Interference of lupus anticoagulant causing antiprothrombin and anti-beta-2-glycoprotein I antibodies on international normalized ratio measurements: comparative analysis of international normalized ratio methods.
anticoagulants
antiphospholipid syndrome
international normalized ratio
lupus coagulation inhibitor
warfarin
Journal
Research and practice in thrombosis and haemostasis
ISSN: 2475-0379
Titre abrégé: Res Pract Thromb Haemost
Pays: United States
ID NLM: 101703775
Informations de publication
Date de publication:
Jul 2024
Jul 2024
Historique:
received:
16
04
2024
revised:
27
05
2024
accepted:
30
05
2024
medline:
7
8
2024
pubmed:
7
8
2024
entrez:
7
8
2024
Statut:
epublish
Résumé
Life-long vitamin K antagonist (VKA) therapy is recommended as a standard of care in antiphospholipid syndrome (APS) patients with thrombosis. Concerns have been raised about the validity of international normalized ratio (INR) measurements in lupus anticoagulant (LA)-positive APS patients because LA may interfere with phospholipid-dependent coagulation tests and could elevate INR measurements. Here, we aimed to determine the interference of antigen-specific monoclonal and isolated patient antibodies with LA activity on INR measurements. Pooled normal plasma and control plasma from patients on VKA (without LA) were incubated with monoclonal and isolated patient immunoglobulin G antiprothrombin and anti-beta-2-glycoprotein I antibodies that express LA activity. INR was determined before and after addition using 3 laboratory assays (Owren STA-Hepato Prest, Quick STA-NeoPTimal, and Quick STA-Neoplastine R) and 1 point-of-care test device (CoaguChek Pro II). Antiprothrombin and anti-beta-2-glycoprotein I antibodies with LA activity interfered with recombinant human thromboplastin reagents (Quick STA-Neoplastine R and CoaguChek Pro II), particularly when added to plasma of VKA-treated controls. This effect was most evident on point-of-care test INR measurements, while the recombinant Quick reagent exhibited a lesser degree of interference. In contrast, tissue-derived thromboplastin reagents (Owren STA-Hepato Prest and Quick STA-NeoPTimal) remained largely unaffected by these antibodies, both in pooled normal plasma and VKA anticoagulated control plasma. Among these reagents, the Owren INR reagent exhibited the lowest sensitivity to the influence of LA antibodies. This observed difference in sensitivity is independent of the plasma dilution factor or the presence of factor V or fibrinogen in Owren reagent. INR reagents that utilize recombinant human thromboplastin are more sensitive to the presence of monoclonal and patient-derived antibodies with LA activity. Consequently, APS patients positive for LA should be monitored using tissue-derived thromboplastin reagents, given its reduced susceptibility to interference by LA-causing antibodies.
Sections du résumé
Background
UNASSIGNED
Life-long vitamin K antagonist (VKA) therapy is recommended as a standard of care in antiphospholipid syndrome (APS) patients with thrombosis. Concerns have been raised about the validity of international normalized ratio (INR) measurements in lupus anticoagulant (LA)-positive APS patients because LA may interfere with phospholipid-dependent coagulation tests and could elevate INR measurements.
Objectives
UNASSIGNED
Here, we aimed to determine the interference of antigen-specific monoclonal and isolated patient antibodies with LA activity on INR measurements.
Methods
UNASSIGNED
Pooled normal plasma and control plasma from patients on VKA (without LA) were incubated with monoclonal and isolated patient immunoglobulin G antiprothrombin and anti-beta-2-glycoprotein I antibodies that express LA activity. INR was determined before and after addition using 3 laboratory assays (Owren STA-Hepato Prest, Quick STA-NeoPTimal, and Quick STA-Neoplastine R) and 1 point-of-care test device (CoaguChek Pro II).
Results
UNASSIGNED
Antiprothrombin and anti-beta-2-glycoprotein I antibodies with LA activity interfered with recombinant human thromboplastin reagents (Quick STA-Neoplastine R and CoaguChek Pro II), particularly when added to plasma of VKA-treated controls. This effect was most evident on point-of-care test INR measurements, while the recombinant Quick reagent exhibited a lesser degree of interference. In contrast, tissue-derived thromboplastin reagents (Owren STA-Hepato Prest and Quick STA-NeoPTimal) remained largely unaffected by these antibodies, both in pooled normal plasma and VKA anticoagulated control plasma. Among these reagents, the Owren INR reagent exhibited the lowest sensitivity to the influence of LA antibodies. This observed difference in sensitivity is independent of the plasma dilution factor or the presence of factor V or fibrinogen in Owren reagent.
Conclusion
UNASSIGNED
INR reagents that utilize recombinant human thromboplastin are more sensitive to the presence of monoclonal and patient-derived antibodies with LA activity. Consequently, APS patients positive for LA should be monitored using tissue-derived thromboplastin reagents, given its reduced susceptibility to interference by LA-causing antibodies.