Assessment of Preanalytical Cerebrospinal Fluid Handling and Storage Factors on Measurement of Aβ1-42, Aβ1-40, and pTau181 Using an Automated Chemiluminescent Platform.
Journal
The journal of applied laboratory medicine
ISSN: 2576-9456
Titre abrégé: J Appl Lab Med
Pays: England
ID NLM: 101693884
Informations de publication
Date de publication:
07 May 2024
07 May 2024
Historique:
received:
22
10
2023
accepted:
11
03
2024
medline:
7
5
2024
pubmed:
7
5
2024
entrez:
7
5
2024
Statut:
aheadofprint
Résumé
Standardizing cerebrospinal fluid (CSF) laboratory protocols will improve the reliability and availability of clinical biomarker testing required for prescription of novel Alzheimer disease (AD) therapies. This study evaluated several preanalytical handling and storage factors common to β-amyloid1-42 (Aβ1-42), β-amyloid1-40 (Aβ1-40), and phosphorylated tau (pTau181) concentrations including storage at different temperatures, extended cap contact, various mixing methods, and multiple freeze-thaw cycles. Aβ1-42, Aβ1-40, and pTau181 concentrations were measured using LUMIPULSE G1200 automated assays. Samples were collected in polypropylene tubes of various volumes. Sample cap-contact was evaluated by storing samples in upright and inverted positions at either 4°C for 1 week or -80°C for 1 month. To assess mixing methods, samples were freeze-thawed and mixed by inversion, vortex, horizontal roller, or unmixed prior to assay sampling. The impact of successive freeze-thaw cycles was assessed through freezing, thawing, and analyzing CSF samples. Short-term storage at 4°C did not affect Aβ1-42, Aβ1-40, or pTau181 measurements in any tube type. Tube cap contact affected Aβ1-42 in 2.5 mL tubes and pTau181 levels in 10 mL tubes. No difference was observed between mixing methods. After 4 freeze-thaw cycles, Aβ1-42 significantly decreased but Aβ1-40 remained unchanged. Utilizing the Aβ1-42/Aβ1-40 ratio, Aβ1-42 values normalized, maintaining ratio values within ±5% of baseline measurements. Storage of CSF at 4°C for 1 week or -80°C for 1 month did not significantly affect Aβ1-42, Aβ1-40, pTau181, or associated ratio measurements. Tube cap-contact impacted pTau181 and pTau181/Aβ1-42 values in larger tubes. Mixing methods are equivalent. The Aβ1-42/Aβ1-40 ratio compensates for freeze-thaw variability up to 4 cycles.
Sections du résumé
BACKGROUND
BACKGROUND
Standardizing cerebrospinal fluid (CSF) laboratory protocols will improve the reliability and availability of clinical biomarker testing required for prescription of novel Alzheimer disease (AD) therapies. This study evaluated several preanalytical handling and storage factors common to β-amyloid1-42 (Aβ1-42), β-amyloid1-40 (Aβ1-40), and phosphorylated tau (pTau181) concentrations including storage at different temperatures, extended cap contact, various mixing methods, and multiple freeze-thaw cycles.
METHODS
METHODS
Aβ1-42, Aβ1-40, and pTau181 concentrations were measured using LUMIPULSE G1200 automated assays. Samples were collected in polypropylene tubes of various volumes. Sample cap-contact was evaluated by storing samples in upright and inverted positions at either 4°C for 1 week or -80°C for 1 month. To assess mixing methods, samples were freeze-thawed and mixed by inversion, vortex, horizontal roller, or unmixed prior to assay sampling. The impact of successive freeze-thaw cycles was assessed through freezing, thawing, and analyzing CSF samples.
RESULTS
RESULTS
Short-term storage at 4°C did not affect Aβ1-42, Aβ1-40, or pTau181 measurements in any tube type. Tube cap contact affected Aβ1-42 in 2.5 mL tubes and pTau181 levels in 10 mL tubes. No difference was observed between mixing methods. After 4 freeze-thaw cycles, Aβ1-42 significantly decreased but Aβ1-40 remained unchanged. Utilizing the Aβ1-42/Aβ1-40 ratio, Aβ1-42 values normalized, maintaining ratio values within ±5% of baseline measurements.
CONCLUSIONS
CONCLUSIONS
Storage of CSF at 4°C for 1 week or -80°C for 1 month did not significantly affect Aβ1-42, Aβ1-40, pTau181, or associated ratio measurements. Tube cap-contact impacted pTau181 and pTau181/Aβ1-42 values in larger tubes. Mixing methods are equivalent. The Aβ1-42/Aβ1-40 ratio compensates for freeze-thaw variability up to 4 cycles.
Identifiants
pubmed: 38712812
pii: 7666101
doi: 10.1093/jalm/jfae033
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Informations de copyright
© Association for Diagnostics & Laboratory Medicine 2024. All rights reserved. For commercial re-use, please contact reprints@oup.com for reprints and translation rights for reprints. All other permissions can be obtained through our RightsLink service via the Permissions link on the article page on our site—for further information please contact journals.permissions@oup.com.