Electromagnetic navigation system for computed tomography-guided synchronous percutaneous lung biopsy and microwave ablation of pulmonary nodules: a prospective, single-center, single-arm clinical study.
CT-guided interventions
Electromagnetic navigation
microwave ablation
percutaneous lung biopsy
pulmonary nodule
Journal
International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group
ISSN: 1464-5157
Titre abrégé: Int J Hyperthermia
Pays: England
ID NLM: 8508395
Informations de publication
Date de publication:
2024
2024
Historique:
medline:
28
10
2024
pubmed:
28
10
2024
entrez:
27
10
2024
Statut:
ppublish
Résumé
The purpose of this study was to clinically evaluate the safety and effectiveness of the electromagnetic navigation (EMN) system designed for computed tomography (CT)-guided synchronous percutaneous lung biopsy and microwave ablation (MWA) of pulmonary nodules. This prospective, single-center, single-arm clinical cohort study was conducted in Beijing Hospital from March 2023 to May 2023. Patients who underwent CT-guided synchronous percutaneous lung biopsy and MWA via the EMN system were prospectively enrolled in our study. All the interventional procedures were performed by the same interventional radiologist. The technical success rate, the technical efficacy rates of biopsy and MWA were assessed as the primary outcomes. Preoperative, intraoperative, and postoperative variables were also recorded and analyzed for each patient. A total of 48 patients were enrolled in the study. The technical success rate was 100%. The technical efficacy rate of biopsy was 95.8% (46/48), and the technical efficacy rate of WMA was 100% (48/48) with no recurrence during follow-up. The total and subpleural needle trajectory length and distance error were 8.3 ± 2.6 cm, 3.6 ± 1.6 cm, and 1.84 ± 1.08 mm, respectively. The median numbers of needle adjustments and CT acquisitions were 1 (range 1-3) and 3 (range 3-5), respectively. The time to reach the target and procedure time were 4.4 ± 1.7 and 19.7 ± 5.2 min, respectively. The dose length product was 748.8 ± 221.8 mGy*cm. The median postoperative hospital stay was 1 (range 1-7) days. No major complications (grade ≥3) occurred and only seven minor complications (14.6%) occurred, including six cases of pneumothorax and one case of hemoptysis. The radiologists achieved high satisfaction scores after surgery. The EMN system is feasible, safe and effective for CT-guided synchronous percutaneous lung biopsy and MWA of pulmonary nodules.
Sections du résumé
BACKGROUND
UNASSIGNED
The purpose of this study was to clinically evaluate the safety and effectiveness of the electromagnetic navigation (EMN) system designed for computed tomography (CT)-guided synchronous percutaneous lung biopsy and microwave ablation (MWA) of pulmonary nodules.
METHODS
UNASSIGNED
This prospective, single-center, single-arm clinical cohort study was conducted in Beijing Hospital from March 2023 to May 2023. Patients who underwent CT-guided synchronous percutaneous lung biopsy and MWA via the EMN system were prospectively enrolled in our study. All the interventional procedures were performed by the same interventional radiologist. The technical success rate, the technical efficacy rates of biopsy and MWA were assessed as the primary outcomes. Preoperative, intraoperative, and postoperative variables were also recorded and analyzed for each patient.
RESULTS
UNASSIGNED
A total of 48 patients were enrolled in the study. The technical success rate was 100%. The technical efficacy rate of biopsy was 95.8% (46/48), and the technical efficacy rate of WMA was 100% (48/48) with no recurrence during follow-up. The total and subpleural needle trajectory length and distance error were 8.3 ± 2.6 cm, 3.6 ± 1.6 cm, and 1.84 ± 1.08 mm, respectively. The median numbers of needle adjustments and CT acquisitions were 1 (range 1-3) and 3 (range 3-5), respectively. The time to reach the target and procedure time were 4.4 ± 1.7 and 19.7 ± 5.2 min, respectively. The dose length product was 748.8 ± 221.8 mGy*cm. The median postoperative hospital stay was 1 (range 1-7) days. No major complications (grade ≥3) occurred and only seven minor complications (14.6%) occurred, including six cases of pneumothorax and one case of hemoptysis. The radiologists achieved high satisfaction scores after surgery.
CONCLUSION
UNASSIGNED
The EMN system is feasible, safe and effective for CT-guided synchronous percutaneous lung biopsy and MWA of pulmonary nodules.