Effect of loading frequency on deformations at the bone-implant interface.


Journal

Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine
ISSN: 2041-3033
Titre abrégé: Proc Inst Mech Eng H
Pays: England
ID NLM: 8908934

Informations de publication

Date de publication:
Dec 2019
Historique:
pubmed: 29 9 2019
medline: 9 4 2020
entrez: 28 9 2019
Statut: ppublish

Résumé

This study considers the time-dependent behaviour of bone in the context of loosening of metal implants, which is one of the typical complications of joint replacement and fracture-fixation surgeries. We employed viscoelastic properties developed from our previous experimental studies for trabecular bone in a representative bone-implant construct, which was subjected to cyclic loading at varying loading frequencies. We found that the separation between the bone and the implant is a function of loading frequency and increases with number of loading cycles applied. Our analysis shows that at the start of cyclic loading, a higher frequency results in a lower displacement response of bone at the bone-implant interface; however, after the bone-implant system has been subjected to a large number of cycles (>500 cycles in this study), higher interfacial displacements are observed at higher loading frequencies. In other words, higher loading frequencies will not result in bone-implant separation if limited number of cycles are applied. In all cases, interfacial displacements increase as bone volume ratio decreases. This simple approach can be used to evaluate the mechanical environment in bone-implant systems due to cyclic loading which commonly used time-independent models that are unable to simulate. The approach can also be used to evaluate implant loosening due to cyclic loading.

Identifiants

pubmed: 31560261
doi: 10.1177/0954411919877970
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

1219-1225

Auteurs

Shuqiao Xie (S)

School of Engineering, Institute for Bioengineering, The University of Edinburgh, Edinburgh, UK.

Krishnagoud Manda (K)

School of Engineering, Institute for Bioengineering, The University of Edinburgh, Edinburgh, UK.
School of Mechanical and Aerospace Engineering, Queen's University Belfast, Belfast, UK.

Pankaj Pankaj (P)

School of Engineering, Institute for Bioengineering, The University of Edinburgh, Edinburgh, UK.

Articles similaires

Hemiarthroplasty in young patients.

Hazimah Mahmud, Dong Wang, Andra Topan-Rat et al.
1.00
Humans Male Hemiarthroplasty Middle Aged Aged

High-throughput Bronchus-on-a-Chip system for modeling the human bronchus.

Akina Mori, Marjolein Vermeer, Lenie J van den Broek et al.
1.00
Humans Bronchi Lab-On-A-Chip Devices Epithelial Cells Goblet Cells
Silicon Dioxide Water Hot Temperature Compressive Strength X-Ray Diffraction
Humans Arthroplasty, Replacement, Knee Osteoarthritis, Knee Awards and Prizes Biomechanical Phenomena

Classifications MeSH