Quality assurance for cancer patient safety: Clinical assessment for precise angles in linac during radiation therapy.

Medical physics SBRT Winston-Lutz cancer treatment machine learning quality assurance radiation therapy

Journal

Tumori
ISSN: 2038-2529
Titre abrégé: Tumori
Pays: United States
ID NLM: 0111356

Informations de publication

Date de publication:
02 Aug 2024
Historique:
medline: 3 8 2024
pubmed: 3 8 2024
entrez: 3 8 2024
Statut: aheadofprint

Résumé

Quality assurance for stereotactic body radiation treatment requires that isocentric verification be ensured during gantry rotation at various angles. This study examined statistical parameters on Winston-Lutz tests to distinguish the deviation of angles from isocenter during gantry rotation using machine learning. The Varian TrueBeam linac was aligned with the marked lines on the Ruby phantom. Eight images were captured while the gantry was rotating at a 45° shift. The statistical features were derived from IsoCheck EPID software. The decision tree model was applied to these Winston-Lutz tests to cluster data into two groups: precise and error angles. At 90° and 270° angles, the gantry exhibits isocentric stability compared to other angles. In these angles, the most statistical features were inside the range. Most variations were observed at 0° and 180° angles. In most tests, the angles 45°, 135°, 225°, and 315° showed reasonable performance and with less variation. The comprehensive statistical analyses for gantry rotation of angles assists expert radiotherapists in determining the contribution of each feature that highly affects gantry movement at specific angles. Misalignment between radiation isocenter and imaging isocenter, tuning of the beam at each angle, or a slight change in the position of the Ruby phantom can further improve the inaccuracy that causes the most variations. Better precision can effectively increase patient safety and quality during cancer treatment.

Identifiants

pubmed: 39096026
doi: 10.1177/03008916241261450
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

3008916241261450

Déclaration de conflit d'intérêts

Declaration of conflicting interestsThe author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Auteurs

Sana Salahuddin (S)

Institute of Physics, The Islamia University of Bahawalpur, Pakistan.
Department of Medical Physics, IRCCS Azienda Ospedaliero-Universitaria di Bologna, Italy.

Saeed Ahmad Buzdar (SA)

Institute of Physics, The Islamia University of Bahawalpur, Pakistan.

Khalid Iqbal (K)

Shaukat Khanum Memorial Cancer Hospital and Research Center, Lahore, Pakistan.

Muhammad Adeel Azam (MA)

Department of Advanced Robotics, Italian Institute of Technology (IIT), Genova, Italy.

Mamona Aslam (M)

Institute of Physics, The Islamia University of Bahawalpur, Pakistan.

Saima Altaf (S)

Institute of Physics, The Islamia University of Bahawalpur, Pakistan.

Ayesha Ikhlaq (A)

Institute of Physics, The Islamia University of Bahawalpur, Pakistan.

Muhammad Usman Mustafa (MU)

Institute of Physics, The Islamia University of Bahawalpur, Pakistan.

Lidia Strigari (L)

Department of Medical Physics, IRCCS Azienda Ospedaliero-Universitaria di Bologna, Italy.

Classifications MeSH