LET response variability of Gafchromic
EBT3 film dosimetry
LET dependence
proton dosimetry
Journal
Medical physics
ISSN: 2473-4209
Titre abrégé: Med Phys
Pays: United States
ID NLM: 0425746
Informations de publication
Date de publication:
Jun 2019
Jun 2019
Historique:
received:
01
11
2018
revised:
01
11
2019
accepted:
02
02
2019
pubmed:
12
2
2019
medline:
23
11
2019
entrez:
12
2
2019
Statut:
ppublish
Résumé
To establish a method of accurate dosimetry required to quantify the expected linear energy transfer (LET) quenching effect of EBT3 film used to benchmark the dose distribution for a given treatment field and specified measurement depth. In order to facilitate this technique, a full analysis of film calibration which considers LET variability at the plane of measurement and as a function of proton beam quality is demonstrated. Additionally, the corresponding uncertainty from the process was quantified for several measurement scenarios. The net change in optical density (OD) from a single version of Gafchromic A strong LET dependence was observed near the Bragg peak (BP) consistent with previous studies performed with earlier versions of EBT3 film. The influence of range straggling on the film's LET response appears to have a uniform effect toward the BP regardless of the nominal beam energy. Proximal to this depth, the film's response decreased with decreasing energy at the same dose-average LET. The opposite trend was observed for depths past the BP. Changes in the SOBP energy modulation showed a linear relationship between the film's relative response and dose-averaged LET. Relative effectiveness factors (RE) were observed to range between 2%-7% depending on the width of the SOBP and depth of the film. Using the field-specific calibration technique, a total k = 1 uncertainty in the absorbed dose to water was estimated to range from 4.68%-5.21%. While EBT3 film's strong LET dependence is a common problem in proton beam dosimetry, this work has shown that the LET dependence can be taken into account by carefully considering the depth and energy modulation across the film using field-specific corrections. RE factors were determined with a combined k = 1 uncertainty of 3.57% for SOBP environments and between 3.17%-4.69% for uniform, monoenergetic fields proximal to the distal 80% of the BP.
Substances chimiques
Cobalt Radioisotopes
0
Protons
0
Cobalt-60
5C8182XDPZ
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
2716-2728Informations de copyright
© 2019 American Association of Physicists in Medicine.