Validation of Acuros for total body irradiation at extended distance.
Acuros
Eclipse
TBI dose validation
extended distance
Journal
Journal of applied clinical medical physics
ISSN: 1526-9914
Titre abrégé: J Appl Clin Med Phys
Pays: United States
ID NLM: 101089176
Informations de publication
Date de publication:
18 Jul 2024
18 Jul 2024
Historique:
revised:
23
05
2024
received:
02
04
2024
accepted:
20
06
2024
medline:
18
7
2024
pubmed:
18
7
2024
entrez:
18
7
2024
Statut:
aheadofprint
Résumé
Standardized and accurately reported doses are essential in conventional total body irradiation (TBI), especially lung doses. This study evaluates the accuracy of the Acuros algorithm in predicting doses for extended-distance TBI. Measurements and calculations were done with both 6 and 18 MV. Tissue Maximum Ratio (TMR), output and off axis ratios (OAR) were measured at 200 and 500 cm source to detector distance and compared to Acuros calculated values. Two end-to-end tests were carried out, one with an in-house phantom (solid water and Styrofoam) with inserted ion chambers and the other was with the Imaging and Radiation Oncology Core (IROC) TBI anthropomorphic phantom equipped with TLDs. The end-to-end test was done for 6 and 18 MV both with and without lung blocks. The source to midplane distance for both phantoms were at 518 and 508 cm respectively. Lung blocks were placed at the phantom surface and a beam spoiler was positioned 30 cm from the surface of the phantoms as per our clinical set up. The agreement between measured and calculated TMR, output and off axis ratios for both 6 and 18 MV were within 2%. Ion chamber measurements in both the Styrofoam and solid water for both energies carried out with and without lung blocks were within 2% of calculated values. TLD measured doses for both 6 and 18 MV in the IROC phantom were within 5% of calculated doses which is within the uncertainty of the TLD measurement. The results indicate that the clinical beam model for Acuros 16.1 commissioned at standard clinical distances is capable of calculating doses accurately at extended distances up to 500 cm.
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
e14468Informations de copyright
© 2024 The Author(s). Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.
Références
Shinde A, Yang D, Frankel P, et al. Radiation‐related toxicities using organ sparing total marrow irradiation transplant conditioning regimens. Int J Radiat Oncol Biol Phys. 2019;105(5):1025‐1033. doi: 10.1016/j.ijrobp.2019.08.010
Vogel J, Hui S, Hua CH, et al. Pulmonary toxicity after total body irradiation—critical review of the literature and recommendations for toxicity reporting. Front Oncol. 2021;11:708906. doi: 10.3389/fonc.2021.708906
Sampath S, Schultheiss TE, Wong J. Dose response and factors related to interstitial pneumonitis after bone marrow transplant. Int J Radiat Oncol Biol Phys. 2005;63(3):876‐884.
Fog LS, Wirth A, MacManus M, et al. Total body irradiation in Australia and New Zealand: results of a practice survey. Phys Eng Sci Med. 2020;43:825‐835.
Hoeben BAW, Pazos M, Albert MH, et al. Towards homogenization of total body irradiation practices in pediatric patients across SIOPE affiliated centers. A survey by the SIOPE radiation oncology working group. Radiother Oncol. 2021;155:113‐119. doi: 10.1016/j.radonc.2020.10.032
Wong JYC, Filippi AR, Dabaja BS, Yahalom J, Specht L. Total body irradiation: guidelines from the International Lymphoma Radiation Oncology Group (ILROG). Int J Radiat Oncol Biol Phys. 2018;101(3):521‐529. doi: 10.1016/j.ijrobp.2018.04.071
Esiashvili N, Lu X, Ulin K, et al. Higher reported lung dose received during total body irradiation for allogeneic hematopoietic stem cell transplantation in children with acute lymphoblastic leukemia is associated with inferior survival: a report from the children's oncology group. Int J Radiat Oncol Biol Phys. 2019;104(3):513‐521. doi: 10.1016/j.ijrobp.2019.02.034
Rassiah P, Esiashvili N, Olch AJ, et al. Practice patterns of pediatric total body irradiation techniques: a children's oncology group survey. Int J Radiat Oncol Biol Phys. 2021;111(5):1155‐1164.
Marks LB, Bentzen SM, Deasy JO, et al. Radiation dose‐volume effects in the lung. Int J Radiat Oncol Biol Phys. 2010;76(Suppl 3):S70‐S76. doi: 10.1016/j.ijrobp.2009.06.091
Bailey DW, Wang IZ, Lakeman T, Hales LD, Singh AK, Podgorsak MB. TBI lung dose comparisons using bilateral and anteroposterior delivery techniques and tissue density corrections. J Appl Clin Med Phys. 2015;16(2):291‐301.
Lavallee MC, Gingras L, Chretien M, Aubin S, Cote C, Beaulieu L. Commissioning and evaluation of an extended SSD photon model for PINNACLE3: an application to total body irradiation. Med Phys. 2009;36(8):3844‐3855. doi: 10.1118/1.3171688
Sarfaraz M, Yu C, Chen D, Der L. A translational couch technique for total body irradiation. J Appl Clin Med Phys. 2001;2(4):201‐209.
Luk SMH, Wallner K, Glenn MC, et al. Effect of total body irradiation lung block parameters on lung doses using three‐dimensional dosimetry. J Appl Clin Med Phys. 2022;23(4):e13513. doi: 10.1002/acm2.13513
Hui SK, Das R, Thomadsen B, Henderson D. CT‐based analysis of dose homogeneity in total body irradiation using lateral beam. J Appl Clin Med Phys. 2004;5(4):71‐79.
Hussain A, Villarreal‐Barajas E, Brown D, Dunscombe P. Validation of the Eclipse AAA algorithm at extended SSD. J Appl Clin Med Phys. 2010;11(3):90‐100.
Lamichhane N, Studenski MT. Improving TBI lung dose calculations: can the treatment planning system help? Med Dosim. 2020;45(2):168‐171.
Lamichhane N, Patel VN, Studenski MT. Going the distance: validation of Acuros and AAA at an extended SSD of 400 cm. J Appl Clin Med Phys. 2016;17(2):63‐73. doi: 10.1120/jacmp.v17i2.5913
Giebel S, Miszczyk L, Slosarek K, et al. Extreme heterogeneity of myeloablative total body irradiation techniques in clinical practice: a survey of the acute leukemia working party of the european group for blood and marrow transplantation. Cancer. 2014;120(17):2760‐2765. doi: 10.1002/cncr.28768
Deye J, Rodgers J, Wu R. NCRP Report No. 151: Structural Shielding Design and Evaluation for Megavoltage X‐and gamma‐ray Radiotherapy Facilities. National Council on Radiation Protection; 2005.