Feasibility of a circulation model for the assessment of endovascular recanalization procedures and periprocedural thromboembolism in-vitro.


Journal

Scientific reports
ISSN: 2045-2322
Titre abrégé: Sci Rep
Pays: England
ID NLM: 101563288

Informations de publication

Date de publication:
22 11 2019
Historique:
received: 04 03 2019
accepted: 29 10 2019
entrez: 24 11 2019
pubmed: 24 11 2019
medline: 25 11 2020
Statut: epublish

Résumé

Aim of this study was to establish a simple and highly reproducible physiological circulation model to investigate endovascular device performance. The developed circulation model included a pneumatically driven pulsatile pump to generate a flow rate of 2.7 L/min at 70 beats per minute. Sections from the superficial femoral arteries were used in order to simulate device/tissue interaction and a filter was integrated to analyze periinterventional thromboembolism of white, red and mixed thrombi. The working fluid (3 L) was a crystalloid solution constantly tempered at 36.5 °C. To evaluate the model, aspiration thrombectomy, stent-implantation and thrombectomy with the Fogarty catheter were performed. Usability of the model was measured by the System Usability Scale (SUS) - Score. Histological specimens were prepared and analyzed postinterventional to quantify tissue/device interaction. Moreover, micro- and macroembolism were evaluated for each thrombus entity and each device. Results were tested for normality using the D'Agostino-Pearson test. Statistical comparisons of two groups were performed using the Student's t-test. All devices were able to remove the occlusions after a maximum of 2 attempts. First-pass-recanalization was not fully achieved for aspiration thrombectomy of mixed thrombi (90.6%), aspiration thrombectomy of red thrombi (84.4%) and stent-implantation in occlusions of red thrombi (92.2%). Most micro- and macroembolism were observed using the Fogarty catheter and after stent-implantation in occlusions of white thrombi. Histological examinations revealed a significant reduction of the vascular layers suggesting vascular damage after use of the Fogarty catheter (327.3 ± 3.5 μm vs. 440.6 ± 3.9 μm; p = 0.026). Analysis of SUS rendered a mean SUS-Score of 80.4 which corresponds to an excellent user acceptability of the model. In conclusion, we describe a stable, easy to handle and reproducible physiological circulation model for the simulation of endovascular thrombectomy including device performance and thromboembolism.

Identifiants

pubmed: 31757980
doi: 10.1038/s41598-019-53607-2
pii: 10.1038/s41598-019-53607-2
pmc: PMC6874641
doi:

Types de publication

Evaluation Study Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

17356

Références

Biomed Res Int. 2017;2017:2362769
pubmed: 28555191
J Stroke. 2016 Sep;18(3):286-296
pubmed: 27733029
Adv Exp Med Biol. 2017;906:195-213
pubmed: 27664152
Expert Rev Cardiovasc Ther. 2015 Apr;13(4):429-44
pubmed: 25728744
J Neurointerv Surg. 2018 Oct;10(10):953-957
pubmed: 29463619
Vascular. 2018 Apr;26(2):163-168
pubmed: 28828935
J Thromb Thrombolysis. 2017 Apr;43(3):352-360
pubmed: 28070820
J Vis Exp. 2014 Aug 20;(90):
pubmed: 25178087
Curr Cardiol Rep. 2017 Aug;19(8):68
pubmed: 28646446
J Vasc Surg. 2001 May;33(5):1128-30
pubmed: 11331864
Annu Rev Biomed Eng. 1999;1:427-62
pubmed: 11701496
Blood. 2017 Oct 19;130(16):1795-1799
pubmed: 28811305
J Vasc Surg. 2017 Sep;66(3):891-901
pubmed: 27693032
ANZ J Surg. 2014 Nov;84(11):877-80
pubmed: 23953009

Auteurs

René Rusch (R)

Department of Cardiovascular Surgery, University Hospital of Schleswig-Holstein, Kiel, Germany. rene.rusch@uksh.de.

Jens Trentmann (J)

Department of Radiology and Neuroradiology, University Hospital of Schleswig-Holstein, Kiel, Germany.

Lars Hummitzsch (L)

Department of Anesthesiology and Intensive Care Medicine, University Hospital of Schleswig-Holstein, Kiel, Germany.

Melanie Rusch (M)

Department of Orthopedics and Trauma Surgery, University Hospital of Schleswig-Holstein, Kiel, Germany.

Schekeb Aludin (S)

Department of Radiology and Neuroradiology, University Hospital of Schleswig-Holstein, Kiel, Germany.

Assad Haneya (A)

Department of Cardiovascular Surgery, University Hospital of Schleswig-Holstein, Kiel, Germany.

Martin Albrecht (M)

Department of Anesthesiology and Intensive Care Medicine, University Hospital of Schleswig-Holstein, Kiel, Germany.

Jost Philipp Schäfer (JP)

Department of Radiology and Neuroradiology, University Hospital of Schleswig-Holstein, Kiel, Germany.

Thomas Puehler (T)

Department of Cardiovascular Surgery, University Hospital of Schleswig-Holstein, Kiel, Germany.

Jochen Cremer (J)

Department of Cardiovascular Surgery, University Hospital of Schleswig-Holstein, Kiel, Germany.

Rouven Berndt (R)

Department of Cardiovascular Surgery, University Hospital of Schleswig-Holstein, Kiel, Germany.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH