Comparative LC-MS/MS and HPLC-UV Analyses of Meropenem and Piperacillin in Critically Ill Patients.
Anti-Bacterial Agents
/ blood
Chromatography, High Pressure Liquid
/ instrumentation
Chromatography, Liquid
/ methods
Critical Care
/ methods
Critical Illness
Drug Monitoring
/ methods
Humans
Meropenem
/ blood
Piperacillin
/ blood
Reproducibility of Results
Tandem Mass Spectrometry
/ methods
Ultraviolet Rays
Journal
Clinical laboratory
ISSN: 1433-6510
Titre abrégé: Clin Lab
Pays: Germany
ID NLM: 9705611
Informations de publication
Date de publication:
01 Sep 2019
01 Sep 2019
Historique:
entrez:
19
9
2019
pubmed:
19
9
2019
medline:
11
2
2020
Statut:
ppublish
Résumé
Therapeutic drug monitoring (TDM) of beta-lactam antibiotics has become a valuable tool to guide dosing in critically ill patients. The main goal of the study was to compare two routinely used techniques for beta-lactam TDM in intensive care unit (ICU) patient samples, namely isotope dilution liquid chromatography tandem mass spectrometry (ID-LC-MS/MS) and high-performance liquid chromatography combined with ultra-violet detection (HPLC-UV). A set of 80 sera/plasma samples from ICU patients receiving therapeutic meropenem or piperacillin dosage was investigated. Sample duplicates and quality assessment samples were assayed in parallel with an in-house LC-MS/MS and a commercially available IVD HPLC-UV kit. A pharmacokinetic and pharmacodynamic (PK/PD) target with ≥ 22.5 mg/L for piperacillin and ≥ 8.0 mg/L for meropenem was used for medical assessment of trough sample (n = 40) antibiotic concentrations. There was no difference between serum and Li-heparin plasmas. Concentration deviations were found for 4% of meropenem and 17% of piperacillin samples. Eliminating the influence of the systemic bias of approximately 10% for piperacillin, measurement discrepancies ≥ 25% between LC-MS/MS and HPLC-UV analyses were only observed for ≈ 4 - 6% of all samples. In the same way, identical PK/PD target attainment rates of 50 - 60% could be obtained. After correction of the analytical bias for piperacillin measurements, both methods showed comparable results, also with respect to clinical decision limits. HPLC-UV analysis is an adequate TDM methodology for testing of beta-lactam antibiotics in centers where no special knowledge in LC-MS/MS based TDM is present. However, potential matrix effects, interferences, and calibration issues for both methods must be taken into account.
Sections du résumé
BACKGROUND
BACKGROUND
Therapeutic drug monitoring (TDM) of beta-lactam antibiotics has become a valuable tool to guide dosing in critically ill patients. The main goal of the study was to compare two routinely used techniques for beta-lactam TDM in intensive care unit (ICU) patient samples, namely isotope dilution liquid chromatography tandem mass spectrometry (ID-LC-MS/MS) and high-performance liquid chromatography combined with ultra-violet detection (HPLC-UV).
METHODS
METHODS
A set of 80 sera/plasma samples from ICU patients receiving therapeutic meropenem or piperacillin dosage was investigated. Sample duplicates and quality assessment samples were assayed in parallel with an in-house LC-MS/MS and a commercially available IVD HPLC-UV kit. A pharmacokinetic and pharmacodynamic (PK/PD) target with ≥ 22.5 mg/L for piperacillin and ≥ 8.0 mg/L for meropenem was used for medical assessment of trough sample (n = 40) antibiotic concentrations.
RESULTS
RESULTS
There was no difference between serum and Li-heparin plasmas. Concentration deviations were found for 4% of meropenem and 17% of piperacillin samples. Eliminating the influence of the systemic bias of approximately 10% for piperacillin, measurement discrepancies ≥ 25% between LC-MS/MS and HPLC-UV analyses were only observed for ≈ 4 - 6% of all samples. In the same way, identical PK/PD target attainment rates of 50 - 60% could be obtained.
CONCLUSIONS
CONCLUSIONS
After correction of the analytical bias for piperacillin measurements, both methods showed comparable results, also with respect to clinical decision limits. HPLC-UV analysis is an adequate TDM methodology for testing of beta-lactam antibiotics in centers where no special knowledge in LC-MS/MS based TDM is present. However, potential matrix effects, interferences, and calibration issues for both methods must be taken into account.