedom from arrhythmia with proactive esophageal cooling compared to luminal esophageal temperature (LET) monitoring during pulmonary vein isolation (PVI). One hypothesized mechanism is improved lesion contiguity (as measured by the Continuity Index) with the use of cooling. We aimed to compare the Continuity Index of PVI cases using proactive esophageal cooling to those using LET monitoring.
Methods UNASSIGNED
Continuity Index was calculated for PVI cases at two different hospitals within the same health system using a slightly modified Continuity Index to facilitate both real-time calculation during observation of PVI cases and retrospective determination from recorded cases. The results were then compared between proactively cooled cases and those using LET monitoring.
Results UNASSIGNED
Continuity Indices for a total of 101 cases were obtained; 77 cases using proactive esophageal cooling and 24 cases using traditional LET monitoring. With proactive esophageal cooling, the average Continuity Index was 2.7 (1.3 on the left pulmonary vein, and 1.5 on the right pulmonary vein). With LET monitoring, the average Continuity Index was 27.3 (14.3 on the left, and 12.9 on the right), for a difference of 24.6 (p < 0.001).
Conclusion UNASSIGNED
Proactive esophageal cooling during PVI is associated with significantly improved lesion contiguity when compared to LET monitoring. This finding may offer a mechanism for the greater freedom from arrhythmia seen with proactive cooling in long-term follow-up.