Evaluation of a new transpalpebral tonometer for self-measuring intraocular pressure.
Journal
PloS one
ISSN: 1932-6203
Titre abrégé: PLoS One
Pays: United States
ID NLM: 101285081
Informations de publication
Date de publication:
2024
2024
Historique:
received:
27
07
2023
accepted:
07
04
2024
medline:
15
5
2024
pubmed:
15
5
2024
entrez:
15
5
2024
Statut:
epublish
Résumé
We developed a novel transpalpebral self-tonometer called the TapEye tonometer (TET) based on palpation of the upper eyelid. Our goal was to evaluate a method for improving the accuracy of measuring intraocular pressure (IOP) through the eyelid. Participants underwent standardized training by technicians and were required to be able to use the TET for study inclusion. Subsequently, a noncontact tonometer and Goldmann applanation tonometer (GAT) were used. All participants were instructed to measure their IOPs using the three tonometers at baseline (visit 1) and at 1 month (visit 2). At visit 2, the corrected IOP value measured by the TET (c-TET) was calculated using the difference between the TET and GAT measurements obtained at visit 1. No significant correlations were found between the TET and GAT measurements at any visit, but the correlation between the c-TET and GAT measurements at visit 2 was significant. The mean difference between the c-TET and GAT measurements was 0.4 ± 3.7 mmHg in the right eye and 0.5 ± 3.4 mmHg in the left eye. After correcting the IOP based on the difference between the TET and GAT measurements at the initial visit, the corrected IOP value of the TET was correlated with that of the GAT at the second visit. The TET has the potential to address an unmet need by providing a tool for minimally invasive IOP measurements. Clinical trial registration number: jRCTs032220268.
Identifiants
pubmed: 38748752
doi: 10.1371/journal.pone.0302568
pii: PONE-D-23-21962
doi:
Types de publication
Journal Article
Evaluation Study
Langues
eng
Sous-ensembles de citation
IM
Pagination
e0302568Informations de copyright
Copyright: © 2024 Hidenaga Kobashi. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Déclaration de conflit d'intérêts
H.K.: CEO and equity owner, Toneasy Inc.; Patent, Toneasy Inc. This does not alter our adherence to PLOS ONE policies on sharing data and materials.