BrachyView: Reconstruction of seed positions and volume of an LDR prostate brachytherapy patient plan using a baseline subtraction algorithm.
Baseline subtraction
Brachytherapy
Intraoperative dosimetry
Prostate
Journal
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
ISSN: 1724-191X
Titre abrégé: Phys Med
Pays: Italy
ID NLM: 9302888
Informations de publication
Date de publication:
Oct 2019
Oct 2019
Historique:
received:
09
04
2019
revised:
13
09
2019
accepted:
19
09
2019
pubmed:
30
9
2019
medline:
15
2
2020
entrez:
30
9
2019
Statut:
ppublish
Résumé
BrachyView is a novel in-body imaging system developed with the objective to provide real-time intraoperative dosimetry for low dose rate (LDR) prostate brachytherapy treatments. The BrachyView coordinates combined with conventional transrectal ultrasound (TRUS) imaging, provides the possibility to localise the effective position of the implanted seeds inside the prostate volume, providing a unique tool for intra-operative verification of the quality of the implantation. This research presents the first complete LDR brachytherapy plan reconstructed by the BrachyView system and is used to evaluate the effectiveness of an imaging algorithm with baseline subtraction. A plan featuring 98 I-125 brachytherapy seeds, with an average activity of 0.248 mCi, were implanted into a prostate gel phantom under TRUS guidance. Images of implanted seeds were obtained by the BrachyView after the implantation of seeds. The baseline subtraction algorithm is applied as a pixel-to-pixel counts subtraction and is applied to every second projection obtained after the implantation of each needle. Seed positions and effectiveness of the baseline reconstruction in the identification of seeds were verified by a high-resolution post-implant CT scan. A complete brachytherapy plan has been reconstructed with a 100% detection rate. This is possible due to the effectiveness of the baseline subtraction, with its application an overall increase of 11.3% in position accuracy and 8.2% increase in detection rate was noted. It has been demonstrated that the BrachyView system shows the potential to be a solution to providing clinics with the means for intraoperative dosimetry for LDR prostate brachytherapy treatments.
Identifiants
pubmed: 31563727
pii: S1120-1797(19)30444-2
doi: 10.1016/j.ejmp.2019.09.237
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
66-76Informations de copyright
Copyright © 2019 Associazione Italiana di Fisica Medica. All rights reserved.