To determine the key image anatomical parameters that are relevant to cochlear implantation (CI) using temporal bone high-resolution computed tomography (HRCT) scans and to identify age group differen...
The data of 346 temporal bone HRCTs of 173 children from 5 months to 18 years of age were retrospectively selected and reviewed. Parameters of the human temporal bone that are relevant to CI key surgi...
There was no significant difference in the above morphological values by gender (p > 0,05). Two-sided FRW was not significantly different (p > 0,05), but MT and angle A were significantly different (p...
Based on preoperative imaging analysis at different ages, middle ear development was stable at 5 months of age, allowing early CI in infancy with severe to severe hearing loss at this age. These data ...
Evaluate the rate of preserved vestibular function in pediatric cochlear implant surgery....
Retrospective case review....
Pre- and post-operative vestibular tests were compared in children who underwent cochlear implantation at a tertiary level pediatric hospital over a 4-year period....
Data from 59 implanted ears in 44 children was included. Median age was 2.8 years at initial testing (range 7 months - 21 years) with 1:1 male/female ratio. Implant surgeries were 26 unilateral, 13 bi...
Preservation rates of vestibular function following cochlear implant surgery were higher in this cohort than what has been reported in many earlier studies. Contemporary, less traumatic electrodes and...
4, Case Series Laryngoscope, 134:1913-1918, 2024....
Cochlear ossification (CO) after meningitis can cause profound sensorineural hearing loss (SNHL). Cochlear implantation (CI) is the ideal treatment strategy for CO....
To explore the strategy for CI in patients with CO after meningitis....
In this retrospective study, the medical records of patients diagnosed with profound SNHL due to CO after meningitis and who underwent CI in our department between September 2010 and September 2021 we...
The data of 26 patients with unilateral CI were reviewed. All patients underwent preoperative temporal high-resolution computed tomography (HRCT) and 22 patients magnetic resonance imaging (MRI). The ...
Preoperative temporal bone HRCT and MRI are essential for determining whether a patient is suitable for CI and surgical planning. A false-negative diagnosis is possible when diagnosing CO, but combine...
The present study describes the treatment of patients at a tertiary institution who experienced device failure after Cochlear Implantation (CI), as well as identifying prodromic symptoms that could as...
Retrospective database review (January 2000-May 2017)....
Single tertiary hospital....
Factors recorded included the etiology of hearing loss; age at first and revision CI surgeries; surgical information, including operation time and approach; electrical outcomes after implantation; dev...
From January 2000 to May 2017, 1431 CIs were performed, with 27 (1.9%) undergoing revision surgeries due to device failure. The most common etiology of hearing loss was idiopathic (12/27), followed by...
The most suggestive symptom preceding device failure was intermittent loss of signal. Patients who present with this symptom should undergo electrical examination for suspected device failure. Audiolo...
Cochlear implantation (CI) for deafblindness may have more impact than for non-syndromic hearing loss. Deafblind patients have a double handicap in a society that is more and more empowered by fast co...
In February 2020, Advanced Bionics initiated a field action notice to remove the MS Ultra (V1) implant from circulation. In this study, we quantify a single site's incidence with device failure and ex...
Retrospective/cohort study....
Tertiary healthcare center....
Forty-nine adult patients (52 devices) were implanted between October 2017 and December 2019, with the following exclusion criteria: <18 years of age, medical/surgical failures, and lack of English-la...
Diagnostic....
Postoperative speech perception (AzBio sentence test) scores at 12 months postactivation were compared with repeat testing at 6- to 12-month intervals. Degree of change in impedances from 1 month post...
To date, 21% (11/52) of devices have met all three criteria for a confirmed failure with an average decline in speech perception of 27.7 ± 22%. The average length of time for detection of device failu...
Impedance values can be used as a reliable indicator of device malfunction, thus facilitating patient counseling and early intervention. No correlation was identified between impedance changes and spe...
To assess the distance burden for access to cochlear implant (CI)-related services and to assess whether socioeconomic disadvantage or level of education and occupation influenced uptake of CIs....
Retrospective case review....
A CI services provider operating across multiple centers....
All patients undergoing CI surgery in a 2-year period between March 2018 and February 2020....
Diagnosis of hearing loss, CI surgery, and subsequent habilitation and mapping....
Distance traveled by patients to their audiological diagnostic, CI surgery hospital, and habilitation sites; subjects' index of relative socioeconomic advantage and disadvantage (IRSAD) and index of e...
n = 201 children and n = 623 adults. There was a significant difference across IRSAD domains for children (p < 0.0001) and adults (p < 0.0001), and IEO in children (p = 0.015) and adults (p < 0.0001) ...
The burden of distance for access to CI in Australia is significant for the upper quartile who may not live within the large city centers. Greater consideration needs to be given regarding barriers to...
This study aimed to describe the factors affecting early and late cochlear implantation....
A total of 159 patients from the Hospital Canselor Tuanku Muhriz (HCTM) Cochlear Implant Programme were recruited in this retrospective cross-sectional study. All paediatric Cochlear Implant (CI) reci...
A total of 83 (52%) patients were female. Chinese race constituted most of the patients, which was 90/159 (57%). The majority were from middle-income families (M40); 89 (56%). The most common aetiolog...
UNHS was a significant factor contributing to early and late implantation. The median age of diagnosis of hearing loss was 18 months (interquartile range; 15); the age of CI was 34 months (interquarti...
A retrospective cross-sectional analysis was conducted of the US Food and Drug Administration's MAUDE (Manufacturer and User Facility Device Experience) database, to evaluate the complication profile ...
A review of the MAUDE database was conducted from 1 January 2010 to 31 December 2020. Complications, including infection, extrusion, facial nerve stimulation, meningitis and cerebrospinal fluid leak, ...
A total of 31 857 adverse events were analysed. Implants of manufacturer C were associated with a statistically higher rate of infection (0.97 per cent), cerebrospinal fluid leak (0.07 per cent), extr...
Consideration of patient risk factors along with cochlear implant manufacturers can heighten awareness of cochlear implant complications pre-operatively, intra-operatively and post-operatively....
To date, auditory rehabilitation mainly focuses on the person with hearing impairment (PHI). This study aimed to analyse the burden of hearing loss on significant others (SOs), and to explore the impa...
Third-party disability (SOS-HEAR) and quality of life (Nijmegen Cochlear Implant Questionnaire) were evaluated in 41 PHI scheduled for CI surgery and their close partners pre- and 6-month post-implant...
Hearing loss imposes a burden on SOs, particularly in relation to changes in communication and socialisation. Third-party disability was higher in SOs of PHI with lower educational background (...
Audiological rehabilitation should expand to include SOs in the rehabilitation process, as the burden experienced by SOs might persist even after CI....