Effect of computed tomography number-relative electron density conversion curve on the calculation of radiotherapy dose and evaluation of Monaco radiotherapy treatment planning system.


Journal

Australasian physical & engineering sciences in medicine
ISSN: 1879-5447
Titre abrégé: Australas Phys Eng Sci Med
Pays: Netherlands
ID NLM: 8208130

Informations de publication

Date de publication:
Jun 2019
Historique:
received: 26 05 2018
accepted: 27 02 2019
pubmed: 9 3 2019
medline: 18 12 2019
entrez: 9 3 2019
Statut: ppublish

Résumé

The accuracy of a computed tomography (CT)-relative electron density (RED) curve may have an indirect impact on the accuracy of dose calculation by a treatment planning system (TPS). This effect has not been previously quantified for input of different CT-RED curves from different CT-scan units in the Monaco TPS. This study aims to evaluate the effect of CT-RED curve on the dose calculation by the Monaco radiotherapy TPS. Four CT images of the CIRS phantom were obtained by different CT scanners. The accuracy of the dose calculation in the three algorithms of the Monaco TPS (Monte Carlo, collapse cone, and pencil beam) is also evaluated based on TECDOC 1583. The CT-RED curves from the CT scanners were transferred to the Monaco TPS to audit the different algorithms of the TPS. The dose values were measured with an ionization chamber in the CIRS phantom. Then, the dose values were calculated by the Monaco algorithms in the corresponding points. For the Monaco TPS and based on TECDOC 1583, the accuracy of the dose calculation in all the three algorithms is within the agreement criteria for most of the points evaluated. For low dose regions, the differences between the calculated and measured dose values are higher than the agreement criteria in a number of points. For the majority of the points, the algorithms underestimate the calculated dose values. It was also found that the use of different CT-RED curves can lead to minor discrepancies in the dose calculation by the Monaco TPS, especially in low dose regions. However, it appears that these differences are not clinically significant in most of the cases.

Identifiants

pubmed: 30848440
doi: 10.1007/s13246-019-00745-6
pii: 10.1007/s13246-019-00745-6
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

489-502

Auteurs

Mohsen Hasani (M)

Department of Radiotherapy Physics, Cancer Research Centre, Cancer Institute, Tehran University of Medical Sciences, Tehran, Iran.

Bagher Farhood (B)

Department of Medical Physics and Radiology, Faculty of Paramedical Sciences, Kashan University of Medical Sciences, Kashan, Iran. bffarhood@gmail.com.

Mahdi Ghorbani (M)

Cancer Research Center, Shahid Beheshti University of Medical Sciences, Tehran, Iran. mhdghorbani@gmail.com.

Hamideh Naderi (H)

Department of Radiotherapy Physics, Cancer Institute, Qom University of Medical Sciences, Qom, Iran.

Sepideh Saadatmand (S)

Department of Radiotherapy Physics, Cancer Institute, Qom University of Medical Sciences, Qom, Iran.

Saeed Karimkhani Zandi (S)

Department of Radiotherapy Physics, Cancer Institute, Qom University of Medical Sciences, Qom, Iran.

Courtney Knaup (C)

Comprehensive Cancer Centers of Nevada, Las Vegas, NV, USA.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH