Clinical outcomes of the Myval transcatheter heart valve system in patients with severe aortic valve stenosis: a two-year follow-up observational study.
Myval valve
aortic valve stenosis
transcatheter aortic valve implantation
Journal
Archives of medical science : AMS
ISSN: 1734-1922
Titre abrégé: Arch Med Sci
Pays: Poland
ID NLM: 101258257
Informations de publication
Date de publication:
2024
2024
Historique:
received:
04
08
2023
accepted:
13
12
2023
medline:
17
5
2024
pubmed:
17
5
2024
entrez:
17
5
2024
Statut:
epublish
Résumé
Limited data exist on long-term follow-up of severe aortic stenosis (SAS) patients who have undergone transcatheter aortic valve implantation (TAVI) with a new generation, balloon expandable Myval transcatheter heart valve (THV). Thus, we sought to investigate the performance and 2-year clinical outcome of the Myval THV system based on Valve Academic Research Consortium-3 (VARC-3) criteria. A multi-centre, registry-based, observational study was conducted, which included 207 consecutive degenerative SAS patients, from Turkey ( Technical success was observed in 204 (99%), device success was observed in 189 (91%), early safety was observed in 161 (78%), and clinical efficacy was observed in 163 (79%) patients. The 30-day death rate was 7.7%; of these, 3.4% were due to cardiovascular reasons. All-cause and cardiovascular mortality rates were 9.7% and 4.3% at 1-year follow-up, and 17.4% and 9.7% at 2-year follow-up, respectively. Incidence of ≥ moderate paravalvular leak (PVL) at 30 days, 1 year and 2 years of follow-up were 3.4%, 4.3% and 4.8%. A total of 11.1% of patients required a permanent pacemaker implantation (PPI) at 30 days after implantation, while the cumulative rate of PPI at 2 years was 12.1%. In this cohort of patients with SAS, the Myval was found to be safe and effective in up to 2 years of follow-up.
Identifiants
pubmed: 38757027
doi: 10.5114/aoms/176937
pii: 176937
pmc: PMC11094825
doi:
Types de publication
Journal Article
Langues
eng
Pagination
410-419Informations de copyright
Copyright: © 2024 Termedia & Banach.
Déclaration de conflit d'intérêts
The authors declare no conflict of interest.