Multi-institutional evaluation of end-to-end protocol for IMRT/VMAT treatment chains utilizing conventional linacs.
End-to-end test
IMRT
OSLD
Quality assurance
VMAT
Journal
Medical dosimetry : official journal of the American Association of Medical Dosimetrists
ISSN: 1873-4022
Titre abrégé: Med Dosim
Pays: United States
ID NLM: 8908862
Informations de publication
Date de publication:
Historique:
received:
18
01
2017
revised:
05
02
2018
accepted:
07
02
2018
pubmed:
25
3
2018
medline:
30
5
2019
entrez:
25
3
2018
Statut:
ppublish
Résumé
We conducted a multi-institutional assessment of a recently developed end-to-end monthly quality assurance (QA) protocol for external beam radiation therapy treatment chains. This protocol validates the entire treatment chain against a baseline to detect the presence of complex errors not easily found in standard component-based QA methods. Participating physicists from 3 institutions ran the end-to-end protocol on treatment chains that include Imaging and Radiation Oncology Core (IROC)-credentialed linacs. Results were analyzed in the form of American Association of Physicists in Medicine (AAPM) Task Group (TG)-119 so that they may be referenced by future test participants. Optically stimulated luminescent dosimeter (OSLD), EBT3 radiochromic film, and A1SL ion chamber readings were accumulated across 10 test runs. Confidence limits were calculated to determine where 95% of measurements should fall. From calculated confidence limits, 95% of measurements should be within 5% error for OSLDs, 4% error for ionization chambers, and 4% error for (96% relative gamma pass rate) radiochromic film at 3% agreement/3 mm distance to agreement. Data were separated by institution, model of linac, and treatment protocol (intensity-modulated radiation therapy [IMRT] vs volumetric modulated arc therapy [VMAT]). A total of 97% of OSLDs, 98% of ion chambers, and 93% of films were within the confidence limits; measurements were found outside these limits by a maximum of 4%, < 1%, and < 1%, respectively. Data were consistent despite institutional differences in OSLD reading equipment and radiochromic film calibration techniques. Results from this test may be used by clinics for data comparison. Areas of improvement were identified in the end-to-end protocol that can be implemented in an updated version. The consistency of our data demonstrates the reproducibility and ease-of-use of such tests and suggests a potential role for their use in broad end-to-end QA initiatives.
Identifiants
pubmed: 29571934
pii: S0958-3947(18)30020-7
doi: 10.1016/j.meddos.2018.02.002
pii:
doi:
Types de publication
Evaluation Study
Journal Article
Multicenter Study
Langues
eng
Sous-ensembles de citation
IM
Pagination
61-66Informations de copyright
Copyright © 2018 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.