Method development for quantitative determination of seven statins including four active metabolites by means of high-resolution tandem mass spectrometry applicable for adherence testing and therapeutic drug monitoring.
Atorvastatin
/ blood
Cardiovascular Diseases
/ drug therapy
Chromatography, High Pressure Liquid
/ methods
Drug Monitoring
/ methods
Guidelines as Topic
Humans
Hydroxymethylglutaryl-CoA Reductase Inhibitors
/ blood
Limit of Detection
Lovastatin
/ blood
Medication Adherence
Pravastatin
/ blood
Quality Control
Reference Standards
Tandem Mass Spectrometry
/ methods
HMG-CoA reductase inhibitors
TDM
adherence monitoring
cardiovascular diseases
statins
Journal
Clinical chemistry and laboratory medicine
ISSN: 1437-4331
Titre abrégé: Clin Chem Lab Med
Pays: Germany
ID NLM: 9806306
Informations de publication
Date de publication:
28 04 2020
28 04 2020
Historique:
received:
25
07
2019
accepted:
19
09
2019
pubmed:
31
10
2019
medline:
21
5
2021
entrez:
31
10
2019
Statut:
ppublish
Résumé
Background Statins are used to treat and prevent cardiovascular diseases (CVDs) by reducing the total serum cholesterol concentration. Unfortunately, dose-related side effects and sub-optimal response, attributed to non-adherence amongst others, were described. Therefore, a fast and sensitive liquid chromatography-high-resolution tandem mass spectrometry (LC-HRMS/MS) method for adherence testing and therapeutic drug monitoring of all currently marketed statins and their active metabolites in human blood plasma should be developed, validated and tested for applicability. Methods Atorvastatin, fluvastatin, lovastatin, pitavastatin, pravastatin, rosuvastatin, and simvastatin, as well as ortho- and para-hydroxy-atorvastatin, lovastatin hydroxy acid and simvastatin hydroxy acid were included and several internal standards (IS) tested. Validation was performed according to the guideline of the European Medicines Agency including selectivity, carry-over, accuracy, precision, matrix effects, dilution integrity and analyte stability. Finally, applicability was tested using 14 patient samples submitted for regular toxicological analysis. Results Due to an analytical interference of atorvastatin-d5, diazepam-d5 and pentobarbital-d5 were chosen as IS for positive and negative ionization mode, respectively. All statins and metabolites fulfilled the validation acceptance criteria except for fluvastatin, which could not be quantified reliably and reproducibly, most probably due to instability. Analyses of human plasma samples revealed concentrations of statins and metabolites below the reference plasma concentrations in the case of eight patients. However, nothing was known concerning patients' adherence and time between intake and sampling. Conclusions An LC-HRMS/MS method for identification and quantification of atorvastatin, lovastatin, pitavastatin, pravastatin, rosuvastatin, simvastatin and four active metabolites was successfully developed and applicability demonstrated.
Identifiants
pubmed: 31665111
doi: 10.1515/cclm-2019-0763
pii: /j/cclm.ahead-of-print/cclm-2019-0763/cclm-2019-0763.xml
doi:
pii:
Substances chimiques
Hydroxymethylglutaryl-CoA Reductase Inhibitors
0
Lovastatin
9LHU78OQFD
Atorvastatin
A0JWA85V8F
Pravastatin
KXO2KT9N0G
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM