Biodegradable-Polymer Versus Polymer-Free Drug-Eluting Stents for the Treatment of Coronary Artery Disease.
Absorbable Implants
Cause of Death
Coronary Artery Disease
/ diagnostic imaging
Coronary Thrombosis
/ mortality
Drug-Eluting Stents
Humans
Myocardial Infarction
/ mortality
Percutaneous Coronary Intervention
/ adverse effects
Polymers
Prosthesis Design
Randomized Controlled Trials as Topic
Risk Factors
Treatment Outcome
Biodegradable-polymer stent
Coronary artery disease
Drug-eluting stent
Percutaneous coronary intervention
Polymer-free stent
Journal
Cardiovascular revascularization medicine : including molecular interventions
ISSN: 1878-0938
Titre abrégé: Cardiovasc Revasc Med
Pays: United States
ID NLM: 101238551
Informations de publication
Date de publication:
10 2019
10 2019
Historique:
received:
07
11
2018
revised:
07
12
2018
accepted:
11
12
2018
pubmed:
24
12
2018
medline:
23
6
2020
entrez:
23
12
2018
Statut:
ppublish
Résumé
Biodegradable-polymer (BP) and polymer-free (PF) drug eluting stents (DES) were developed to reduce the risk of delayed arterial healing observed with durable-polymer (DP) platforms. Although trials demonstrate BP-DES and PF-DES are non-inferior to DP-DES, there is limited data directly comparing these technologies. We performed a meta-analysis to assess the efficacy and safety of BP-DES versus PF-DES for the treatment of coronary artery disease. Electronic searches were performed identifying randomized trials comparing BP-DES with PF-DES. Co-primary efficacy endpoints were target vessel revascularization (TVR), target lesion revascularization (TLR) and angiographic in-stent late lumen loss (LLL). Co-secondary safety endpoints were all-cause death, myocardial infarction (MI) and stent thrombosis (ST). Of 208 studies, 5 met inclusion criteria including 1975 patients. At mean follow-up (14 ± 5 months), BP-DES were associated with significantly reduced rates of TVR (OR 0.58, 95%CI 0.37-0.92, p = 0.02), TLR (4.7% vs 9.5%) (OR 0.48, 95%CI 0.31-0.75, p = 0.001) and in-stent LLL (pooled mean difference -0.20 mm, 95%CI -0.24 to -0.16, p < 0.001). There was no difference in safety, including all-cause death (OR 1.24, 95%CI 0.68-2.28, p = 0.48), MI (OR 0.92, 95%CI 0.54-1.56, p = 0.75) or ST (OR 1.58, 95%CI 0.67-3.73, p = 0.30). These data suggests that BP-DES are more efficacious when compared with PF-DES for the treatment of CAD.
Sections du résumé
BACKGROUND/PURPOSE
Biodegradable-polymer (BP) and polymer-free (PF) drug eluting stents (DES) were developed to reduce the risk of delayed arterial healing observed with durable-polymer (DP) platforms. Although trials demonstrate BP-DES and PF-DES are non-inferior to DP-DES, there is limited data directly comparing these technologies. We performed a meta-analysis to assess the efficacy and safety of BP-DES versus PF-DES for the treatment of coronary artery disease.
METHODS/MATERIALS
Electronic searches were performed identifying randomized trials comparing BP-DES with PF-DES. Co-primary efficacy endpoints were target vessel revascularization (TVR), target lesion revascularization (TLR) and angiographic in-stent late lumen loss (LLL). Co-secondary safety endpoints were all-cause death, myocardial infarction (MI) and stent thrombosis (ST).
RESULTS
Of 208 studies, 5 met inclusion criteria including 1975 patients. At mean follow-up (14 ± 5 months), BP-DES were associated with significantly reduced rates of TVR (OR 0.58, 95%CI 0.37-0.92, p = 0.02), TLR (4.7% vs 9.5%) (OR 0.48, 95%CI 0.31-0.75, p = 0.001) and in-stent LLL (pooled mean difference -0.20 mm, 95%CI -0.24 to -0.16, p < 0.001). There was no difference in safety, including all-cause death (OR 1.24, 95%CI 0.68-2.28, p = 0.48), MI (OR 0.92, 95%CI 0.54-1.56, p = 0.75) or ST (OR 1.58, 95%CI 0.67-3.73, p = 0.30).
CONCLUSIONS
These data suggests that BP-DES are more efficacious when compared with PF-DES for the treatment of CAD.
Identifiants
pubmed: 30578169
pii: S1553-8389(18)30613-4