HeartMate 3 driveline damage by gradual corrosion due to liquid infiltration: a case report.

Driveline damage Durable LVAD Heartmate 3 Liquid infiltration

Journal

Journal of artificial organs : the official journal of the Japanese Society for Artificial Organs
ISSN: 1619-0904
Titre abrégé: J Artif Organs
Pays: Japan
ID NLM: 9815648

Informations de publication

Date de publication:
27 Aug 2024
Historique:
received: 27 02 2024
accepted: 29 07 2024
medline: 27 8 2024
pubmed: 27 8 2024
entrez: 27 8 2024
Statut: aheadofprint

Résumé

A 31-year-old man with a HeartMate 3 was admitted with a "Driveline Communication Fault" alarm due to liquid infiltration. Eleven months earlier, the connector between the modular and pump cables had gotten wet when he was taking a shower. The cable connector was dried immediately, and no alarm had been observed during follow-up. Subsequently, the modular cable and system controller were replaced, with corrosion found on the modular cable connector. The "Communication Fault" alarm recurred, and complete damage to the communication cables was discovered. The pump was replaced, and the removed pump cable connector showed corrosion as well. If the driveline connector gets wet, the multidisciplinary team should discuss if it should be immediately disconnected and dried, averting the need for future pump replacements due to corrosion.

Identifiants

pubmed: 39190056
doi: 10.1007/s10047-024-01464-w
pii: 10.1007/s10047-024-01464-w
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Informations de copyright

© 2024. The Japanese Society for Artificial Organs.

Références

Merzah AS, Hanke JS, Li T, Merzah FAS, Gabriel M, Derda AA, et al. Outcomes of modular cable exchange in heartmate 3 patients versus conventional driveline repair in heartmate II patients. Artif Organs. 2023;47:380–6.
doi: 10.1111/aor.14413 pubmed: 36148849
Yuzefpolskaya M, Schroeder SE, Houston BA, Robinson MR, Gosev I, Reyentovich A, et al. The society of thoracic surgeons intermacs 2022 annual report: focus on the 2018 heart transplant allocation system. Ann Thorac Surg. 2023;115:311–28.
doi: 10.1016/j.athoracsur.2022.11.023 pubmed: 36462544
Kormos RL, McCall M, Althouse A, Lagazzi L, Schaub R, Kormos MA, et al. Left ventricular assist device malfunctions: it is more than just the pump. Circulation. 2017;136:1714–25.
doi: 10.1161/CIRCULATIONAHA.117.027360 pubmed: 28674109
D’Antonio ND, Maynes EJ, Tatum RT, Prochno KW, Saxena A, Maltais S, et al. Driveline damage and repair in continuous-flow left ventricular assist devices: a systematic review. Artif Organs. 2021;45:819–26.
doi: 10.1111/aor.13901 pubmed: 33377216

Auteurs

Tomo Yoshizumi (T)

Department of Cardiac Surgery, Nagoya University Graduate School of Medicine, 65 Tsurumai, Showa-Ku, Nagoya, 466-8550, Japan. yoshizumi.tomo.s4@f.mail.nagoya-u.ac.jp.

Hideki Ito (H)

Department of Cardiac Surgery, Nagoya University Graduate School of Medicine, 65 Tsurumai, Showa-Ku, Nagoya, 466-8550, Japan.

Sachie Terazawa (S)

Department of Cardiac Surgery, Nagoya University Graduate School of Medicine, 65 Tsurumai, Showa-Ku, Nagoya, 466-8550, Japan.

Yoshiyuki Tokuda (Y)

Department of Cardiac Surgery, Nagoya University Graduate School of Medicine, 65 Tsurumai, Showa-Ku, Nagoya, 466-8550, Japan.

Hajime Sakurai (H)

Department of Cardiac Surgery, Nagoya University Graduate School of Medicine, 65 Tsurumai, Showa-Ku, Nagoya, 466-8550, Japan.

Yuji Narita (Y)

Department of Cardiac Surgery, Nagoya University Graduate School of Medicine, 65 Tsurumai, Showa-Ku, Nagoya, 466-8550, Japan.

Masato Mutsuga (M)

Department of Cardiac Surgery, Nagoya University Graduate School of Medicine, 65 Tsurumai, Showa-Ku, Nagoya, 466-8550, Japan. mutsuga.masato.f5@f.mail.nagoya-u.ac.jp.

Classifications MeSH