Conducted Electrical Weapon Fire Risk in the Presence of Supplemental Oxygen.

conducted electrical weapon flammability oxygen tactical EMS taser

Journal

Journal of special operations medicine : a peer reviewed journal for SOF medical professionals
ISSN: 1553-9768
Titre abrégé: J Spec Oper Med
Pays: United States
ID NLM: 101158402

Informations de publication

Date de publication:
13 Jun 2024
Historique:
pubmed: 12 6 2024
medline: 12 6 2024
entrez: 12 6 2024
Statut: aheadofprint

Résumé

Conducted electrical weapons (CEWs) are nonlethal weapons used in tactical environments. Tactical EMS (TEMS) operators provide patient care in environments where CEWs are present. CEWs may produce a spark that can be a source of ignition. When supplemental oxygen is in use by a TEMS operator, the flammability risk from the CEW is unknown. We measured oxygen levels over 20 minutes in an enclosed space with a supplemental oxygen source maximally flowing. Measurements were taken at various distances from the oxygen source to establish baseline concentrations. These concentrations were replicated within a plexiglass box where a CEW was activated to create a sparking arc between probes embedded in a fresh swine shoulder (with skin intact). Various flammable materials, typically found in a patient care setting were used as potential fuel sources. The highest oxygen levels were achieved directly at the source. At 15cm from the source, the maximum oxygen concentration was 31.5%. Within the box, ignition was only achieved at oxygen concentrations greater than 45% and only when human hair was present as fuel. No ignition was achieved at oxygen levels below this regardless of the tested fuel present. Ignition from a CEW is possible at supplemental oxygen levels greater than 45% when human hair is present. In an enclosed space, oxygen concentration levels of 45% are only present within 15cm of a flush rate oxygen source. The likelihood of CEW-caused ignition in such settings with supplemental oxygen in use is extremely low.

Sections du résumé

BACKGROUND BACKGROUND
Conducted electrical weapons (CEWs) are nonlethal weapons used in tactical environments. Tactical EMS (TEMS) operators provide patient care in environments where CEWs are present. CEWs may produce a spark that can be a source of ignition. When supplemental oxygen is in use by a TEMS operator, the flammability risk from the CEW is unknown.
METHODS METHODS
We measured oxygen levels over 20 minutes in an enclosed space with a supplemental oxygen source maximally flowing. Measurements were taken at various distances from the oxygen source to establish baseline concentrations. These concentrations were replicated within a plexiglass box where a CEW was activated to create a sparking arc between probes embedded in a fresh swine shoulder (with skin intact). Various flammable materials, typically found in a patient care setting were used as potential fuel sources.
RESULTS RESULTS
The highest oxygen levels were achieved directly at the source. At 15cm from the source, the maximum oxygen concentration was 31.5%. Within the box, ignition was only achieved at oxygen concentrations greater than 45% and only when human hair was present as fuel. No ignition was achieved at oxygen levels below this regardless of the tested fuel present.
CONCLUSION CONCLUSIONS
Ignition from a CEW is possible at supplemental oxygen levels greater than 45% when human hair is present. In an enclosed space, oxygen concentration levels of 45% are only present within 15cm of a flush rate oxygen source. The likelihood of CEW-caused ignition in such settings with supplemental oxygen in use is extremely low.