Relative profile measurements in 1.5T MR-linacs: investigation of central axis deviations.

MR-linac Unity effective point of measurement relative dosimetry

Journal

Physics in medicine and biology
ISSN: 1361-6560
Titre abrégé: Phys Med Biol
Pays: England
ID NLM: 0401220

Informations de publication

Date de publication:
13 Aug 2024
Historique:
medline: 14 8 2024
pubmed: 14 8 2024
entrez: 13 8 2024
Statut: aheadofprint

Résumé

In 1.5T MR-linacs, the absorbed dose central axis (CAX) deviates from the beam's CAX due to inherent profile asymmetry. In addition, a measured CAX deviation may be biased due to potential lateral (to the beam) effective point of measurement (EPOM

Identifiants

pubmed: 39137816
doi: 10.1088/1361-6560/ad6ed7
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Informations de copyright

Creative Commons Attribution license.

Auteurs

Anastasios Episkopakis (A)

Medical Physics Laboratory, Medical School, National and Kapodistrian University of Athens, Mikras Asias 75, Goudi, Athens, 11527, GREECE.

Vasiliki Margaroni (V)

Medical Physics Laboratory, Medical School, National and Kapodistrian University of Athens, Mikras Asias 75, Goudi, Athens, 11527, GREECE.

Pantelis Karaiskos (P)

Medical Physics Laboratory, Medical School, National and Kapodistrian University of Athens, Mikras Asias 75, Goudi, Athens, 11527, GREECE.

Efi Koutsouveli (E)

Medical Physics Department, Hygeia Hospital, Kifisias Avenue and 4 Erythrou Stavrou, Maroussi, Athens, 15123, GREECE.

Nikolas Marinos (N)

Global Clinical Operations, Elekta Ltd, Fleming Way, Crawley, West Sussex, RH10 9RR, UNITED KINGDOM OF GREAT BRITAIN AND NORTHERN IRELAND.

Eleftherios P Pappas (EP)

Medical Physics Laboratory, Medical School, National and Kapodistrian University of Athens, 75, Mikras Asias, Athens, 11527, GREECE.

Classifications MeSH