Disintegration of Multiple Parts of a Bone Transport Nail During Planned Removal: A Case Report.


Journal

JBJS case connector
ISSN: 2160-3251
Titre abrégé: JBJS Case Connect
Pays: United States
ID NLM: 101596828

Informations de publication

Date de publication:
01 Jul 2024
Historique:
medline: 13 9 2024
pubmed: 13 9 2024
entrez: 13 9 2024
Statut: epublish

Résumé

A 34-year-old man with a distal tibia bone defect was treated with an intramedullary bone transport nail (Precice Bone Transport System, NuVasive). During planned removal after successful treatment, 7 separate subcomponents of the nail became disconnected and had to be separately removed using specialized instrumentation. This occurred despite adherence to the manufacturer's recommended technique for nail removal and in the absence of clinical or radiographic evidence of implant failure. When planning for implant removal, surgeons should be aware of potential intraoperative disconnection of subcomponents of this magnetic bone transport nail and ensure that equipment for retrieval (e.g., very long endoscopy forceps) is available.

Identifiants

pubmed: 39270038
doi: 10.2106/JBJS.CC.24.00159
pii: 01709767-202409000-00055
doi:

Types de publication

Case Reports Journal Article

Langues

eng

Sous-ensembles de citation

IM

Informations de copyright

Copyright © 2024 by The Journal of Bone and Joint Surgery, Incorporated.

Déclaration de conflit d'intérêts

Financial Disclosure Statement: No author associated with this article has any relevant conflicts of interest to report. Disclosure: The Disclosure of Potential Conflicts of Interest forms are provided with the online version of the article (http://links.lww.com/JBJSCC/C439).

Références

Paley D, Herzenberg JE, Paremain G, Bhave A. Femoral lengthening over an intramedullary nail. A matched-case comparison with Ilizarov femoral lengthening. J Bone Joint Surg Am. 1997;79(10):1464-80.
Kocaoglu M, Eralp L, Kilicoglu O, Burc H, Cakmak M. Complications encountered during lengthening over an intramedullary nail. J Bone Joint Surg Am. 2004;86(11):2406-11.
Frost MW, Rahbek O, Traerup J, Ceccotti AA, Kold S. Systematic review of complications with externally controlled motorized intramedullary bone lengthening nails (FITBONE and PRECICE) in 983 segments. Acta Orthop. 2021;92(1):120-7.
Noonan KJ, Leyes M, Forriol F, Cañadell J. Distraction osteogenesis of the lower extremity with use of monolateral external fixation. A study of two hundred and sixty-one femora and tibiae. J Bone Joint Surg Am. 1998;80(6):793-806.
Papakostidis C, Bhandari M, Giannoudis PV. Distraction osteogenesis in the treatment of long bone defects of the lower limbs: effectiveness, complications and clinical results; a systematic review and meta-analysis. Bone Joint J. 2013;95-B(12):1673-80.
Ilizarov GA. The tension-stress effect on the genesis and growth of tissues: Part I. The influence of stability of fixation and soft-tissue preservation. Clin Orthop Relat Res. 1989;238:249-81.
Hlukha LP, Alrabai HM, Sax OC, Hammouda AI, McClure PK, Herzenberg JE. Mechanical failures in magnetic intramedullary lengthening nails. J bone Joint Surg Am.2023;105(2):113-27.
Paley D. PRECICE intramedullary limb lengthening system. Expert Rev Med Devices. 2015;12(3):231-49.
Health C for D and R. UPDATE: NuVasive Specialized Orthopedics' Precice Devices - Letter to Health Care Providers. Washington, DC: FDA; 2023. Available at: https://www.fda.gov/medical-devices/letters-health-care-providers/update-nuvasive-specialized-orthopedics-precice-devices-letter-health-care-providers. Accessed February 20, 2024.
Precice Bone Transport Tibia Surgical Technique Guide, Washington, DC, USA 2024. Available at: https://www.nuvasive.com/wp-content/uploads/2020/09/PBT-ST-Tibia.pdf. Accessed February 20, 2024.
Precice Bone Transport System. NuVasive; 2023. Available at: https://www.nuvasive.com/precice-bone-transport/. Accessed March 16, 2024.
Panagiotopoulou VC, Davda K, Hothi HS, Henckel J, Cerquiglini A, Goodier WD, Skinner J, Hart A, Calder PR. A retrieval analysis of the Precice intramedullary limb lengthening system. Bone Joint Res. 2018;7(7):476-84.
Tiefenböck TM, Wozasek GE. Unusual complication with an intramedullary lengthening device 15 months after implantation. Injury. 2015;46(10):2069-72.
Hidden KA, Dahl MT, Ly TV. Management of a broken PRECICE femoral nail at an ununited distraction osteogenesis site: a case report. JBJS Case Connect. 2020;10(1):e0267.
Johnson MA, Karkenny AJ, Arkader A, Davidson RS. Dissociation of a femoral intramedullary magnetic lengthening nail during routine hardware removal: a case report. JBJS Case Connect. 2021;11(1):e20.00950.
Nuvasive. Urgent Recall Notification PRECICE Stryde, Precise Plate and PRECICE Bone Transport; 2021. Available at: https://www.nuvasive.com/wp-content/uploads/2021/02/NSO-Precice-FSN-United-States-Biodur.pdf. Accessed February 20, 2024.

Auteurs

Ryan Halvorson (R)

Department of Orthopaedic Surgery, University of California, San Francisco, California.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs