Repeatability and Reproducibility of Neonatal Echocardiography: The Copenhagen Baby Heart Study.
Acquisition
Agreement
Echocardiography
Measurements
Neonate
Reliability
Journal
Journal of the American Society of Echocardiography : official publication of the American Society of Echocardiography
ISSN: 1097-6795
Titre abrégé: J Am Soc Echocardiogr
Pays: United States
ID NLM: 8801388
Informations de publication
Date de publication:
07 2019
07 2019
Historique:
received:
01
02
2019
pubmed:
12
5
2019
medline:
11
11
2020
entrez:
12
5
2019
Statut:
ppublish
Résumé
The Copenhagen Baby Heart Study (CBHS) is a population-based cohort study of neonates (N = 25,000), including echocardiography. Echocardiography in neonates is mainly focused on congenital heart disease (CHD), whereas general aspects of cardiac dimensions and function in neonates without CHD remain to be further addressed. This study was conducted to assess the reliability of neonatal echocardiography and validity of echocardiographic methods used in the CBHS. Reliability and agreement were tested for two-dimensional (2D), M-mode, spectral Doppler, and tissue velocity echocardiography for the following. (1) Measurements: seven sonographers independently performed two measurement rounds: (a) measurement of the same 50 echocardiograms (n = 350 echocardiograms measured) and (b) repeated measurement of 25 of the 50 echocardiograms (n = 175 echocardiograms measured). (2) Acquisition: four sonographers independently performed two rounds of echocardiographic acquisition and subsequent measurement of the same 22 neonates (n = 176 acquisitions and measures). Intra- and interobserver variabilities were assessed by determinations of coefficient of variation (CV), intraclass correlation coefficient (ICC), Bland-Altman plot, and 95% limits of agreement. (1) Measurements: we found intra- and interobserver ICC ≥ 0.67 for 2D parameters, except for left ventricular (LV) wall thicknesses and LV diameter (interobserver); ICC ≥ 0.84 for tricuspid annular plane systolic excursion (TAPSE); ICC ≥ 0.93 for pulsed-wave Doppler (PW); ICC ≥ 0.84 for continuous-wave Doppler; and ICC ≥ 0.87 for tissue velocity parameters. We found CV < 15% for all parameters except LV wall thicknesses. (2) Acquisition: we found intra- and interobserver ICC ≥ 0.69 for 2D parameters, except for LV wall thicknesses, aortic valve annulus (interobserver), and LV end-systolic diameter (interobserver); ICC = 0.45-0.49 for TAPSE; ICC = 0.48-0.64 for PW; and ICC ≥ 0.70 for continuous wave. We found CV < 15% for all parameters. Reliability of echocardiographic measurements and acquisition of cardiac dimensions and function were good for most parameters but lower for TAPSE (acquisition) and PW Doppler (acquisition) and poor for LV wall thicknesses. In general, echocardiography of cardiac dimensions and function in the neonate is reliable.
Sections du résumé
BACKGROUND
The Copenhagen Baby Heart Study (CBHS) is a population-based cohort study of neonates (N = 25,000), including echocardiography. Echocardiography in neonates is mainly focused on congenital heart disease (CHD), whereas general aspects of cardiac dimensions and function in neonates without CHD remain to be further addressed.
PURPOSE
This study was conducted to assess the reliability of neonatal echocardiography and validity of echocardiographic methods used in the CBHS.
METHODS
Reliability and agreement were tested for two-dimensional (2D), M-mode, spectral Doppler, and tissue velocity echocardiography for the following. (1) Measurements: seven sonographers independently performed two measurement rounds: (a) measurement of the same 50 echocardiograms (n = 350 echocardiograms measured) and (b) repeated measurement of 25 of the 50 echocardiograms (n = 175 echocardiograms measured). (2) Acquisition: four sonographers independently performed two rounds of echocardiographic acquisition and subsequent measurement of the same 22 neonates (n = 176 acquisitions and measures). Intra- and interobserver variabilities were assessed by determinations of coefficient of variation (CV), intraclass correlation coefficient (ICC), Bland-Altman plot, and 95% limits of agreement.
RESULTS
(1) Measurements: we found intra- and interobserver ICC ≥ 0.67 for 2D parameters, except for left ventricular (LV) wall thicknesses and LV diameter (interobserver); ICC ≥ 0.84 for tricuspid annular plane systolic excursion (TAPSE); ICC ≥ 0.93 for pulsed-wave Doppler (PW); ICC ≥ 0.84 for continuous-wave Doppler; and ICC ≥ 0.87 for tissue velocity parameters. We found CV < 15% for all parameters except LV wall thicknesses. (2) Acquisition: we found intra- and interobserver ICC ≥ 0.69 for 2D parameters, except for LV wall thicknesses, aortic valve annulus (interobserver), and LV end-systolic diameter (interobserver); ICC = 0.45-0.49 for TAPSE; ICC = 0.48-0.64 for PW; and ICC ≥ 0.70 for continuous wave. We found CV < 15% for all parameters.
CONCLUSIONS
Reliability of echocardiographic measurements and acquisition of cardiac dimensions and function were good for most parameters but lower for TAPSE (acquisition) and PW Doppler (acquisition) and poor for LV wall thicknesses. In general, echocardiography of cardiac dimensions and function in the neonate is reliable.
Identifiants
pubmed: 31076139
pii: S0894-7317(19)30101-4
doi: 10.1016/j.echo.2019.02.015
pii:
doi:
Banques de données
ClinicalTrials.gov
['NCT02753348']
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM
Pagination
895-905.e2Commentaires et corrections
Type : CommentIn
Informations de copyright
Copyright © 2019 American Society of Echocardiography. Published by Elsevier Inc. All rights reserved.