Impedance Based Automatic Detection of Ventilations During Mechanical Cardiopulmonary Resuscitation.
Journal
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
ISSN: 2694-0604
Titre abrégé: Annu Int Conf IEEE Eng Med Biol Soc
Pays: United States
ID NLM: 101763872
Informations de publication
Date de publication:
Jul 2019
Jul 2019
Historique:
entrez:
18
1
2020
pubmed:
18
1
2020
medline:
23
2
2020
Statut:
ppublish
Résumé
Monitoring ventilation rate is key to improve the quality of cardiopulmonary resuscitation (CPR) and increase the probability of survival in the event of an out-of-hospital cardiac arrest (OHCA). Ventilations produce discernible fluctuations in the thoracic impedance signal recorded by defibrillators. Impedance-based detection of ventilations during CPR is challenging due to chest compression artifacts. This study presents a method for an accurate detection of ventilations when chest compressions are delivered using a piston-driven mechanical device. Data from 223 OHCA patients were analyzed and 399 analysis segments totaling 3101 minutes of mechanical CPR were extracted. A total of 18327 ventilations were annotated using concurrent capnogram recordings. An adaptive least mean squares filter was used to remove compression artifacts. Potential ventilations were detected using a greedy peak detector, and the ventilation waveform was characterized using 8 waveform features. These features were used in a logistic regression classifier to discriminate true ventilations from false positives produced by the greedy peak detector. The classifier was trained and tested using patient wise 10-fold cross validation (CV), and 100 random CV partitions were created to statistically characterize the performance metrics. The peak detector presented a sensitivity (Se) of 99.30%, but a positive predictive value (PPV) of 54.43%. The best classifier configuration used 6 features and improved the mean (sd) Se and PPV of the detector to 93.20% (0.06) and 94.43% (0.04), respectively. When used to measure per minute ventilation rates for feedback to the rescuer, the mean (sd) absolute error in ventilation rate was 0.61 (1.64) min
Identifiants
pubmed: 31945835
doi: 10.1109/EMBC.2019.8856822
doi:
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM