Tricuspid Annuloplasty Rings: A Quantitative Comparison of Size, Nonplanar Shape, and Stiffness.


Journal

The Annals of thoracic surgery
ISSN: 1552-6259
Titre abrégé: Ann Thorac Surg
Pays: Netherlands
ID NLM: 15030100R

Informations de publication

Date de publication:
11 2020
Historique:
received: 05 11 2019
revised: 05 02 2020
accepted: 26 02 2020
pubmed: 7 4 2020
medline: 15 12 2020
entrez: 7 4 2020
Statut: ppublish

Résumé

Functional tricuspid regurgitation due to annular and ventricular dilatation is increasingly recognized as a significant source of morbidity and mortality. To repair the annulus, surgeons implant one of many annuloplasty devices that differ in size, 3-dimensional (3D) shape, and stiffness. However, there have been no quantitative comparisons between various available devices. Three-dimensional scanning, micro-computed tomography imaging, analytical methods, and mechanical tests were used to compare 3 Edwards Lifesciences (Irvine, CA) and 3 Medtronic (Minneapolis, MN) annuloplasty devices of all available sizes. We measured in-plane metrics of maximum diameter, perimeter, area, height, as well as elevation and curvature profiles. Furthermore, we computed bending stiffness as well as the maximum and minimum axes of the bending stiffness. Most annular prostheses differed little in their in-plane geometries but varied significantly in height. In-plane properties deviated significantly from measurements of healthy human tricuspid annuli. Height of the Edwards' MC3 and Medtronic's Contour 3D resembled healthy human tricuspid valve annuli, whereas the Edwards' Physio and Classic, and Medtronic's TriAd, did not. Additionally, the elevation profiles of the MC3 and Contour 3D and curvature profiles between all devices were consistent and matched those of healthy human annuli. The tested devices also differed in their bending stiffness, both in terms of absolute values and their maximum and minimum axes. Contoured devices, such as Edwards' MC3 and Medtronic's Contour 3D, most accurately resembled the healthy human tricuspid annulus but differed significantly in bending stiffness. To what extent prosthesis properties and shape affect tricuspid valve function remains to be determined.

Sections du résumé

BACKGROUND
Functional tricuspid regurgitation due to annular and ventricular dilatation is increasingly recognized as a significant source of morbidity and mortality. To repair the annulus, surgeons implant one of many annuloplasty devices that differ in size, 3-dimensional (3D) shape, and stiffness. However, there have been no quantitative comparisons between various available devices.
METHODS
Three-dimensional scanning, micro-computed tomography imaging, analytical methods, and mechanical tests were used to compare 3 Edwards Lifesciences (Irvine, CA) and 3 Medtronic (Minneapolis, MN) annuloplasty devices of all available sizes. We measured in-plane metrics of maximum diameter, perimeter, area, height, as well as elevation and curvature profiles. Furthermore, we computed bending stiffness as well as the maximum and minimum axes of the bending stiffness.
RESULTS
Most annular prostheses differed little in their in-plane geometries but varied significantly in height. In-plane properties deviated significantly from measurements of healthy human tricuspid annuli. Height of the Edwards' MC3 and Medtronic's Contour 3D resembled healthy human tricuspid valve annuli, whereas the Edwards' Physio and Classic, and Medtronic's TriAd, did not. Additionally, the elevation profiles of the MC3 and Contour 3D and curvature profiles between all devices were consistent and matched those of healthy human annuli. The tested devices also differed in their bending stiffness, both in terms of absolute values and their maximum and minimum axes.
CONCLUSIONS
Contoured devices, such as Edwards' MC3 and Medtronic's Contour 3D, most accurately resembled the healthy human tricuspid annulus but differed significantly in bending stiffness. To what extent prosthesis properties and shape affect tricuspid valve function remains to be determined.

Identifiants

pubmed: 32251659
pii: S0003-4975(20)30440-9
doi: 10.1016/j.athoracsur.2020.02.064
pii:
doi:

Types de publication

Comparative Study Journal Article Research Support, Non-U.S. Gov't

Langues

eng

Sous-ensembles de citation

IM

Pagination

1605-1614

Informations de copyright

Copyright © 2020 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Auteurs

Mrudang Mathur (M)

Walker Department of Mechanical Engineering, The University of Texas at Austin, Austin, Texas.

Marcin Malinowski (M)

Division of Cardiothoracic Surgery, Spectrum Health, Grand Rapids, Michigan; Department of Cardiac Surgery, Medical University of Silesia School of Medicine in Katowice, Katowice, Poland.

Tomasz A Timek (TA)

Division of Cardiothoracic Surgery, Spectrum Health, Grand Rapids, Michigan.

Manuel K Rausch (MK)

Department of Aerospace Engineering and Engineering Mechanics, The University of Texas at Austin, Austin, Texas; Department of Biomedical Engineering, The University of Texas at Austin, Austin, TX; Oden Institute for Computational Engineering and Science, The University of Texas at Austin, Austin, Texas. Electronic address: manuel.rausch@utexas.edu.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH