Effect of systemic vascular resistance on the agreement between stroke volume by non-invasive pulse wave analysis and Doppler ultrasound in healthy volunteers.
Journal
PloS one
ISSN: 1932-6203
Titre abrégé: PLoS One
Pays: United States
ID NLM: 101285081
Informations de publication
Date de publication:
2024
2024
Historique:
received:
17
10
2023
accepted:
27
03
2024
medline:
7
5
2024
pubmed:
7
5
2024
entrez:
7
5
2024
Statut:
epublish
Résumé
Stroke volume can be estimated beat-to-beat and non-invasively by pulse wave analysis (PWA). However, its reliability has been questioned during marked alterations in systemic vascular resistance (SVR). We studied the effect of SVR on the agreement between stroke volume by PWA and Doppler ultrasound during reductions in stroke volume in healthy volunteers. In a previous study we simultaneously measured stroke volume by PWA (SVPWA) and suprasternal Doppler ultrasound (SVUS). We exposed 16 healthy volunteers to lower body negative pressure (LBNP) to reduce stroke volume in combination with isometric hand grip to elevate SVR. LBNP was increased by 20 mmHg every 6 minutes from 0 to 80 mmHg, or until hemodynamic decompensation. The agreement between SVPWA and SVUS was examined using Bland-Altman analysis with mixed regression. Within-subject limits of agreement (LOA) was calculated from the residual standard deviation. SVRUS was calculated from SVUS. We allowed for a sloped bias line by introducing the mean of the methods and SVRUS as explanatory variables to examine whether the agreement was dependent on the magnitude of stroke volume and SVRUS. Bias ± limits of agreement (LOA) was 27.0 ± 30.1 mL. The within-subject LOA was ±11.1 mL. The within-subject percentage error was 14.6%. The difference between methods decreased with higher means of the methods (-0.15 mL/mL, confidence interval (CI): -0.19 to -0.11, P<0.001). The difference between methods increased with higher SVRUS (0.60 mL/mmHg × min × L-1, 95% CI: 0.48 to 0.72, P<0.001). PWA overestimated stroke volume compared to Doppler ultrasound during reductions in stroke volume and elevated SVR in healthy volunteers. The agreement between SVPWA and SVUS decreased during increases in SVR. This is relevant in settings where a high level of reliability is required.
Sections du résumé
BACKGROUND
BACKGROUND
Stroke volume can be estimated beat-to-beat and non-invasively by pulse wave analysis (PWA). However, its reliability has been questioned during marked alterations in systemic vascular resistance (SVR). We studied the effect of SVR on the agreement between stroke volume by PWA and Doppler ultrasound during reductions in stroke volume in healthy volunteers.
METHODS
METHODS
In a previous study we simultaneously measured stroke volume by PWA (SVPWA) and suprasternal Doppler ultrasound (SVUS). We exposed 16 healthy volunteers to lower body negative pressure (LBNP) to reduce stroke volume in combination with isometric hand grip to elevate SVR. LBNP was increased by 20 mmHg every 6 minutes from 0 to 80 mmHg, or until hemodynamic decompensation. The agreement between SVPWA and SVUS was examined using Bland-Altman analysis with mixed regression. Within-subject limits of agreement (LOA) was calculated from the residual standard deviation. SVRUS was calculated from SVUS. We allowed for a sloped bias line by introducing the mean of the methods and SVRUS as explanatory variables to examine whether the agreement was dependent on the magnitude of stroke volume and SVRUS.
RESULTS
RESULTS
Bias ± limits of agreement (LOA) was 27.0 ± 30.1 mL. The within-subject LOA was ±11.1 mL. The within-subject percentage error was 14.6%. The difference between methods decreased with higher means of the methods (-0.15 mL/mL, confidence interval (CI): -0.19 to -0.11, P<0.001). The difference between methods increased with higher SVRUS (0.60 mL/mmHg × min × L-1, 95% CI: 0.48 to 0.72, P<0.001).
CONCLUSION
CONCLUSIONS
PWA overestimated stroke volume compared to Doppler ultrasound during reductions in stroke volume and elevated SVR in healthy volunteers. The agreement between SVPWA and SVUS decreased during increases in SVR. This is relevant in settings where a high level of reliability is required.
Identifiants
pubmed: 38713665
doi: 10.1371/journal.pone.0302159
pii: PONE-D-23-32409
doi:
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM
Pagination
e0302159Informations de copyright
Copyright: © 2024 Lie et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Déclaration de conflit d'intérêts
The authors have declared that no competing interests exist.