The Use of Dynamic Navigation for the Placement of Zygoma Implants. A Cadaver Study to Compare Accuracy.
Dynamic Navigation
Guided implant surgery
Maxillofacial implants
Zygoma implants
Journal
The Journal of oral implantology
ISSN: 1548-1336
Titre abrégé: J Oral Implantol
Pays: United States
ID NLM: 7801086
Informations de publication
Date de publication:
18 Oct 2024
18 Oct 2024
Historique:
medline:
18
10
2024
pubmed:
18
10
2024
entrez:
18
10
2024
Statut:
aheadofprint
Résumé
This study aimed to compare the angular, platform, and apical deviation of zygoma implants placed with the aid of a dynamic navigation device compared to the implants placed freehand. Eight cadaver heads were used for the study. Pre-operative CBCT scans were made for the heads, and an implant planning software was used to plan zygoma implants bilaterally in each scanned head. A total of 16 zygoma implants were placed using each surgical technique. Post-implant CBCT scans were merged with the pre-operative plan to evaluate and compare the accuracy of the implants using each technique. Angular and linear deviations of the implants were measured and compared. The effect of implant length and position on the deviation was also investigated. The results showed a mean angular deviation of 2.44 ± 1.10 degrees for implants placed using DN compared to 6.63 ± 4.81degrees for implants placed freehand. The mean apical deviation for implants placed using DN and for implants placed freehand was 2.14 ± 1.02 mm and 3.80 ± 2.49 mm, respectively. Finally, the mean platform deviation for implants placed using DN and for implants placed freehand was 1.66 ± 0.9 mm and 2.81 ± 2.1 mm, respectively. The analysis showed a significant difference between the two angular and apical deviation techniques. ( p=0.001 and 0.029 respectively). Zygoma implants placed with a dynamic navigation device resulted in less apical and angular deviation than implants placed freehand..
Identifiants
pubmed: 39423334
pii: 503615
doi: 10.1563/10.1563_aaid-joi-D-24-00026
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Déclaration de conflit d'intérêts
Authors report no conflict of interest reported by the authors.